HPLC法测定五层共挤输液用袋中抗氧剂1178的含量 被引量:3

Determination of Antioxygen 1178 in 5-Layer Co-extrusion Bags Used for Infusion by HPLC
下载PDF
导出
摘要 目的:建立五层共挤输液用袋中抗氧剂1178的含量测定方法。方法:色谱柱:Uitimat XB-C8(250 mm ×4.6 mm,5μm),流动相:甲醇-水梯度洗脱,流速:1.0 ml·min-1,检测波长:223 nm,柱温:35℃,进样量20μl。结果:抗氧剂1178在1.64~205.10μg·ml-1范围内线性关系良好(r=0.9999),平均回收率为92.05%,RSD=1.94%(n=9)。结论:建立的方法结果准确、稳定性好、专属性强,可用于五层共挤输液用袋中抗氧剂1178的含量测定。 Objective:To establish a method for the determination of antioxygen 1178 in 5-layer co-extrusion bags used for infu-sion. Methods:A Uitimat XB-C8 (250 mm × 4. 6 mm,5μm) column was used,the mobile phase was methanol-water with gradient elu-tion, the flow rate was 1. 0 ml·min-1 ,the detection wavelength was 223 nm,the column temperature was 35℃ and the injection vol-ume was 20 μl. Results:The concentration of antioxygen 1178 had the linear relationship within the range of 1. 64-205. 10 μg·ml-1 (r=0. 999 9),and the average recovery was 92. 05% (RSD=1. 94%, n=9). Conclusion:The method is accurate,stable and spe-cific,and can be used to determine antioxygen 1178 in 5-layer co-extrusion bags used for infusion.
出处 《中国药师》 CAS 2015年第5期876-877,共2页 China Pharmacist
关键词 抗氧剂1178 高效液相色谱 五层共挤输液用袋 Antioxygen 1178 HPLC 5-Layer co-extrusion bags used for infusion