相位对比法磁共振成像测量主动脉和肺动脉起始部净血流量的准确性研究 被引量:6

Measurement Accuracy of Net Blood Flow Volume of the Initial Part of Aorta and Main Pulmonary Artery by Using PC MRI
下载PDF
导出
摘要 目的采用相位对比法(PC)磁共振成像(MRI)测量健康志愿者的主、肺动脉起始部净血流量,并进行比较,以验证PC MRI测量主、肺动脉起始部净血流量的准确性及可重复性。方法选取2012年1月—2013年12月昆明医科大学第二附属医院进行体检的健康志愿者31例,采用PC MRI测量心动周期内主、肺动脉起始部净血流量,观察PC MRI测量主、肺动脉起始部净血流量的准确性;前后两次测量9例健康志愿者主、肺动脉起始部净血流量,观察PC MRI测量主、肺动脉起始部净血流量的可重复性。结果主、肺动脉起始部时间-净血流量图呈逐渐上升曲线,相对应心脏的收缩期,主、肺动脉起始部净血流量曲线表现为倾斜度较大、光滑且连续的曲线;相对应心脏的舒张期,主、肺动脉起始部净血流量曲线表现为逐渐向上、波形较平缓、光滑且连续的曲线。主动脉起始部净血流量(72.8±12.2)ml,肺动脉起始部净血流量(73.8±13.8)ml,差异无统计学意义(t=-1.924,P=0.64)。主动脉起始部净血流量与肺动脉起始部净血流量呈正相关(r=0.94,P<0.05),方程为:Y=-3.825+1.066X。第1次与第2次测量主、肺动脉起始部净血流量比较,差异均无统计学意义(P>0.05)。第1次与第2次测量主动脉起始部净血流量呈正相关(r=0.91,P<0.05),方程为:Y=-6.669+1.078X。第1次与第2次测量肺动脉起始部净血流量呈正相关(r=0.92,P=0.001),方程为:Y=5.391+0.923X。结论 PC MRI是一种无创、安全、准确、信息量大的检查方法,且不受肺气、骨骼的影响,是研究主、肺动脉起始部净血流量的好方法。 Objectjve In order to evaluate the accuracy and repeatability of PC MRI in measuring the net blood flow volume of aorta and main pulmonary artery,the net blood flow volume of aorta and main pulmonary artery is measured by using PC MRI among healthy volunteers,the results were compared. Methods 31 healthy volunteers who received health examination in the Second Affiliated Hospital of Kunming Medical University from January 2012 to December 2013,were selected as study subjects,the net blood flow volume of the initial part of aorta and pulmonary artery in the cardiac circle was measured by using PC MRI,the accuracy of PC MRI in measuring the net blood flow volume was analyzed. To find out the measurement repeatability of net blood flow volume of aorta and main pulmonary artery by using PC MR,the net blood flow volume of aorta and pulmonary artery is measured by using PC MRI among 9 healthy volunteers twice. Results The time - net blood flow volume diagram for the initial part of aorta and pulmonary artery showed a rising curve. In systole,the time - net blood flow volume diagram for the initial part of aorta and pulmonary artery showed a smooth,continuous curve with high inclination;in diastole,the time - blood flow volume diagram for the initial part of aorta and pulmonary artery showed a smooth,continuous and rising curve with mildnbsp;waveform. There was no significant difference in the net blood flow volume between the initial part of aorta artery〔(72. 8 ± 12. 2) ml〕and the initial part of pulmonary artery〔(73. 8 ± 13. 8)ml〕(t = - 1. 924,P = 0. 64). The net blood flow volume of aorta was positively correlated with the net blood flow volume of main pulmonary artery( r = 0. 94,P 〈 0. 05),the equation:Y= - 3. 825 + 1. 066X. There was no significant difference in the net blood flow volume of aorta and main pulmonary artery between the first time and the second time(P 〉 0. 05). The net blood flow volume of aorta for the first time was positively correlated with the net blood flow volume of aorta for the second time(r = 0. 91,P 〈 0. 05),the equation:Y = - 6. 669 + 1. 078X. The net blood flow volume of main pulmonary artery for the first time was positively correlated with the net blood flow volume of main pulmonary artery for the second time(r = 0. 92,P = 0. 001),the equation:Y = 5. 391 + 0. 923X. Conclusjon PC MRI is a noninvasive,safe,accurate and informative method, and is not affected by air and bone, so it is a good method for hemodynamic study of aorta and pulmonary artery.
出处 《中国全科医学》 CAS CSCD 北大核心 2015年第15期1840-1844,共5页 Chinese General Practice
基金 教育部博士点基金新教师类(20115317120003) 云南省教育厅重点项目(2012Z091)
关键词 磁共振成像 主动脉 肺动脉 Magnetic resonance imaging Aorta Pulmonar artery
  • 相关文献

参考文献13

  • 1Henk CB,Grampp S,Koller J,et al.Elimination of errors caused by first - orderaliasing in velocity encoded cine - MRmeasurements of postoperative jets afteraortic coarctation : in vitro and in vivovalidation [J]. Eur Radiol, 2002, 12(6): 1523 -1531.
  • 2Tariq U,Hsiao A, Alley M, et al. Venousand arterial flow quantification are equallyaccurate and precise with parallel imagingcompressed sensing 4D phase contrast MRI[J]‘ J Magn Reson Imaging, 2013,37(6): 1419-1426.
  • 3Meier D,Maier S, BOsiger P. Quantitativeflow measurements on phantoms and on bloodvessels with MR [ J ]. Magn Reson Med,1988,8 (1) : 25 -34.
  • 4Duerk JL,Pattany PM. In - plane flowvelocity quantification along the phaseencoding axis in MRI [ J ]. Magn ResonImaging, 1988,6 (3) : 321 -333.
  • 5Nayler DL, Firmin DN, Longmore DB.Blood flow imaging by cine magneticresonance [ J ]. J Comput Assist Tomogra,1986,10 (5) : 715 -722.
  • 6Kondo C,Caputo GR,Semelka R, et al.Right and left ventricular stroke volumemeasurements with velocity - encoded cineMR imaging : in vitro and in vivo validation[J]. AJR Am J Roentgenol, 1991,157(1): 9-16.
  • 7Silverman JM, Julien PJ, Herfkens R J, et al. Quantitative differential pulmonary perfusion: MR imaging versus radionuclide lung scanning [ J ]. Radiology, 1993, 189 (3): 699-701.
  • 8Rees S,Firmin D, Mohiaddin R, et al.Application of flow measurements bymagnetic resonance velocity mapping tocongenital heart disease [ J] . Am J Cardiol,1989, 64 (14) : 953 -956.
  • 9Macfarlane PW, Lawrie TDV. Comprehensiveelectrocardiology [ M ]. New York:Pergamon Press, 1989 : 407 - 457.
  • 10James SH, Wald R,Wintersperger BJ,et ai.Accuracy of right and left ventricularfunctional assessment by short - axis vs axialcine steady - state free - precession magneticresonance imaging : intrapatient correlationwith main pulmonary artery and ascendingaorta phase — contrast flow measurements[J]. Can Assoc Radiol J, 2013, 64 (3):213 -219.

二级参考文献14

  • 1李彩英,田建明,刘辉,陆建平,黄勃源,刘怀军.利用磁共振相位对比法对非搏动流体模型的定量测量研究[J].中国医学影像技术,2006,22(4):535-537. 被引量:10
  • 2李彩英,黄勃源,崔彩霞,刘怀军,孟令惠.MRI电影相位对比与超声对照对正常志愿者餐前、餐后门静脉血流定量研究[J].放射学实践,2007,22(5):471-474. 被引量:6
  • 3Iotz J, Meier C, Leppert A, et al. Cardiovascular Flow Measure ment with Phase-contrast MR Imaging; Basic Facts and Implementation[J]. RadioGraphics,2002,22(3) :651-671.
  • 4Greil G,Geva T, Maier SE, et al. Effect of Acquisition Parameters on the Accuracy of Velocity Encoded Cine Magnetic Resonance Imaging Blood Flow Measurements[J].Magn Reson lmaging, 2002,15(1) :47-54.
  • 5Bakker CJ, Hoogeveen RM, Viergever MA. Construction of a Protocol for Measuring Blood Flow by Two-dimensional Phase -contrast MRA[J]. J Magn Reson Imaging,1999,9(1):119-127.
  • 6Buonocore MH, Bogren H. Factors Influencing the Accuracy and Precision of Velocity-encoded Phase Imaging [J].Magn Reson Med,1992,26(1):141-154.
  • 7Tang C,Blatter DD, Parker DL. Accuracy of Phase-contrast Flow Measurements in the Presence of Partial-volume Effects[J]. J Magn Reson Imaging, 1993,3 (2) : 377-385.
  • 8Stahlbe W, Sondergaard L, Thomsen C, et al. Quantification of Complex Flow Using MR Phase Imaging: a Study of Parameters Influency the Phase velocity Relation[J]. Magn Reson Imaging, 1992,10(1) :13-23.
  • 9Burkart DJ,Johnson CD, Ehman RL, et al. Evaluation of Portal Venous Hypertension with Cine Phase-contrast MR Flow Measurements: High Association of Hyperdynamic Portal Flow with Variceal Hemorrhage[J]. Radiology, 1993,188(3):643-648.
  • 10Lorie R,Norbert J, Stephen C, et al. Arterial and Venous Blood Flow: Noninvasive Quantitation with MR Imaging[J]. Radiology, 1992,185(3):809-812.

共引文献6

同被引文献17

引证文献6

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部