低管电压头颈CTA图像质量与辐射剂量的定量评估 被引量:2

Evaluation of Image Quality and Radiation Dose for Low-tube Voltage Head and Neck CTA
下载PDF
导出
摘要 目的:通过尝试直接降低管电压的方法控制辐射剂量,并评价图像质量,探讨其临床应用价值。方法:前瞻性选取2013年9月~2014年7月拟行头颈CTA检查的患者40例,并满足入组条件。患者分为A、B两组。A组患者采用低管电压设置,B组患者采用常规管电压设置。评价两组患者双侧颈动脉客观图像质量(包括CT值、SNR与CNR)与主观图像质量、辐射剂量。P〈0.05认为差异具有统计学意义。结果:A组SNR为11.39±3.80,B组总SNR为12.48±4.48,t=-1.660,P=0.099,差异无统计学意义。A组总对CNR为10.12±3.55,B组总CNR为10.44±4.14,t=-0.485,P=0.628,差异无统计学意义。A、B两组可评估率比较差异无统计学意义(χ2=1.013,P=0.314)。A组有效剂量为2.29 m Sv±0.12 m Sv,B组有效剂量为6.69 m Sv±0.31 m Sv,t=-59.756,P=0.000,差异具有统计学意义。结论:通过降低管电压至80 k V行头颈CTA扫描可以大幅度减低辐射剂量,而图像质量无明显影响,可以进行初步临床应用。 Objective: The purpose of this prospective study was to compare with image quality and radiation dose between low voltage and conventional head and neck CTA protocols. Methods: 40 consecutive patients were prospectively enrolled From September 2013 to July 2013, and satisfy the following conditions. Patients in group A was with low tube voltage settings, patients in group B was with traditional tube voltage settings. Two groups were evaluated for image quality, mean attenuation, noise, signal-to-noise ratio (SNR) and contrast-to-noise (CNR) in carotid artery and radiation dose. A P value below 0.05 was considered to be statistically significant. Results: The mean attenuation, noise, SNR and CNR was no statistically significant for group A versus group B (b 〉 0.05). The percentage of assessable coronary artery segments was no statistically significant for group B versus group A (χ2 = 1.013, P = 0.314). The mean effective dose was statistically significant for the group A was 2.29 mSv versus 6.69 mSv for the group B (t = -59.756, P = 0.000). Conclusion Low tube voltage for head and neck CTA technique could offer assessable image quality and substantially reduced effective radiation dose.
出处 《CT理论与应用研究(中英文)》 2015年第1期87-94,共8页 Computerized Tomography Theory and Applications
基金 内蒙古医科大学青年创新基金(YKD2013QNCX024)
关键词 体层摄影术 X线计算机 颈动脉 图像质量 辐射剂量 tomography X-ray computed carotid artery image quality radiation dose
  • 相关文献

参考文献11

  • 1Bongartz G,Golding SJ,Jurik AJ.European guidelines for multislice computed tomography:Report EUR 16262 EN 2004[R].Luxembourg:European Commission,2004.
  • 2Guidelines on radiation dose to the patient[M]//European Commission.European Guidelines on Quality Criteria for Computed Tomography.Report EUR 16262 EN.Office for Official Publication of the European Communities,Luxembourg,2010.
  • 3Eller A,Wuest W,Kramer M,et al.Carotid CTA:Radiation exposure and image quality with the use of attenuation-based,automated kilovolt selection[J].American Journal of Neuroradiology,2014,35(2):237-241.
  • 4Nakaura T,Kidoh M,Nakamura S,et al.Low-dose abdominal CT protocols with a tube voltage setting of 100 k Vp or 80 k Vp:Performance of radiation dose reduction and influence on visual contrast[J].Clinical Radiology,2014,pii:S0009-9260(14)00116-0.doi:10.1016/j.crad.2014.03.013.
  • 5李玮,刘建新,王霄英,朱颖.低电压、低对比剂剂量头颈CTA的可行性研究[J].放射学实践,2013,28(5):482-485. 被引量:45
  • 6吴国庚,王妍焱,周诚,焦晟,曹会志,沈云.64层螺旋CT超低剂量前门控轴面扫描在冠状动脉血管成像的应用[J].中华放射学杂志,2009,43(12):1255-1259. 被引量:39
  • 7唐坤,李瑞,林洁,郑祥武,曹国全.80kV低管电压扫描对脑CT血管成像图像质量及辐射剂量的影响[J].中国医学影像学杂志,2013,21(3):177-180. 被引量:23
  • 8Beitzke D,Wolf F,Edelhauser G,et al.Computed tomography angiography of the carotid arteries at low k V settings:A prospective randomised trial assessing radiation dose and diagnostic confidence[J].European Radiology,2011,21(11):2434-2444.
  • 9Bahner ML,Bengel A,Brix G,et al.Improved vascular opacification in cerebral computed tomography angiography with 80 k Vp[J].Investigative Radiology,2005,40(4):229-234.
  • 10Ertl-Wagner BB,Hoffmann RT,Bruning R,et al.Multidetector row CT angiography of the brain at various kilovoltage settings[J].Radiology,2004,231(2):528-535.

二级参考文献41

  • 1王霄英,薛华丹.泌尿生殖系统影像学2011年度进展报告[J].中国继续医学教育,2011,3(8):75-85. 被引量:23
  • 2李明利,金征宇,张云庆,王沄,赵文敏,陆菁菁.多层螺旋CT头颈部低剂量对比剂血管成像的可行性研究[J].临床放射学杂志,2005,24(3):214-217. 被引量:42
  • 3张森,杜湘珂,李剑颖.心电图调制电流对于容积CT心脏扫描图像质量及曝光剂量影响的评估[J].中华放射学杂志,2006,40(9):974-976. 被引量:45
  • 4Hunold P, Vogt FM, Schmermund A, et al. Radiation exposure during cardiac CT: effective doses at multi-detector row CT and electron-beam CT. Radiology,2003,226 : 145-152.
  • 5Trabold T, Buchgeister M, Kuttner A, et al. Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo, 2003, 175: 1051-1055.
  • 6Leber AW, Knez A, Becker A, et al. Accuracy of multidetector spiral computed tomography in identifying and differentiating the composition of coronary atherosclerotic plaques: a comparative study with intracoronary ultrasound. J Am Call Cardiol, 2004, 43 : 1241-1247.
  • 7Leber AW, Becker A, Knez A, et al. Accuracy of 64-slice computed tomography to classify and quantify plaque volumes in the proximal coronary, system : a comparative study using intravascular ultrasound. J Am Coil Cardiol, 2006, 47:672-677.
  • 8Mnllet NR, Cademartiri F, van Mieghem CA, et al. High- resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography.Circulation, 2005, 112:2318-2323.
  • 9Raft GL, Gallagher MJ, O'Neill WW, et al. Diagnostic accuracy of noninvasive coronary angiography using 64-slice spiral computed tomography. J Am Coll Cardiol, 2005, 46:552-557.
  • 10Hsieh J, Londt J, Vass M, et al. Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys, 2006, 33:4236-4248.

共引文献109

同被引文献25

引证文献2

二级引证文献16

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部