期刊文献+

低剂量CT检查在头部外伤中的诊断价值 被引量:1

Clinical value of low-dose CT diagnosing head trauma
原文传递
导出
摘要 目的:探讨头部外伤诊断中应用低剂量CT检查的临床价值,为临床诊断提供参考。方法选择头部外伤患者48例,初次CT头部检查应用常规剂量进行,复查分别使用70%、50%、30%低剂量CT检查。对比不同剂量CT检查图像结果及射线剂量。结果随着扫描剂量降低,图像优良率大致呈降低趋势,但四种剂量扫描图像优良率相近,对比差异无统计学意义(P>0.05);常规剂量扫描时辐射水平达96.59 mGy,随着扫描剂量降低,辐射水平明显降低,低剂量扫描辐射水平与常规剂量对比差异均有统计学意义( t=4.293, P<0.05;t=7.547,P<0.01;t=9.438,P<0.01)。结论头部外伤应用CT检查时,低剂量CT检查结果可信,辐射剂量小,可优先考虑使用。 Objective To investigate the clinical value of low -dose CT diagnosing head trauma ,providing a reference for clinical diagnosis .Methods 48 patients with head trauma were selected ,routine CT dose was applied in the initial diagnosis ,then 70%,50%,30%( compared to routine CT dose ) CT dose were applied in review diagnosis . Image results and radiation dose were compared in different doses .Results With the scanning dose reduction ,excel-lent rate of image decreased roughly ,but rates of four doses were similar to each other ,there was not significant statis-tical difference (P〉0.05);during routine scanning,the radiation level was up to 96.59mGy,with scan dose reduc-tion,radiation levels were significantly lower ,compared with routine radiation level ,low-dose radiation level had sig-nificant statistical difference respecttively (t =4.293,P 〈0.05;t =7.547,P 〈0.01;t =9.438,P 〈0.01). Conclusion In CT diagnosis for head trauma ,low-dose CT checkups are credible with small dose radiation ,which can be a priority diagnosis option .
作者 陆重勋
出处 《中国基层医药》 CAS 2015年第5期646-647,共2页 Chinese Journal of Primary Medicine and Pharmacy
关键词 头部外伤 诊断 低剂量 体层摄影术 X线计算机 Head trauma Diagnosis Low dose Tonography,X-ray computed
  • 相关文献

参考文献9

二级参考文献45

共引文献86

同被引文献8

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部