摘要 目的比较暗室下采用传统瞳孔直径测量工具和全自动电脑验光仪瞳孔测量结果。方法分别用Haab瞳孔尺和Nidek ARK-530A全自动电脑验光仪测量于本院行准分子激光手术患者31例(62眼)的双眼暗视瞳孔水平直径,对结果进行统计学分析。结果传统瞳孔直径测量工具Haab瞳孔尺测得的暗视平均瞳孔直径右眼为(5.4±0.2)mm,左眼(5.4±0.3)mm。Nidek ARK-530A全自动电脑验光仪测得的暗视瞳孔直径右眼为(6.5±0.4)mm,左眼(6.5±0.2)mm。两组比较差异有统计学意义(P<0.01)。结论两种方法测得的暗室瞳孔直径不同,Nidek ARK-530A全自动电脑验光仪测量瞳孔直径准确性更高,更接近文献记载的暗视瞳孔直径。
出处 《广东医学院学报》 2014年第6期846-847,共2页 Journal of Guangdong Medical College
  • 相关文献

参考文献6

  • 1Holladay J T, Janes J A. Toppgraphic changes in corneal asphericity and effective optical zone after laser in situ keratomileusis [J]. Cataract Refract Surg 2002, 28(6):942-947.
  • 2Villa C, Gutierrez R, Jimenez J R, et al. Night vision disturbances after successful lasik surgery[J]. BR J Ophthalmol, 2007, 91(8): 1031-1037.
  • 3宋慧,汤欣,袁佳琴.远红外瞳孔测量仪对屈光手术患者适应度的评估[J].中国实用眼科杂志,2005,23(9):941-943. 被引量:6
  • 4Winn B, Whitaker D, Ellintt D B, et al. Facter affecting light-a-dapted pupil size in nomal human subjects[J]. Invest Ophthalmol Via Sci, 1993, 95(3):1132-1136.
  • 5Kohnen T, Terzi E, Buhren J, et al. Comparison of a digital and ahandheld infrared pupillometer for determining acotopic pupil diameter [J]. J Cataract Refract Surg, 2003, 29(1): 112-117.
  • 6Nixon W S. Pupil size in refractive surgery[J]. J Cataract Refract Surg, 1997, 23(6):1435-1436.

二级参考文献9

  • 1O' Brart DPS, Corbett MC, Lohmann CP, et al. The effects of ablation diameter on the outcome of excimer laser photorefractive keratectomy; a prospective, randomized, double - blind study.Arch Ophthalmol, 1995, 113:438-443
  • 2Roberts CW, Koester CJ. Optical zone diameters for photorefractive corneal surgery. Invest Ophthalmol Vis Sci, 1993, 34:2275- 2281
  • 3Michael C. Preoperative measurement of scotopic pupil dilation using an office pupillometer. J Cataract Refract Srug, 1998, 24:1594 - 1597
  • 4Thomas K, et al. Comparison of a digital and a handheld infrared pupillometer for determining scotopic pupil diameter. J Cataract Refract Surg, 2003, 29:112-117
  • 5Ludtke H, Wilhelm B, Adler M, et al. Mathematics procedures in data recording and processing of papillary fatigue waves. Vison Res, 1998, 38:2889-2896
  • 6Maguire LJ, Bechara S. Epithelial distortiions at the ablation zone margin after excimer laser photorefractive keratectomy for myopia.AmJ Ophthalmol, 1994, 117:809-810
  • 7Nixon WS. Pupil size in refractive surgery. J Cataract Refract Surg, 1997, 23:1435-1436
  • 8Emanuel SR, Christa LG, et al. Use of a digital infrared pupillometer to assess patient suitability for refractive surgery. J Cataract Refract Surg, 2002, 28: 1433- 1438
  • 9Eva- Maria S, Martin B, et al. Scotopic measurement of normal pupils: Colvard versus Video Vision Analyzer infrared pupillometer.J Cataract Refract Surg, 2000, 26:859 - 866

共引文献5

同被引文献2

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部