基于脉搏波识别的电子血压计智能充气研究 被引量:5

Research of Electronic Sphygmomanometer Intelligent Aeration Based on Pulse Wave Identification
下载PDF
导出
摘要 通过对多款常用的国内外电子血压计的血压实验测试,发现对高血压或低血压情况进行测量时,存在着由于最大充气压力与本次血压测量不符,而带来反复充放气和充气压力过高的问题。为了解决这些问题,找到合适的最大充气压力,提出了两个基于脉搏波识别的智能充气方案,并进行了700组血压实验,对两个方案进行实验验证。实验证明所提出的两个方案具有很好的稳定性以及准确性,可以有效地解决血压计测量中的问题,同时得出了方案二即在充气过程中预估最大充气压力为最优方案。 Through various common domestic and foreign electronic sphygmomanometers to test blood pressure, we find that when measuring high blood pressure or low blood pressure, there is a mismatch between the maximum inflation pressure and the blood pressure measurement, which often results in repeatedly inflating and deflating as well as the problem of high inflation pressure. In order to solve these problems and find a suitable maximum inflation pressure, two intel igent pneumatic solutions based on identifying of pulse wave are suggested and 700 groups of blood pressure experiments are done, then the two solutions are verified by experiments. The experiment proved that these solutions proposed have good stability and accuracy, they can solve the problems appeared in measuring blood pressure effectively, at the same time, the second solution that estimate the maximum inflation pressure during inflation is considered as the best one.
出处 《中国医疗器械杂志》 CAS 2015年第2期90-94,共5页 Chinese Journal of Medical Instrumentation
基金 深圳市科创委项目(SW201110039 SDSY20120612094855904) 广东省科技厅项目(2012A032200025)
关键词 电子血压计 智能充气 预充气压力值 实验验证 electronic sphygmomanometer intel igent aeration maximal pressure value of aeration experimental verification
  • 相关文献

参考文献8

  • 1余守章,岳云,陈秉学.临床监测学[M].广东:科技出版社,1997:1-3.
  • 2Myers MG, Mclnnis NH, Fodor G J, et al. Comparison between an automated and manual sphygmomanometer in a population survey[J]. Am J Hypertens, 2008, 21(3): 280-283.
  • 3Mattu, Gurdial S, Perry, et al. Comparison of the oscillomet- ric blood pressure monitor(BPM-100 Beta)with the auscultatory mercury sphygmomanometer[J]. Blood Pressure Monit, 2001,6(3): 153-159.
  • 4钱峰,刘晔.基于示波法的电子血压计实现[J].仪器仪表学报,2006,27(z2):1534-1535. 被引量:21
  • 5何涛,吴夷,杜堃.医疗器械软件的要求和评价[J].中国医疗器械杂志,2011,35(5):379-382. 被引量:7
  • 6O'Brien E, Waeber B, Parati G, et al. Blood pressure measuring devices: recommendations of the European Society of Hypertension[J]. BMJ, 2001, 322: 531- 536.
  • 7李雪晴.基于示波法和高斯拟合腕式电子血压计设计与验证[D].中国科学技术大学,2011:47-54.
  • 8Wan Y, Heneghan C, Stevens R, et al. Determining which automatic digital blood pressure device performs adequately: a systematic review[J]. J Human Hypertens, 2010, 24(7): 431-438.

二级参考文献12

  • 1史学玲.功能安全标准的理论特点与管理模式[J].仪器仪表标准化与计量,2006(4):2-4. 被引量:7
  • 2何立民.MCS-51系列单片机应用系统设计[M].北京:北京航空航天大学出版社,1993..
  • 3齐颂扬.医学仪器[M].北京:高等教育出版社,1990..
  • 4YY/T0664-2008/IEC62304:2006,医疗器械软件软件生存周期过程[S].
  • 5GB/T16260.1-2006/ISO/IEC9126-1:2001,软件工程产品质量第1部分:质量模型[S].
  • 6GB/T20438.1-2006/IEC61508-1:1998,电气电子可编程电子安全相关系统的功能安全第1部分一般要求[S].
  • 7GB/T20438.4-2006/IEC61508-4:1998,电气/电子/可编程电子安全相关系统的功能安全第4部分:定义和缩略语[S].
  • 8YY/T0708-2009/IEC60601-1-4:2000,医用电气设备第1-4部分:安全通用要求并列标准:可编程医用电气系统[S].
  • 9IEC62366-2007,医疗设备可用性工程学对医疗设备的应用[S].
  • 10IEC60601-1-6,医用电气设备第1-6部分:基本安全和必要性能通用要求-并列标准:可用性[S].

共引文献26

同被引文献40

引证文献5

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部