摘要 应用AA3型连续流动分析仪测定地表水中挥发酚,增加长光程比色池(LWCC)以降低方法检出限,达到更低的检测要求。经实验论证,在0-0.0500 mg/L范围内,线性良好(r=0.9999),检出限为0.0002 mg/L,满足地表水检测要求。相较手工分析方法大大提高分析效率,并且减轻对实验人员的危害。 A Continuous Flow Analysis( CFA) method for determination of volatile phenol in water was developed.For lower detection limit,Liquid Waveguide Capillary Cell( LWCC) was added. Statistics showed that in the range of 0 -0. 0500 mg / L,detection limit could be lowered to 0. 0002 mg / L,with good correlation coefficient( r = 0. 9999) at the same time. This method could be used in the field of environmental monitoring. Compared to manual method,CFA was more efficient and less hazardous.
作者 高裕雯
出处 《广州化工》 CAS 2015年第8期144-145,共2页 GuangZhou Chemical Industry
关键词 长光程比色池 连续流动 挥发酚 检出限 LWCC Continuous Flow Analysis volatile phenol detection limit
  • 相关文献

参考文献4

  • 1中华人民共和国卫生部国家标准化管理委员会.GB5749-2006生活饮用水卫生标准[S].北京:中华人民共和国卫生部,2006.
  • 2中华人民共和国环境保护部.HJ503-20199水质挥发酚的测定4-氨基安替比林分光光度法[S].北京:中华人民共和国环境保护部,2009.
  • 3刘明,魏磊,周天泽.长光程比色池简介[J].分析仪器,1983(3):86-88.
  • 4丁宇,周睿,劳宝法.流动注射分光光度法测定饮用水中挥发酚的方法研究[J].上海预防医学,2010,22(7):349-351. 被引量:18