Sysmex CS5100全自动凝血分析仪性能评价 被引量:4

Performance evaluation of Sysmex CS5100 automatic blood coagulation analyzer
下载PDF
导出
摘要 目的验证Sysmex CS5100全自动凝血分析仪的系统性能,确定该检测系统是否稳定、准确、可靠。方法对Sysmex CS5100全自动凝血分析仪上的准确度、不精密度、线性、检测限、携带污染率进行评价。测试20例健康体检者的凝血酶原时间(PT)、活化部分凝血活酶时间(APTT)、凝血酶时间(TT)、纤维蛋白原(FIB)验证厂家提供的参考区间。结果准确度试验测定结果全部合格;批内最大变异系数(CV)3.82%、批间最大CV值为6.24%;线性试验FIB相关系数(r)为0.998 9、D-二聚体(DD)r为0.996 7、纤维蛋白(原)降解产物(FDP)r为0.998 5;检测限最大CV为6.13%;携带污染率最大值为-6.45%,验证全部合格。结论 Sysmex CS5100全自动凝血分析仪是一个精密度好、准确度高、检测范围宽、携带污染率低的检测系统。 Objective To verify the systematic performance of the Sysmex CS5100 automatic coagulation analyzer for determining whether this detection system is stable, accurate, and reliable. Methods The accuracy, precision,linearity, limit of detection and the carrying-over contamination rate of the Sysmex CS5100 automatic coagulation analyzer were evaluated. Prothrombin time (PT), activated partial thromboplastin time (APTT) ,thrombin time (TT) and fibrinogen (FIB) were detected by this analyzer in 20 cases of healthy physical examination were examined to verify the reference intervals provided by the manufacturer. Results The accura- cy examination results were all qualified;the maximum intra-assay coefficient of variation(CV) was 3.82 G, and maximum inter-day CV was 6.24 ~;the linear correlation coefficient (r) of FIB was 0. 998 9,which of D-dimer (D-D) was 0. 996 7 and which of fibrin (ogen) degradation products (FDP) was 0. 998 5the maximum CV of detection limit was 6.13 % ;carrying-over contamination rate was --6.45%. The verification results were all qualified. Conclusion The Sysmex CS5100 automatic blood coagulation analyzer is a detection system with good precision, high accuracy, wide detection range and low carrying-over contamination rate.
出处 《国际检验医学杂志》 CAS 2015年第9期1259-1261,共3页 International Journal of Laboratory Medicine
关键词 SYSMEX CS5100全自动凝血分析仪 性能 评价 Sysmex CS5100automatic blood coagulation ianalyzerl performance evaluation
  • 相关文献

参考文献11

二级参考文献46

  • 1张晓虹.CA510全自动血凝仪的使用与维护[J].医疗设备信息,2005,20(6):73-74. 被引量:11
  • 2陆红兵,孙文旦,王妍,郁寿仙.STA-Compact型全自动血凝仪的性能评价[J].血栓与止血学,2005,11(4):174-175. 被引量:6
  • 3毕波,吕元.定量检测方法学性能验证的系统设计[J].中华检验医学杂志,2007,30(2):143-145. 被引量:153
  • 4唐宁,曹文静,夏汛生,张碧玉.ACL TOP全自动血凝分析系统的性能评价[J].医疗设备信息,2007,22(2):48-51. 被引量:17
  • 5中国合格评定国家认可委员会.CNAS-CL02医学实验室质量和能力认可准则[S].北京:中国计量出版社,2006:6.
  • 6National Committee for Clinical Laboratory standards.User Demonstration of Performance for Precision andAccuracy; Approved Guideline[S].Wayne,Pennsylvania.NCCLS,2001.
  • 7National Committee for Clinical Laboratory standards.User Demonstration of Performance for Precision and Accuracy;Approved Guideline[S].Wayne,Pennsylvania:NCCLS,2001.
  • 8Clinical and Laboratory Standards Institute.H57-A Protocol for the evaluation,validation,and implementation of coagulometers;approved guideline[S].Wayne,PA:CLSI,2008.
  • 9Clinical and Laboratory Standards Institute.P7-A2 Interference testing in clinical chemistry; approved guideline-second edition[S].Wayne,PA:CLSI,2005.
  • 10Clinical and Laboratory Standards Institute.AUTO 8-A Managing and validating laboratory information systems; approved guideline[S].Wayne,PA:CLSI,2006.

共引文献165

同被引文献22

引证文献4

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部