摘要
The stability parameters of implants (ITV, ISQ & PTV) according to different sizes of controlled bone defects made in implant osteotomies were analyzed and the correlation among the three kinds of implant stability parameters was tested in this study. 45 tapped screw-type dental implants were inserted in three types of implant osteotomies made in 8 fresh-frozen pig femoral bones: Typel - without coronal bone defect, Type2 - with 3 mm coronal bone defects, and Type3 - with 6 mm coronal bone defects. The insertion torque values, ISQ & PTV of implants were measured and analyzed statistically. It is concluded that the circumferential coronal bone defects statistically influence the primary stability of implants; ITV, ISQ and PTV are suitable and available to detect the peri-implant coronal bone defects in 3 mm increments, and ITV and PTV are more sensitive to coronal cortical bone loss. There was a strong correlation between ITV and ISQ.
The stability parameters of implants (ITV, ISQ & PTV) according to different sizes of controlled bone defects made in implant osteotomies were analyzed and the correlation among the three kinds of implant stability parameters was tested in this study. 45 tapped screw-type dental implants were inserted in three types of implant osteotomies made in 8 fresh-frozen pig femoral bones: Typel - without coronal bone defect, Type2 - with 3 mm coronal bone defects, and Type3 - with 6 mm coronal bone defects. The insertion torque values, ISQ & PTV of implants were measured and analyzed statistically. It is concluded that the circumferential coronal bone defects statistically influence the primary stability of implants; ITV, ISQ and PTV are suitable and available to detect the peri-implant coronal bone defects in 3 mm increments, and ITV and PTV are more sensitive to coronal cortical bone loss. There was a strong correlation between ITV and ISQ.
基金
Sponsored by the Scientific Research Foundation for the Returned Overseas Chinese Scholars,Ministry of Education,China