期刊文献+

原子荧光和原子吸收法连续测定水中的重金属 被引量:2

Continuous Determination of Heavy Metals in Water by Atomic Fluorescence and Atomic Absorption Spectrometry
下载PDF
导出
摘要 通过剖析原子荧光和原子吸收法的不同消解体系,优化实验条件,确定了原子荧光和原子吸收法连续测定水中的硒、砷、锌、铜的含量的最佳消解条件和最佳仪器测定条件。经加标回收以及标样分析,连续测定的灵敏度高,回收率在95.9%~109%之间,相对标准误差低于4.41%,操作简便快速,结果精确。 By analyzing the different digestion system of atomic fluorescence and atomic absorption method and optimizing experimental conditions, the optimal digestion conditions and the best measurement conditions of instruments for continuous determination selenium, arsenic, zinc, copper content in water by atomic fluorescence and atomic absorption spectrometry were determined. Spiking recovery and standard analysis result showed that continuous measurement had high sensitivity, and the recoveries was between 95.9% and 109%, relative standard deviation was less than 4. 41%, the operation was simple, rapid and accurate.
作者 谢倩
出处 《广州化工》 CAS 2015年第7期111-113,共3页 GuangZhou Chemical Industry
关键词 原子荧光 原子吸收 连续测定 atomic fluorescence atomic absorption selenium arsenic zinc copper continuous determination