期刊文献+

磁导航指导下心房颤动导管消融与手动消融辐射剂量的对比研究 被引量:2

Clinical study of radiation exposure from atrial fibrillation catheter ablation guided by magnetic navigation system
原文传递
导出
摘要 目的 比较磁导航指导下心房颤动导管消融与手动消融时辐射剂量的差异.方法 连续收住入院的94例行房颤(AF)导管消融术患者,前60例为手动消融组(CON组),后34例为磁导航(MNS)指导下导管消融组(MNS组).对比两组患者的皮肤表面累积入射剂量(CD)、剂量面积乘积(DAP)、透视时间,以及医护人员的辐射剂量和透视时间.结果 MNS组和CON组两组患者的CD值分别为(0.54±0.45)和(1.61 ±0.89)Gy(t=2.44,P<0.05),DAP值为(46.86±27.09)和(139.71±76.69) Gy· cm^2(t=3.89,P<0.05),透视时间为(15.60 ±7.52)和(39.50±8.82) min (t=1.96,P<0.05).两组手术医师辐射剂量分别为(22.68±6.87)和(62.74±20.92)μSv(t=2.02,P<0.05),透视时间为(11.48±7.59)和(30.50±14.82)min(t =2.75,P<0.05),助手辐射剂量为(19.38±5.64)和(49.42±10.67)μSv(t =3.58,P<0.05)、透视时间为(8.96±5.88)和(24.49±9.09) min(=4.20,P<0.05),护士辐射剂量为(18.98±4.99)和(47.77±13.65) μSv(t =3.17,P<0.05)、透视时间为(8.33±6.35)和(22.99±13.36) min(t=2.76,P<0.05).结论 与手动消融相比,磁导航指导下心房颤动导管消融具有明显减少医患辐射剂量的优点. Objective To compare the X-ray radiation dose from atrial fibrillation (AF) ablation guided by magnetic navigation system (MNS) and manual procedure (CON).Methods 94 AF patients were randomly divided into MNS group (34 cases) and CON group(60 cases).The dose area product (DAP),cumulative radiation dose (CD),fluoroscopy time of patients,the X-ray exposure time and dose of operators were recorded and analyzed.Results The results from the patients in MNS group and CON group were CD values (0.54 ±0.45) and (1.61 ±0.89) Gy (t =2.44,P 〈0.05),DAP values (46.86 ±27.09) and (139.71 ±76.69) Gy·cm^2(t =3.89,P 〈0.05) and fluoroscopy time (15.60 ±7.52) and (39.50 ± 8.82) min (t =1.96,P 〈 0.05),respectively.The X-ray exposure dose in both groups were (22.68 ± 6.87) and (62.74 ± 20.92) μSv (t =2.02,P 〈 0.05) for operation doctor (19.38 ± 5.64) and (49.42 ± 10.67) μ Sv (t =3.58,P 〈 0.05) for operation assistant and (18.98 ± 4.99) and (47.77 ± 13.65) μ Sv (t =3.17,P 〈 0.05) for nurse,respectively.The X-ray exposure time experienced in both groups were (11.48 ±7.59) and (30.50 ±14.82) min (t =2.75,P 〈0.05) for operation doctor,(8.96 ±5.88) and (24.49 ±9.09) min (t =4.20,P 〈0.05) for operation assistant and (8.33 ±6.35) and (22.99 ± 13.36) min(t =2.76,P 〈 0.05) for nurse,respectively.Conclusions Compared with manual procedure,the MNS applied in AF ablation has the potential to decrease X-ray exposure dose.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2015年第4期307-311,共5页 Chinese Journal of Radiological Medicine and Protection
基金 江苏省无锡市科技发展资金资助项目(CSE32N1302) 无锡市医院管理中心医学科技发展基金面上项目(YGM1110)
关键词 磁导航 心房颤动 心血管介入 X射线 Magnetic navigation system Atrial fibrillation Cardiovascular intervention X-rays
  • 相关文献

参考文献11

  • 1Chida K, Saito H, Otani H, et al. Relationship between fluoroscopic time, dsoe-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures [J]. Am J Roentgenol, 2006, 186(3) : 774-778.
  • 2Smith IR, Rivers JT, Hayes J, et al. Reassessment of radiation risks from electrophysiology procedures compared to coronary angiography[ J]. Heart Lung Circ, 2009,18(3) :191-199.
  • 3January CT, Wann LS, Alpert JS, et al. 2014 AHA/ACC/HRS guideline for the management of patients with atrial fibrillation: executive summary: A report of the American College of Cardiology/American Heart Association task force on practice guidelines and the Heart Rhythm Society[J] Circulation, 2014- 04-10. http://www, ncbi. nlm. nih. gov/pubmed/24682348.
  • 4Miyazaki S, Shah A J, Xhaet O, et al. Remote magnetic navigation with irrigated tip catheter for ablation of paroxysmal atrial fibrillation [ J]. Circ Arrhythm Electrophysiol, 2010, 3 (6) :585-589.
  • 5Proietti R, Pecoraro V, Di Blase L, et al. Remote magnetic with open-irrigated catheter vs. manual navigation for ablation of atrial fibrillation: A systematic review and meta-analysis [ J 1. Eurupaee, 2013, 15(9) :1241-1248.
  • 6Bauernfeind T, Akea F, Schwagten B, et al. The magnetic navigation system allows safety and high efficacy for ablation of arrhythmias[J]. Europaee, 2011, 13(7) :1015-1021.
  • 7Arya A1, Zaker-Shahrak R, Sommer P, et al. Catheter ablation of atrial fibrillation using remote magnetic catheter navigation : acase-control study[J]. Europace, 2011, 13( 1 ) :45-50.
  • 8Pezawas T, Ristl R, Bilinski M, et al. Single, remote-magnetic catheter approach for pulmonary vein isolation in patients with paroxysmal and non-paroxysmal atrial fibrillation [ J ]. Int JCardiol, 2014, 174(1):18-24.
  • 9冯俊,王爱玲,程景林,郭杰.不同类型心血管介入手术辐射剂量分析[J].中华放射医学与防护杂志,2012,32(4):416-419. 被引量:17
  • 10Shurrab M, Danon A, Lashevsky l, et al. Robotically assisted ablation of atrial fibrillation: a systematic review and meta- analysis[ J]. Int J Cardiol, 2013, 169 (3) : 157-165.

二级参考文献9

  • 1Chida K, Saito H, Otani H, et al. Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. AJR Am J Roentgenol, 2006,186(3):774-778.
  • 2Chida K, Kato M, Kagaya Y, et al. Radiation dose and radiation protection for patients and physicians during interventional procedure. J Radiat Res, 2010,51(2): 97-105.
  • 3Valentin J. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP,2000,30(2):7-67.
  • 4Miller DL, Balter S, Cole PE, et al. Radiation doses in interventional radiology procedures: the RAD-IR study. PartⅠ: overall measures of dose. J Vasc Interv Radiol, 2003,14(6):711-727.
  • 5Miller DL, Balter S, Cole PE, et al. Radiation doses in interventional radiology procedures: the RAD-IR study. PartⅡ: skin dose. J Vasc Interv Radiol, 2003, 14(8): 977-990.
  • 6Mercuri M, Xie CC, Levy M, et al. Predictors of increased radiation dose during percutaneous coronary intervention. Am J Cardiol, 2009, 104(9): 1241-1244.
  • 7Kuon E, Empen K, Robinson DM, et al. Efficiency of a minicourse in radiation reducing techniques: a pilot initiative to encourage less irradiating cardiological interventional techniques. Heart, 2005, 91(9): 1221-1222.
  • 8Rivolta A, Emanuelli S, Tessarin C, et al. Method of patient dose evaluation in the angiographic and interventional radiology procedures. Radiol Med, 2005, 110(5-6): 689-698.
  • 9Suzuki S, Furui S, Yamakawa T, et al. Radiation exposure to patients’ skin during cardiac resynchronization therapy. Europace, 2009, 11(12): 1683-1688.

共引文献16

同被引文献10

引证文献2

二级引证文献1

相关作者