血管内支架径向压缩性能的测试方法 被引量:3

Testing Method of the Radial Compression Performance for Intravascular Stents
下载PDF
导出
摘要 以3种不同直径的聚丙烯(PP)单丝为原料,分别编织4种不同密度共12枚血管内支架.采用平面压缩法在不同上平面尺寸和不同接触界面条件下对支架的径向支撑力、径向压强及弹性回复率进行测试和分析.试验结果表明:使用平面压缩法对支架径向支撑力进行测试时,应选用尺寸大于支架外径的上平面;3种接触面测试支架径向压缩性能时,压缩曲线趋势均一致,测试值有一定差异;采用薄膜包覆支架进行压缩时,压缩曲线较为平滑. Twelve braided intravascular stents of four different levels of braiding tightness were produced with three different diameters of polypropylene(PP)monofilaments,and stents were compressed with with plate compression method by pressure foot with different sizes and different contact interfaces.The radial compression strength,radial intensity of pressure and the elastic recovery rate of these stents were tested by the plate compression method.The experimental results showed that,when plate compression method was used to test the compression performance of stents,the size of pressures should be larger than the outer diameter of the stent.When compressed with three different contact interfaces,the compression curves showed same trend and certain differences on the test value.When the stent was coated with rubber film,the compression curve was smoother.
出处 《东华大学学报(自然科学版)》 CAS CSCD 北大核心 2015年第1期16-21,共6页 Journal of Donghua University(Natural Science)
基金 上海市科委资助项目(10411953300) 高等学校学科创新引智计划资助项目(B07024)
关键词 血管内支架 径向压缩性能 平面压缩法 测试 intravascular stent radial compression performance plate compression method test
  • 相关文献

参考文献8

  • 1SCOT G, PATRICK W, SERRUYS M D. Coronary stents: Current status[J]. Journal of American College of Cardiology, 2010,56(10): 1-42.
  • 2张金山,王茂强.当前我国内支架应用中存在的一些问题[J].中华放射学杂志,1997,31(5):295-297. 被引量:20
  • 3I3ARTKOWIAK JOWSA M, BEDZINSK1 R, KOZLOWSKA A, et al. Mechanical, rheological, fatigue, and degradation behavior of PLLA, PGLA and PDGLA as materials for vascular implants[J]. Meccanica, 2013,48(3) : 721-731.
  • 4杨海山,王田蔚,赵永生,王晓彬,史美龙.三种自制可降解性血管内支架的性能分析[J].中华放射学杂志,2007,41(3):293-296. 被引量:2
  • 5SWAGAT I, SABIT A. Design and characterization of polymaricstents[J]. Journal of Industrial Textiles, 2006,35 (3) : 189-200.
  • 6REGIS R, CATHARINE M. Radial force of coronary stents: A eompararive analysis[J]. Catheterization and Cardiovascular Intervent;,ons, 1999,46(3) : 380-391.
  • 7SUBBU V, TAN L P. Collapse pressures of Biodegradable stents[J]. Biomaterials, 2003,24(12) : 2105-2011.
  • 8张佩华,甄慧英,王聪儿,等.一种手工编织血管内支架用的带手柄打孔模具:中国,CN202589711u[P],2012-12-12.

二级参考文献9

  • 1Colombo A, Karvouni E. Biodegradable stents: 'fulilling the mission and stepping away'. Circulation, 2000, 102:371-373.
  • 2Tamai H, Igaki K, Kyo E, et al. Initial and 6-month results of biodegradable poly-1-Lactic acid coronary stents in humans.Circulation, 2000, 102: 399-404.
  • 3Tsuji T, Tamai H, Kyo E, et al. A new biodegradable coronary stent: acute and 3 months clinical and angiographic follow-up.Circulation, 1999, 100: 289-292.
  • 4欧阳埔.管腔内支架的作用机制和性能//张金山,现代腹部介入放射学.北京:科学出版社,2000:239-241.
  • 5Palmaz JC. Intravascular stents: tissue-stent interactions and design considerations. AIR, 1993, 160: 613-618.
  • 6Heldman AW, Cheng L, Jenkins GM, et al. Paclitaxel stent coating inhibits neointimal hyperplasia at 4 weeks in a porcine model of coronary rostenosis. Circulation, 2001, 103:2289-2295.
  • 7欧阳墉.管腔内支架:Ⅰ.作用机制和性能[J].国外医学(临床放射学分册),1997,20(4):210-213. 被引量:13
  • 8陈锐,陈槐卿,韩君,周大利,郑昌琼.聚L-乳酸(PLLA)/β-磷酸三钙(β-TCP)多孔复合材料的制备及其性能研究[J].生物医学工程学杂志,2001,18(2):177-180. 被引量:12
  • 9吕秀阳,杨亦文,任其龙,吴平东.紫杉醇提取工艺的放大与优化[J].农业工程学报,2001,17(4):119-122. 被引量:8