双源CT大螺距扫描模式在儿童胸部低剂量成像中的应用价值 被引量:7

The application value of low-dose of high-pitch scan mode of dual-source computed tomography in the chest of children
下载PDF
导出
摘要 目的探讨双源CT大螺距扫描模式在儿童胸部成像中的辐射剂量及图像质量的应用价值。方法将80例行双源CT胸部检查的患儿随机分为A、B两组。其中A组40例,行大螺距模式扫描,螺距3.0;B组40例,行常规模式扫描,螺距1.2。对A、B两组图像质量、辐射剂量(CT容积剂量指数CTDIvol值、剂量长度乘积DLP值及有效剂量E)及扫描时间进行对比分析。结果 A、B两组均获得了满足临床诊断的图像,图像质量差异无统计学意义(Z=-1.75,P=0.08);A、B两组患儿的CTDIvol值分别平均为1.82mGy、4.50mGy,差异具有统计学意义(t=-32.50,P=0.00),A组患儿较B组患儿下降59.56%。A、B两组患儿的DLP值分别平均为38.41mGy/cm、82.74mGy/cm,差异具有统计学意义(t=-37.94,P=0.00),A组患儿的DLP值较B组患儿下降53.58%。A组患儿的平均扫描时间(0.27s)明显短于B组患儿(2.60s),差异有统计学意义(t=41.31,P=0.00),A组扫描时间仅为B组的10.38%。结论双源CT大螺距扫描模式具有扫描速度快、辐射剂量低等优点,在儿童胸部低剂量扫描中具有重要价值。 Objective To evaluate the radiation dose and image quality of the application of high‐pitch scan mode of dual -source computed tomography in the chest of children .Methods 80 sick children who were going to do chest CT examination by dual‐source CT system were randomly divided into two groups of A B .40 children of group A were examined with a in high‐pitch scan mode(pitch=3.0);40 patients of group B were examined with in conventional pitch mode(pitch=1.2) .The image quality ,radiation dose(CTDIvol ,DLP) and scaning time between the two groups were analysed .Results The Image quality of group A and B meet the requirements of the clinical diagnosis .There is no significant difference in the image quality of group A and B(Z= -1.75 ,P=0.08);The average CTDIvol value of group A and B was 1.82 mGy、4.50 mGy .The CTDIvol value of group A was less than group B(t= -32.5 ,P=0.00) .The CTDIvol value of group A was decreased by 59.56% than group B . The average DLP value of group A and B was 38.41 mGy、82.74 mGy/cm .The average DLP value of group A and B was 38.41 mGy、82.74 mGy .The DLP value of group A was less than group B(t= -37.94 ,P=0.00) .The DLPl value of group A was decreased by 53.58% than group B .The scanning time of group A(0.27 s) was shorter than group B(2.60 s) (t=41.00 ,P=0.00) .The scanning time of group A(0.27 s) was less than group B(2.60 s)(t=41.34 ,P=0.00) .The scanning time of group A was accounted for only 10.38% of group B .Conclusion The high‐pitch scan mode of dual‐source computed tomography has higher scanning speed ,lower radiation dose in the chest ,there is important value in the loe‐dose scanning of children's chest .
出处 《滨州医学院学报》 2015年第1期47-49,共3页 Journal of Binzhou Medical University
关键词 双源CT 大螺距 低剂量 儿童 胸部 dual source CT high-pitch low-dose children chest
  • 相关文献

参考文献10

  • 1Paul JF,Rohnean A,Sigal-Cinqualbre A.Multidetector CT for congenital heart patients:what a paediatric radiologist should know[J].Pediatr Radiol,2010,40(6):869-875.
  • 2Mettler FA Jr,Wiest PW,Locken JA,et al.CT scanning:pattens of use and dose[J].J Radiol Prot,2000,20(4):353-359.
  • 3McNitt-Gray MF.AAPM/RSNA Physics Tutorial for Residents:Topics in CT.Radiation dose in CT[J].Radiographics,2002,22(6):1541-1553.
  • 4Mayo JR,Aldrich J,Muller NL.Radiation exposure at chest CT:a statement of the Fleischner Society[J].Radiology,2003,228(1):15-21.
  • 5Lell M,Marwan M,Schepis T,et al.Prospectively ECG-triggered high-pitch spiral acquisition for coronary CT angiography using dual source CT:technique and initial experience[J].Eur Radial,2009,19(11):2576-2583.
  • 6Reimann AJ,Rinek D,Birinci-Aydogan A,et al.Dual-source computed tomography:advances of improved temporal resolution in coronary plaque imaging[J].Invest Radiol,2007,42(3):196-203.
  • 7Sommer WH,Schenzle JC,Becker CR,et al.Saving dose in triple-rule-out computed tomography examination using a highpitch dual spiral technique[J].Invest Radiol,2010,45(2):64-71.
  • 8Goetti R,Leschka S,Baumuller S,et al.Low dose high-pitch spiral acquisition 128-slice dual-source computed tomography for the evaluation of coronary artery bypass graft patency[J].Invest Radiol,2010,45(6):324-330.
  • 9苗英,钱玉娥,张月俏.16层螺旋CT低剂量扫描在女性骨盆中的应用[J].中华放射医学与防护杂志,2006,26(1):90-91. 被引量:29
  • 10Achenbach S,Marwan M,Ropers D,et al.Coronary computed tomography angiography with a consistent dose below 1mSv using prospectively electrocardiogram-triggered high-pitch spiral acquisition[J].Eur Heart J.2010,31(3):340-346.

二级参考文献9

  • 1刘昌盛,魏文洲,郑晓华,童世平,殷响林,杨洪飙.低剂量CT扫描对婴幼儿颅脑病变检查的防护价值[J].中华放射医学与防护杂志,2004,24(3):270-271. 被引量:67
  • 2赵修义,李国英,陈静,肖红,刘汝勤.CT检查的防护探讨[J].中华放射医学与防护杂志,1995,15(1):57-57. 被引量:10
  • 3Frush DP,Donnelly LF,Rosen NS.Computed tomography and radiation risks:what pediatric health care providers should know.Pediatrics,2003,112:951-957.
  • 4Hidajat N,Wolf M,Nunnemann A,et al.Survey of conventional and spiral CT doses.Radiology,2001,218:395-401.
  • 5Jurik AG.Technique and radiation dose of conventional X-rays and computed tomography of the sacroiliac joint.Radiologe,2004,44:229-233.
  • 6Goddard CC,al-Farsi A.Radiation doses from CT in the Sultanate of Oman.Br J Radiol,1999,72:1073-1077.
  • 7Hidajat N,Maurer J,Schroder R J,et al.Relationships between physical dose quantities and patient dose in CT.Br J Radiol,1999,72:556-561.
  • 8贾明轩,张煊,刘为,王强.CT检查所致病人剂量的简便估算方法[J].中华放射医学与防护杂志,2001,21(5):374-376. 被引量:30
  • 9张朝利,耿振林,费军,王宗烨,焦兴华.多层螺旋CT低mAs扫描技术在胸部的应用[J].中华放射医学与防护杂志,2004,24(1):74-74. 被引量:3

共引文献28

同被引文献35

引证文献7

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部