热塑膜固定联合π形尺摆位在腹部肿瘤患者放疗中的应用 被引量:2

Application of Combination of Thermoplastic Film Fixation andπ-Shaped Ruler Positioning in Radiotherapy for Abdominal Tumors
下载PDF
导出
摘要 目的:探讨热塑膜固定联合π形尺摆位在腹部肿瘤患者放疗中的临床应用价值。方法选取我院2013年12月-2014年3月12例腹部肿瘤患者,均在模拟定位机下做膜和标记魔十字,并在动度较小的骨性标记处的皮肤上标记π形尺激光点,然后摄取其正侧位DR图像;复位时A组使用热塑膜固定摆位,B组使用热塑膜固定联合π形尺摆位,分别摄取两组正侧位DR图像;将复位与做膜时摄取的正侧位DR图像进行二维配准,测量摆位误差。结果B组与A组在X、Y、Z轴方向上的摆位误差均值分别为3.40、4.59、6.92 mm和6.63、7.35、8.15 mm。B组摆位误差小于A组,两组差异有统计学意义(P<0.05)。结论热塑膜固定联合π形尺摆位比单独使用热塑膜固定摆位更准确,可有效减小摆位误差。 Objective To investigate the effectiveness of application of combination of thermoplastic iflm ifxation andπ-shaped ruler positioning in radiotherapy for patients with abdominal tumors. Methods Altogether 12 cases with good self-control ability were selected among abdominal tumor patients who had been treated in the hospital from December 2013 to March 2014. For all the patients, the iflm was made, the magic cross was marked, the skin at the bony landmarks with the smaller dynamic degree was marked with aπ-shaped laser point when they lay on the simulation localization machine. Then, the anteroposterior and lateral DR (Digital Radiography) images were taken. During the restoration, two positioning methods were adopted:the direct positioning of thermoplastic iflm ifxation was adopted in Group A;Group B used the combination of thermoplastic iflm ifxation andπ-shaped ruler positioning. The anteroposterior and lateral DR images were taken. The two-dimensional image matching was performed between images taken during restoration and iflm-making to measure the error. Results The absolute value of errors (mm) between the positioning in X, Y and Z directions of Group B and Group A were respectively 3.40, 4.59, 6.92 mm and 6.63, 7.35, 8.15 mm. The errors significantly decreased when the combination of the thermoplastic film fixation and π-shaped ruler positioning was adopted. The difference of the errors between two positioning methods was statistically significant (P<0.05). Conclusion Application of combination of the thermoplastic iflm ifxation andπ-shaped ruler positioning demonstrated its superior accuracy over the single use of the thermoplastic film fixation, which can effectively reduce errors caused by the positioning.
出处 《中国医疗设备》 2015年第4期96-97,93,共3页 China Medical Devices
基金 四川省卫生厅科研课题(100158) 川北医学院重点培育项目(CBY11-A-ZP04)
关键词 放射治疗 热塑膜固定 π形尺 魔十字标记 摆位误差 radiotherapy thermoplastic iflm ifxation π-shaped ruler magic cross marker positioning errors
  • 相关文献

参考文献10

二级参考文献33

  • 1何晓华,黄继列,陈立新,黄劭敏,梁俊斌,陈国章.头颈肿瘤立体定向分次治疗体位固定技术[J].中国肿瘤,2004,13(9):558-560. 被引量:15
  • 2邱梅.腹部肿瘤放疗使用固定器价值的探讨[J].现代肿瘤医学,2004,12(6):580-581. 被引量:15
  • 3王鑫,胡超苏,应红梅,朱国培,何少琴.鼻咽癌适形放射治疗中的摆位误差分析[J].中国癌症杂志,2006,16(4):272-276. 被引量:17
  • 4刘丽英.腹部肿瘤适形放疗的标记探讨[J].医疗设备信息,2006,21(7):86-87. 被引量:1
  • 5International Organization for Standardization."Quality Management and Quality Assurance Standards-Part 1:Guideliness for Selection and Use."ISO 900O,1994.
  • 6Kutcher,G.J.,L.Coia,M.Gillin,W.F.Hanson,S.Leibel,R.J.Morton,J.R.Palta,J.A.Purdy,L.E.Reinstein,G.K.Svensson,M.Weller,L.Wingfield,"Comprehensive QA for radiation oncology."Report of AAPM Radiation Therapy Committee Task Group 40."Med.Phys.21:581-618.(1994).
  • 7Purdy,J.A.,and C.A.Perez,"Quality assurance in radiation oncology in the United States."Rays 21:505-540(1996).
  • 8刘凌,管峦,孙凌飞,等.应用EPID提高放疗摆位精确性的临床研究[C].2006年国际医学物理学术会议暨全国放射肿瘤物理学学术年会.2006:285-288.
  • 9Fidanzio A,Mameli A,Placidi E,et al.Epid cine acquisition mode for in vivo dosimetry in dynamic are radiation therapy[J].Nuclear Instruments &Methods in Physics Research.B,Beam Interactions with Materials and Atoms,2008,266(4):658-666.
  • 10M.Mohammadi,E.Bezak.Evaluation of dosimetric characteristics of multi-leaf and conventional collimated radiation fields using a scanning liquid ionization chamber EPID[J].Australasian Physical &Engineering Sciences in Medicine,2008,31(4):280-289.

共引文献52

同被引文献14

引证文献2

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部