红外单光子探测器定标方法研究 被引量:1

CALIBRATION METHOD STUDY FOR INFRARED SINGLE-PHOTON DETECTORS
下载PDF
导出
摘要 紫外激光器输出的355 nm激光泵浦BBO晶体,采用晶体的I类相位匹配,利用晶体的角度调谐特性得到的红外波段光源应用于单光子探测器定标,提出了利用参量下转换产生的纠缠光子对定标SPCM单光子探测器量子效率的方法,介绍了实验原理和定标装置。 355 nm UV laser output laser pumping BBO crystal, using I phase matching crystal, infrared light source utilizing crystal angle tuning characteristics are applied to the single-photon detector calibration. We propose the use of parametric down conversion produce entangled photons calibrate SPCM quantum efficiency of the method. We also introduce the principle and experiment of calibration device.
出处 《井冈山大学学报(自然科学版)》 2015年第2期11-14,24,共5页 Journal of Jinggangshan University (Natural Science)
基金 辽宁省激光与光信息技术重点实验室开放基金项目(4771004kfs20)
关键词 纠缠光子 角度调谐 I类相位匹配 探测器定标 entangled photons angle tuning I phase matching detector calibration
  • 相关文献

参考文献9

  • 1Polyakov S V, Migdall A L. High accuracy verification of a correlated-photon-based method for determining photoncounting detection efficiency[J]. Optics express, 2007, 15(4): 1390-1407.
  • 2Klyshko D N. Use of two-photon light for absolute calibration of photoelectric detectors[J]. Quantum Electronics, 1980, 10: 1112-1117.
  • 3Brida G, Castelletto S, Novero C, et al. Quantum-efficiency measurement "of photodetectors by means of correlated photons[J]. JOSA B, 1999, 16(10): 1623-1627.
  • 4Lan Xinju. Lase Technology [M].Wuhan:HuaZhong University of Science and Technology Press,1995: 199-228(in Chinese).
  • 5Castelletto S,Degiovanni I P, Schettini V, et al.Optimizing single-Photon-source heralding efficiency and detection efficiency metrology an 1550nm using periodically poled lithium biobate[J]Metrologia,2006,43 (2):S56-S60.
  • 6Odate S, Yoshizawa A, Fukuda D, et al. Quantum efficiency measurements by bidirectional coincidence counting of correlated photon pairs[J]. Optics letters,2007, 32(21): 3176-3178.
  • 7Brida G, Genovese M, Monticone E, et al. Experimental sub-shot noise quantum imaging versus differential classical imaging[C].0uantum, Nano and Micro Technologies, 2010. ICQNM'I0. Fourth International Conference on. IEEE, 2010: 71-76.
  • 8Avclla A, Bfida Q Degiovanni I P, ct al. Self consistent, absolute calibration technique for photon nurnbcr resolving detectors[J]. Optics express, 2011, 19(23): 23249-23257.
  • 9Peifina Jr J, Haderka O, Allevi A, et al. Absolute calibration of photon-number-resolving detectors with an analog output using twin beams[J]. Applied Physics Letters, 2014, 104(4): 041113.