期刊文献+

CG-5重力仪的漂移与寿命 被引量:17

Zero drift of CG-5 gravimeter and its service lifetime
下载PDF
导出
摘要 漂移及其线性度是重力仪最重要的性能指标,在很大程度上决定了其精度和寿命。对14台CG-5重力仪漂移数据的分析表明,多数仪器使用2年后的平均漂移率约在(0.7~1.7)×10^-5m/s2/d之间。以此为基础,结合厂家提供的相关数据,对CG-5重力仪的使用寿命进行了推测。得到北纬40°地区,仪器平均寿命约10年的结论。 The zero drift and its linearity are the most important performance indexes of gravimeter and also considerably dominate the instrumental accuracy and lifetime. Based on an analysis of drift data from 14 CG-5 gravimeters,the authors found that the average drift rates of the majority of instruments are between 0. 7 and 1. 7 × 10^- 5m / s2/ day in two years. On such a basis in combination with the related data provided by manufacturers,the service lifetimes of CG-5 gravimeters have been inferred in this paper. The normal mean operational lifetime seems to be about 10 years in the area of forty degrees north latitude.
出处 《物探与化探》 CAS CSCD 2015年第2期383-386,共4页 Geophysical and Geochemical Exploration
关键词 CG-5重力仪 漂移 使用寿命 仪器性能 CG-5 gravimeter drift service lifetime instrument performance
  • 相关文献

参考文献5

二级参考文献19

  • 1卢红艳,郑金涵,刘端法,郭凤义.利用实测资料确定LCR重力仪的线性格值和周期误差参数[J].地震地磁观测与研究,2004,25(4):20-25. 被引量:5
  • 2梁伟锋,祝意青,徐云马,刘练,王继英.测点周围环境变化对重力观测影响研究[J].地震地磁观测与研究,2005,26(3):50-55. 被引量:10
  • 3曹金国,王来鹏,翟广卿,操玉顺.2007.CG-5重力仪及应用[M].北京:解放军出版社:1--5.
  • 4刘瑞法.LaCoste-Romberg G型重力仪读数精度的提高.地震,1984,4(3):55-57.
  • 5中国地震局.2007.中华人民共和国地震行业标准DB/T 23-2007《地震观测仪器进网技术要求:重力仪》[S].北京:中国标准出版社:2--10.
  • 6王俊勤,申慧群.航天靶场大地工程测量[M].北京:解放军出版社,2007.265-269.
  • 7GJB1167—1991.军用重力测量规则[S].
  • 8Tim Niebauer.LRS相对重力仪http://www.laureltechnolog-ies.com,2011-12-26.
  • 9朱淑霞.CG-2型重力仪格值变化对流动重力段差测量值的影响[J].地壳形变与地震,1982(04):84-93.
  • 10肖凡,邹雪平,张宏伟等.CG-5相对重力仪测量精度分析[J].大地测量,2011,91(1):40-42.