准分子激光原位角膜磨镶术中眼球动态旋转的临床研究 被引量:2

A clinical study on dynamic cyclotorsion during laser in situ keratomileusis
原文传递
导出
摘要 目的 观察准分子激光原位角膜磨镶术(LASIK)中眼球动态旋转的方向和角度.方法 前瞻性系列病例研究,对近视及近视性散光133例(235只眼)应用阿玛仕准分子激光仪行LASIK手术,按照眼别、术前屈光度分组后,观察术中眼球动态旋转的方向和角度.计量资料采用独立样本t检验进行分析.结果 动态旋转角度范围(绝对值)0°~7.88°.右眼动态旋转角度范围(绝对值)0°~7.77°,平均为1.74°±1.46°;左眼动态旋转角度范围(绝对值)0°~7.88°,平均为1.90°±1.54°,右眼与左眼比较差异无统计学意义(t=0.69,P=0.25).133例(235只眼)均测出了动态旋转:内旋者26只眼(11.06%),旋转角度0°~7.88°,平均为2.03°±2.05°;外旋者12只眼(5.11%),旋转角度0.18°~ 6.81°,平均为2.16°±1.87°;混合性旋转197只眼(83.83%),旋转角度0.01°~5.79°,平均为1.74°±1.34°.全部235眼动态旋转幅度为0.01°~10.5°,平均为3.54°±2.00°;右眼平均幅度为3.26°±1.65°,左眼平均幅度为3.33°±1.71°,两者比较,差异无统计学意义(t=0.93,P =0.20).动态旋转幅度:术前屈光度高度组>屈光度中度组>屈光度低度组,且3组均值比较差异有统计学意义(t=4.10,P=0.001;t=3.04,P=0.002;t =6.06,P=0.001).结论 LASIK术中眼球均发生动态旋转,近视度愈高,旋转角度和总幅度愈大,因此需要实时测量动态旋转并调整激光发射位置进行补偿. Objective To evaluate the direction and degree of dynamic cyclotorsion during laser in situ keratomileusis(LASIK).Methods This retrospective study comprised 235 eyes of 133 patients that had LASIK for myopia with or without astigmatism by Schwind Amaris excimer laser.According to eyes,preoperative diopter of spherical power were divided into groups,dynamic cyclotorsion was assessed during LASIK.Enumeration data were compared with independent samples test.Results Dynamic cyclotorsion degree range from 0° to 7.88° (absolute value).Dynamic cyclotorsion degree of right eyes range from 0° to 7.77° (absolute value)the mean AP was 1.74° ± 1.46°.Dynamic cyclotorsion degree of left eyes range from 0° to 7.88° (absolute value) the mean AP was 1.90° ± 1.54°.There was no significant statistical difference between right eyes and left eyes (t =0.69,P =0.25).235 eyes were successful measured.26 eyes (11.06%) had incyclotorsion,range from 0 degrees to 7.88 degrees,the mean AP was 2.03 ° ± 2.05 °;12 eyes (5.11%) had excyclotorsion,range from 0.18 degrees to 6.81 degrees,the mean AP was 2.16° ± 1.87°;197 eyes(83.83%) had both excyclotorsion and incyclotorsion,range from 0.01 degrees to 5.79 degrees.the mean AP was 1.74° ± 1.34 °.The dynamic cyclotorsion range from 0.01 degrees to 2.24 degrees,the mean AP was 3.54° ± 2.00°.The dynamic cyclotorsion amplitude of right eyes was 3.26° ± 1.65°,left eyes was 3.33° ± 1.71 °.There was no significant statistical difference between right eyes and left eyes(t =0.93,P =0.20).Cyclotorsion amplitude had been compared:high-diopter 〉 mid-diopter 〉 low-diopter.There were statistical differences between three groups(t =4.10,P =0.001;t =3.04,P =0.002;t =6.06,P =0.001).Conclusion Dynamic cyclotorsion occurs during LASIK.The higher myopic degree is,the larger the eye cyclotorsion degree and the total amplitude will be.So Dynamic cyclotorsion should be measured and compensated.
出处 《中华眼外伤职业眼病杂志》 2015年第3期176-179,共4页 Chinese Journal of Ocular Trauma and Occupational Eye Disease
关键词 准分子激光屈光性角膜切削术 眼球动态旋转 Laser in situ keratomileusis(LASIK) Dynamic cyclotorsion
  • 相关文献

参考文献12

  • 1王卫群,蔡慧敏,李铮.应用虹膜识别技术测量准分子激光角膜原位磨镶术中瞳孔中心移位和眼球旋转[J].中华眼外伤职业眼病杂志,2012,34(6):427-430. 被引量:4
  • 2王卫群,张金嵩,赵晓金.有和无虹膜识别波阵面像差引导的准分子激光原位角膜磨镶术治疗近视或近视散光眼临床对比研究[J].中华眼科杂志,2011,47(10):927-933. 被引量:4
  • 3Prakash G, Agarwal A, Kumar DA, et al. Surface ablation with iris recognition and dynamic rotational eye tracking-based tissue saving treatment with the Technolas 217z excimer laser [ J]. J Re- fract Surg, 2011,27(3) :223-231.
  • 4Chang J. Cyclotorsion during laser in situ keratomileusis [ Jl. J Cataract Refract Surg ,2008,34(10) :1720-1726.
  • 5Park SH, Kim M, Joo CK. Measurement of pupil centroid shift pupil and cyelotorsional displacement using iris registration. Oph- thalmologica, 2009,223 ( 3 ) : 166-171.
  • 6Tomita M, Watabe M, Yukawa S, et al. Supplementary effect of static cyelotorsion compensation dynamic with cyclotorsion compen- sation on the refractive and visual outcomes of laser in situ kerato- mileusis for myopic astigmatism I J ] : J Cataract Refract Surg, 2013,39(5 ) :752-758.
  • 7Arba-Mosquera S, Arbelaez MC . Three-month clinical outcomes with static and dynamic cyclotorsion correction using the Schwind Amaris [ J ] . C 2011,30 (9):951-957.
  • 8NarvOez J, Brucks M, Zimmerman G, et al. Intraoperative cyclo- rotation and pupil centroid shift during LASIK and PRK[ J]. J Cat- aract Refract Surg, 2012,28 ( 5 ) :353-357.
  • 9Neuhann IM, Lege BA, Bauer M,et al. Static and dynamic rota- tional eye tracking during LASIK treatment of myopic astigmatism with the Zyoptix laser platform and Advanced Control Eye Tracker [J:. J Cataract Refract Surg ,2010,26(1 ) : 17-27.
  • 10Bharti S, Bains HS . Active cyclotorsion error correction during LASIK for myopia and myopic astigmastim with the Nidek EC-5000 CX III laser[ J]. J Refract Surg, 2007,23 (9) : 1041-1045.

二级参考文献26

  • 1吕帆,王勤美,瞿佳.进一步重视屈光手术的安全性和有效性研究[J].中华眼科杂志,2005,41(6):482-485. 被引量:46
  • 2邱平,王铮,杨斌,黄国富,肖巴伊,顾建军,周胜.波阵面引导与常规准分子激光原位角膜磨镶术后对比敏感度和主观感觉的比较[J].中华眼科杂志,2007,43(4):329-335. 被引量:8
  • 3Bueeler M, Iseli HP, Jankov M, et al. Treatment-induced shifts of ocular reference axes used for measurement centration. J Cataract Refract Surg,2005,31 : 1986-1994.
  • 4Kohnen T, Btihren J, Ktihne C, et al. Wavefront-guided LASIK with the Zyoptix 3. 1 system for the correction of myopia and compound myopic astigmatism with 1-year follow-up: clinical outcome and change in higher order aberrations. Ophthalmology, 2004,111:2175-2185.
  • 5Kim A, Chuck RS. Wavefront-guided customized corneal ablation. Curr Opin Ophthalmol,2008 ,19 :314-320.
  • 6Chernyak DA. From wavefront device to laser: an alignment method for complete registration of the ablation to the cornea. J Refract Surg,2005,21:463-468.
  • 7Waring GO 3rd. Conventional standards for reporting results of refractive surgery. Refract Corneal Surg,1989,5:285-287.
  • 8Koch DD, Kohnen T, Obstbaum SA, et al. Format for reporting refractive surgical data. J Cataract Refract Surg,1998,24:285-287.
  • 9Hori-Komai Y, Sakai C, Toda I, et al. Detection of cyclotorsional rotation during exeimer laser ablation in LASIK. J Refract Surg, 2007,23:911-915.
  • 10Chang J. Cyclotorsion during laser in situ keratomileusis. J Cataract Refract Surg,2008,34 : 1720-1726.

共引文献5

同被引文献11

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部