期刊文献+

电感耦合等离子体原子发射光谱法测定焊剂中的磷元素 被引量:4

ICP-AES method determination flux of P element
下载PDF
导出
摘要 建立了焊剂中磷元素的电感耦合等离子体原子发射光谱(ICP-AES)的测定方法,样品经盐酸、硝酸溶解,于优化选定的仪器条件下测定磷元素.使用本方法与熔炼焊剂化学分析方法-钼蓝光度法测定磷量(JB/T7948.8-1999)结果一致.测定回收率均在96.0%~98.0%之间,标准偏差小于0.8%,该法操作简便,分析快速,结果准确,比钼蓝光度法更为快速。 To establish the determination method of flux in P using ICP-AES. Samples dissolved hydrochloric HC1-HNO3 dissolve. To optimize the selected instrument under the condition of P element. Using this method and the heat flux of chemical analysis molybdenum blue photometric method for determination of P amount (JB/T7948. 8-1999) consistent results. In the determination of recovery 96.0%~98.0%. The standard deflation is less than 0. 8%.The method is simple, rapid analysis theresults are accurate.More quickly than molybdenum blue photometeie method.
作者 刘明 田国峰
出处 《天津化工》 CAS 2015年第2期47-48,共2页 Tianjin Chemical Industry
关键词 焊剂 ICP-AES flux P
  • 相关文献

参考文献3

  • 1中华人民共和国机械行业标准JB/T7948.8-1999熔炼焊剂化学分析方法,钼蓝光度法测定磷量.
  • 2Ebert K H.Analysis of Portland Cement by ICP-AES[J].Atomic tros- copy,1995(3): 102-103.
  • 3Rao R.N.Ta Ⅱ uri M.V.N.K.An overview of recent applications of in- ductively coupled plasma mass spectrometry [ICP-AES] in deter- mination of inorganic impurities in drugs and pharmaceuticals[J]. Journal of Pharmaceutical and Biomedical Analysis,2007.43(1): 1-13.