摘要 A fifth order operational transconductance amplifier-C (OTA-C) Butterworth type low-pass filter with highly linear range and less passband attenuation is presented for wearable bio-telemetry monitoring applications in a UWB wireless body area network. The source degeneration structure applied in typical small transconduc- tance circuit is improved to provide a highly linear range for the OTA-C filter. Moreover, to reduce the passband attenuation of the filter, a cascode structure is employed as the output stage of the OTA. The OTA-based circuit is operated in weak inversion due to strict power limitation in the biomedical chip. The filter is fabricated in a SMIC 0.18-μm CMOS process. The measured results for the filter have shown a passband gain of -6.2 dB, while the -3-dB frequency is around 276 Hz. For the 0.8 Vpp sinusoidal input at 100 Hz, a total harmonic distortion (THD) of-56.8 dB is obtained. An electrocardiogram signal with noise interference is fed into this chip to validate the function of the designed filter. A fifth order operational transconductance amplifier-C (OTA-C) Butterworth type low-pass filter with highly linear range and less passband attenuation is presented for wearable bio-telemetry monitoring applications in a UWB wireless body area network. The source degeneration structure applied in typical small transconduc- tance circuit is improved to provide a highly linear range for the OTA-C filter. Moreover, to reduce the passband attenuation of the filter, a cascode structure is employed as the output stage of the OTA. The OTA-based circuit is operated in weak inversion due to strict power limitation in the biomedical chip. The filter is fabricated in a SMIC 0.18-μm CMOS process. The measured results for the filter have shown a passband gain of -6.2 dB, while the -3-dB frequency is around 276 Hz. For the 0.8 Vpp sinusoidal input at 100 Hz, a total harmonic distortion (THD) of-56.8 dB is obtained. An electrocardiogram signal with noise interference is fed into this chip to validate the function of the designed filter.
出处 《Journal of Semiconductors》 EI CAS CSCD 2015年第5期120-125,共6页 半导体学报(英文版)
基金 Project supported by the National Natural Science Foundation of China(Nos.61161003,61264001,61166004) the Guangxi Natural Science Foundation(No.2013GXNSFAA019333)
关键词 ECG signal fully differential OTA-C filter highly linear range less passband attenuation ECG signal fully differential OTA-C filter highly linear range less passband attenuation
  • 相关文献

参考文献14

  • 1Webster J G.Medical instrumentation:application and design.New York:Wiley,2009.
  • 2Lasanen K,Kostamovaara J.A I-V analog CMOS front-end for detecting QRS complexes in a cardiac signal.IEEE Trans Cir- cuits Syst I:Regular Papers,2005,52(12):2584.
  • 3Chen Fangxiong,Lin Min,Ma Heping,et al.An asymmetric MOSFET-C band-pass filter with on-chip charge pump auto- tuning.Journal of Semiconductors,2009,30(8):085005.
  • 4Wan Chuanchuan,Li Zhiqun,Hou Ningbing.A CMOS Gm-C complex filter with on-chip automatic tuning for wireless sen- sor network application.Journal of Semiconductors,2011,32(5):055002.
  • 5Lee S Y,Cheng C J.Systematic design and modeling of a OTA- C filter for portable ECG detection.IEEE Trans Biomedical Cir- cuits Syst,2009,3(1):53.
  • 6Solis-Bustos S,Silva-Martinez J,Maloberti F,et al.A 60-dB dynamic-range CMOS sixth-order 2.4-Hz low-pass filter for medical applications.IEEE Trans Circuits Syst II,2000,47(12):1391.
  • 7Veeravalli A,Sdnchez-Sinencio E,Silva-Martinez J.Transcon- ductance amplifier structures with very small transconductances:a comparative design approach.IEEE J Solid-State Circuits,2002,37(6):770.
  • 8Lasanen K,Kostamovaara J.A I-V analog CMOS front-end for detecting QRS complexes in a cardiac signal.IEEE Trans Cir- cuits Syst I:Regular Papers,2005,52(12):2584.
  • 9Rodriguez-Villegas E,Yufera A,Rueda A.A 1.25-V micropower Gm-C filter based on FGMOS transistors operating in weak in- version.IEEE J Solid-State Circuits,2004,39(1):100.
  • 10Mahmoud S A,Bamakhramah A,Al-Tunaiji S A.Low-noise low-pass filter for ECG portable detection systems with digitally programmable range.Circuits Systems and Signal Processing,2013,32(5):2029.

同被引文献2

引证文献1

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部