低管电压、低碘浓度对比剂及迭代重建在冠状动脉CT血管成像中的应用 被引量:27

Application feasibility of low tube voltage and lower iodine concentration contrast medium as well as iterative reconstruction in coronary computed tomography angiography
原文传递
导出
摘要 目的 探讨行冠状动脉CT血管成像(CCTA)检查时,低管电压(100 kV)、低碘浓度(270 mgI/ml)对比剂及迭代重建在降低辐射剂量和碘对比剂用量的同时,保持图像质量的可行性.方法 连续入选2013年1至5月行CCTA检查的疑诊冠心病患者80例,将患者分为两组:对照组(40例)采用常规方案,即管电压采用120 kV,碘对比剂浓度为370 mgI/ml,采用滤波反投影法重建,重建卷积核为B26f;试验组(40例)管电压采用100kV,碘对比剂浓度为270 mgI/ml,采用迭代重建,重建卷积核为I26f.其他扫描和对比剂注射参数两组间完全一致.对两组的图像进行重建、测量和质量评分,并计算碘用量和有效辐射剂量.结果 对照组和试验组患者的体质指数分别为(25.3±3.0) kg/m2和(25.4±3.0) kg/m2(P=0.852),图像信噪比分别为22.2 ±5.5和23.6±5.8(P=0.277),对比度噪声比分别为35.6±17.6和41.1±17.6(P =0.163),图像质量分别为(4.70 ±0.52)分和(4.63±0.59)分(P =0.545).对照组和试验组碘用量分别为(25.27±0.94)g和(18.49±0.75)g(P <0.001),试验组较对照组少26.8%.对照组和试验组的有效辐射剂量分别为(3.52±1.16) mSv和(2.31±0.73) mSv(P <0.001),试验组较对照组少34.4%.结论 行CCTA检查时,采用低管电压、低碘浓度对比剂和迭代重建可以在保持冠状动脉图像质量不降低的基础上,显著减少碘对比剂用量及有效辐射剂量. Objective To investigate the feasibility of using lower iodine concentration 270 mgI/ml) contrast medium,lower X-ray tube voltage (100 kVp) and iterative reconstruction (IR) to reduce both iodine load and radiation dose but keep the image quality of coronary CT angiography (CCTA).Methods A total of 80 consecutive patients with suspected coronary artery disease were prospectively assigned to one of two groups via computer central system from January to May 2013.The control group (n =40) was scanned using dual-source CCTA protocols of 120 kV,370 mgI/ml Iopromide and filtered back projection reconstruction with a vascular algorithm (B26f).The study group (n =40) was scanned using 100 kV,270 mgI/ml Iodixanol and sinogram affirmed iterative reconstruction with a vascular algorithm (I26f).Other scan parameters and contrast injection protocol were similar between the two groups.Attenuation in the ascending aorta and coronary arteries along with image noise were measured.Images were reconstructed,measured and graded,and iodine load and effective radiation dose were calculated.Results The body mass index ((25.3 ± 3.0) kg/m2 vs.(25.4 ± 3.0) kg/m2,P =0.852),image quality scores (4.70 ±0.52 vs.4.63 ±0.59,P =0.545),mean signal-to-noise ratios (22.2 ±5.5 vs.23.6 ± 5.8,P =0.277),and contrast-to-noise ratios (35.6 ± 17.6 vs.41.1 ± 17.6,P =0.163) were similar between the control group and study group.Mean iodine loads were significantly reduced in the study group ((18.49 ± 0.75) g) compared to control group ((25.27 ± 0.94) g),P < 0.001).Mean effective radiation doses were also significantly reduced in the study group ((2.31 ± 0.73) mSv) compared to that in control group ((3.52 ± 1.16) mSv),P < 0.001).Conclusion Use of low X-ray tube voltage and iterative reconstruction allows lower iodine load and effective radiation dose application at CCTA without image quality reduction.
出处 《中华心血管病杂志》 CAS CSCD 北大核心 2015年第3期234-238,共5页 Chinese Journal of Cardiology
基金 国家“十一五”科技支撑计划(2007BAI05B02)
关键词 体层摄影术 X线计算机 造影剂 辐射剂量 Tomography, X-ray computed Contrast media Radiation dosage
  • 相关文献

参考文献17

  • 1Achenbach S, Ropers D, Kuettner A, et al. Contrast-enhanced coronary artery visualization by dual-source computed tomography - initial experience [J]. Eur J Radiol, 2006,57 (3) : 331-335.
  • 2Johnson TR, Nikolaou K, Wintersperger BJ, et al. Dual-source CT cardiac imaging: initial experience [J]. Eur Radiol, 2006,16 (7) : 1409-1415.
  • 3Scanlon PJ, Faxon DP, Audet AM, et al. ACC/ AHA guidelines for coronary angiography. A report of the American College of Cardiology/ American Heart Association Task Force on practice guidelines (Committee on Coronary Angiography). Developed in collaboration with the Society for Cardiac Angiography and Interventionsj L]. JAm Coll Cardiol, 1999,33(6) :1756-1824.
  • 4Gruberg L, Mintz GS, Mehran R, et al. The prognostic implications of further renal function deterioration within 48 h of interventional coronary procedures in patients with pre-existent chronic renal insufficiency [J]. J Am Coll Cardiol , 2000 ,36 ( 5 ) : 1542-1548.
  • 5Achenbach S, Moselewski F, Ropers D, et al. Detection of calcified and non calcified coronary atherosclerotic plaque by contrast-enhanced, submillimeter multidetector spiral computed tomography: a segment-based comparison with intravascular ultrasoundj J}. Circulation,2004,109(1) :14-17.
  • 6Shuman WP, Branch KR, May JM, et al. Whole-chest 64-MDCT of emergency department patients with nonspecific chest pain: Radiation dose and coronary artery image quality with prospective ECG triggering versus retrospective ECG gating [J]. AJR Am J Roentgenol,2009, 192 (6) : 1662-1667.
  • 7Austen WG, Edwards JE, Frye RL, et al. A reporting system on patients evaluated for coronary artery disease. Report of the Ad Hoc Committee for Grading of Coronary Artery Disease, Council on Cardiovascular Surgery, American Heart Association [J] . Circulation, 1975 ,51 (4 Suppl ) : 5 -40.
  • 8Moscariello A, Takx RA, Schoepf VJ, et al. Coronary CT angiography: image quality, diagnostic accuracy, and potential for radiation dose reduction using a novel iterative image reconstruction technique-comparison with traditional filtered back projection [J]. Eur Radiol ,2011,21 (10) :2130-2138.
  • 9Small GR, Kazmi M, Dekemp RA, et al. Established and emerging dose reduction methods in cardiac computed tomography [J]. J Nucl Cardiol,2011,18(4) :570-579.
  • 10Leipsic J, Labounty TM, Heilbron B, et al. Estimated radiation dose reduction using adaptive statistical iterative reconstruction in coronary CT angiography: the ERASIR study [J]. AJR Am J Roentgenol, 2010,195 (3) :655-660.

二级参考文献43

  • 1覃羽乔,徐永芳,梁桂民,陈娜莹,安爱萍,郭琳芳,陆建建,张获华,栾青,刘凤翔,汤洪洋,陈燕园.超重和肥胖与高血压、糖尿病的关系[J].中国公共卫生,2004,20(6):656-657. 被引量:59
  • 2杨晓光,孔灵芝,翟凤英,马冠生,金水高,中国居民营养与健康状况调查技术执行组.中国居民营养与健康状况调查的总体方案[J].中华流行病学杂志,2005,26(7):471-474. 被引量:174
  • 3杨晓光,李艳平,马冠生,胡小琪,王京钟,崔朝辉,王志宏,于文涛,杨正雄,翟凤英.中国2002年居民身高和体重水平及近10年变化趋势分析[J].中华流行病学杂志,2005,26(7):489-493. 被引量:81
  • 4Husmann L,Valenta I,Gaemperli O,et al.Feasibility of lowdose coronary CT angiography:first experience with prospective ECG-gating Eur Heart J,2008,29:191-197.
  • 5Coles DR,Smail MA,Negus IS,et al.Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography.J Am Coll Cardiol,2006,47:1840-1845.
  • 6Brenner DJ,Hall EJ.Computed tomography:an increasing source of radiation exposure.N Engl J Med,2007,357:2277-2284.
  • 7Hausleiter J,Meyer T,Hermann F,et al.Estimated radiation dose associated with cardiac CT angiography.JAMA,2009,301:500-507.
  • 8Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiation dose optimization.Radiology,2004,230:619-628.
  • 9Mollet NR,Cademartiri F,de Feyter PJ.Non-invasive multislice CT coronary imaging.Heart,2005,91:401-407.
  • 10Szucs-Farkas Z,Kurmann L,Strautz T,et al.Patient exposure and image quality of low-dose pulmonary computed tomography angiography:comparison of 100-and 80-kVp protocols.Invest Radiol,2008,43:871-876.

共引文献1798

同被引文献193

引证文献27

二级引证文献148

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部