不同植入扭矩对种植体骨结合影响的临床初步探究 被引量:2

Study on the effects of different insertion torque values on implant osseointegration
下载PDF
导出
摘要 目的利用共振频率分析法观察不同的种植体植入扭矩值对不同时间点种植体稳定性以及骨结合效果的影响。方法根据纳入及排除标准选择牙列缺损成年患者49名,共植入86颗种植体,手术由同一高年资种植科医生操作完成。根据种植体植入就位后的所测最终扭矩值(Itv)进行分组:A低扭矩组,Itv为0~15 N·cm;B中扭矩组,Itv为16~35 N·cm;C高扭矩组,Itv为36~50 N·cm。利用共振频率分析仪分别对种植即刻、术后8周、术后12周的稳定指数(ISQ)进行测量分析,比较组内以及不同组间不同时间点种植体稳定性以及骨结合效果的差异。结果 1各扭矩组内比较,随着时间的延长,所有组ISQ均呈上升趋势,其中低、中扭矩组内各时间点间均有统计学差异,高扭矩组内各时间点间均无统计学差异。2各扭矩组间第8周ISQ值无明显差异,低扭矩组第8周ISQ值增加幅度最大;各扭矩组间第12周ISQ值无明显差异。结论种植体植入8周以内是形成良好的种植体骨结合的关键时期;不同的种植体植入扭矩值对于8周以后的骨结合效果影响差异不显著。 Objective To study the effect of different insertion torque values on implant stability and osseointegration by the means of Resonance Frequency Analysis. Methods 49 patients were chosen randomly according to the following standards,with 86 screws implanted. Furthermore,the same experienced clinician finished all surgeries. All implants were divided into three groups based on the insertion torque values: Group A,Itv 0 ~ 15 N·cm; Group B,Itv 20 ~ 40 N·cm; Group C,Itv 〉40 N·cm. The dental implant stability was measured to evaluate the osteointegration by Resonance frequency analysis when implants were placed immediately and at the time point of 8 weeks and 12 weeks after the surgeries. Results 1ISQ improved as the time extended in all groups. Statistical difference was shown in low and medium group while there' s no significant difference in high insertion torque value groups. 2 Low ISQ group showed the most obvious increase during 8 weeks; there was no significant difference among the three groups in 8 week and 12 week.Conclusions The 8-week time point is a critical time for implants healing to develop a high-quality osseointegration. What is more,there was no obvious difference of the osseointegration observed at the 8-week time point when ISQ changed among the groups.
出处 《口腔医学》 CAS 2015年第3期201-204,共4页 Stomatology
关键词 骨结合 稳定系数 植入扭矩 共振频率 牙种植体 osseointegration ISQ insert torque resonance frequency analysis implants
  • 相关文献

参考文献16

  • 1Nkenke E ,Fenner M. Indications for immediate loading of implants and implant success [ J ]. C lin Oral Implants Res, 2006,17 ( Suppl 2) :19 -34.
  • 2Irinaks T, Wiebe C. Initiahorque stability of a new bone condensing dental implant: a cohort study of 140 on seductively placed ira-plants[ J]. J Oral Implantol,2009,35 (6) :277 - 282.
  • 3余粤海,胡启斌,张勇平,邱小燕,谢小茹,管晓佩.高扭矩植入种植体的近期疗效的对比研究[J].中国口腔种植学杂志,2011,16(4):207-210. 被引量:7
  • 4Brnemark PI( September 1983). Osseointegratlon and its experi- mental background [ J ]. J Prosthet Dent, 1983,50 ( 3 ) :399 - 410.
  • 5Sennerby L, Meredith N. Resonance frequency analysis:measuring implant stability and osseointegration [J ]. Compcnd Contln Educ Dent,1998,19(5 ) :493 -498,500,502,quiz 504.
  • 6周炜,宋应亮.共振频率分析在口腔种植稳定性测量中的应用[J].国际生物医学工程杂志,2006,29(2):120-123. 被引量:7
  • 7SiC, strUm M,Sennerby L, Nilson H, et al. Reconstruction of the a- trophic edentulous maxilla with free iliac crest grafts and implants: a 3-year report of a prospective clinical study [ J ]. Clin Implant Dent Relat Res,2007,9(1) :46 -59.
  • 8Turkyilmaz I, Aksoy U, Mcglumphy EA. Two alternative surgical techniques for enh, ing primary implant stability in the posterior maxilla:a clinical study including bone density, inseItion torque, and resonance frequency analysis data[ J]. Clin Implant Dent Relat Res,2008,10(4) :231 -237.
  • 9Nedir R, Bischo FM, Szmukler-Moncler S, et al. Predicting osseoin- tegration by means of implant primary stability [ J ]. Clin Oral Im- plants Res ,2004,15 ( 5 ) :520.
  • 10庄秀妹,朱呈小露,邓飞龙,刘芹,廖展澎,吴亚霖,谢馨.Osstell ISQ^(TM)监测种植体稳定性的临床研究[J].中国口腔种植学杂志,2013,18(3):133-137. 被引量:21

二级参考文献119

  • 1李宁华,区品中,朱汉民,杨定焯,赵晰,高芳堃,郑宏,张大新,李恩,马翰章,邓力平,张韵,朱晓颖,安珍,刘力克,郑蘋如,李颖新.应用不同骨密度诊断标准计算骨质疏松症患病率比较研究[J].中国组织工程研究与临床康复,2001,10(24):114-115. 被引量:10
  • 2汪昆,李德华,于敬宇,李玉龙,刘宝林.双皮层骨种植影响牙种植体稳定性的有限元固有频率分析[J].实用口腔医学杂志,2005,21(3):356-358. 被引量:3
  • 3杨德圣,刘洪臣,顾晓明,魏务建,李华.牙种植体挤压植入后早期骨界面的组织学变化[J].中华老年口腔医学杂志,2005,3(2):113-114. 被引量:7
  • 4周炜,宋应亮.共振频率分析在口腔种植稳定性测量中的应用[J].国际生物医学工程杂志,2006,29(2):120-123. 被引量:7
  • 5Irinaks T,Wiebe C,Initial torque stability of a new bone condensing dental implant:a cohort study of 140 on seductively placed implants.J of Oral Implantology,2009;35:277-282.
  • 6Ueda M, Matsuki M, Jacobsson M, et al.Relationship between insertion torque and removal torque analyzed in fresh temporal bone.Int J Oral Maxillofac Implants.1991;6:442-447.
  • 7Drago C J,Lazzara RJ.Immediate occlusal loading of Osseotite implants in mandibular edentulous patients:a prospective observational report with 18-month data.J Prosthodont.2006;15:187-194.
  • 8Ottoni JM,Oliveira ZF,Mansini R,et al.Correlation between placement torque and survival of singletooth implants.Int J Oral Maxillofac Implants.2005;20:769-776.
  • 9Rabel A,Kohler SG,Schmidt-Westhausen AM.Clinical study on the primary stability of two dental implant systems with resonance frequency analysis.Clin Oral Investig.2007;11:257-265.
  • 10Olsson M,Friberg B,Nilson H,Kultje C.MkⅡ-a modified self-tapping Branemark implant:3-year results of a controlled prospective pilot study.Int J Oral Maxillofac Implants.1995;10:15-21.

共引文献61

同被引文献17

引证文献2

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部