期刊文献+

二维矩阵用于加速器调强适形放射治疗质量控制检测的比较研究 被引量:2

Research on the use of two-dimensional array for IMRT quality control of medical accelerators
下载PDF
导出
摘要 目的:研究二维矩阵应用于调强适形放射治疗(IMRT)质量控制检测的可行性。方法:选取不同医院用于IMRT治疗的6台加速器,使用市面上常见的3种品牌二维矩阵对其重要的质量控制项目进行检测分析。结果:6台加速器的不同质量控制项目检测符合相应标准,不同矩阵间未出现明显差异。结论:使用二维矩阵进行IMRT质量控制检测简单可靠,可以作为IMRT加速器质量控制检测的有效方法。 Objective:To investigate the feasibility of two-dimensional array (2D-array) for IMRT quality control. Methods:Three kinds of 2D-array commercially available were used to measure several key QC procedures for six medical accelerators from different hospitals. Results: The measurement results can meet the standard requirement, no significant differences are found among the arrays. Conclusion:The simple and reliable method based on 2D-array is helpful with IMRT QC of medical accelerator.
出处 《中国医学装备》 2015年第5期62-65,共4页 China Medical Equipment
基金 卫生行业科研专项(201002009)"辐射危害控制与核辐射卫生应急处置关键技术研究及其应用" 中国疾控中心辐射安全所青年科学研究所长基金(201405)"螺旋断层放疗质量控制计量验证方法的研究"
关键词 二维矩阵 医用加速器 调强适形放射治疗质量控制 多叶准直器 Gamma分析 Two-dimensional array Medical accelerator Intensity modulated radiation therapy quality control Multileaf collimator
  • 相关文献

参考文献15

二级参考文献61

  • 1吴昊,蒋璠,韩树奎,包尚联.常见探测器的空间分辨率比较[J].中国医学物理学杂志,2011,28(2):2501-2504. 被引量:9
  • 2邹华伟,贾明轩,吴荣,肖福大,董晓琪.调强适形放射治疗计划剂量学的验证[J].中华放射医学与防护杂志,2004,24(3):256-258. 被引量:11
  • 3马金利,蒋国樑.调强适形放射治疗剂量学验证体系的建立[J].中国肿瘤,2004,13(8):485-489. 被引量:8
  • 4肖锋,孙朝阳,石梅,张学成,陆宙,陆军,刘晓莉.鼻咽癌动态调强放疗计划的剂量学验证[J].现代肿瘤医学,2007,15(2):178-180. 被引量:17
  • 5周颖娟,黄劭敏,邓小武.用放射性铬胶片进行调强放疗剂量验证的研究[J].中华放射肿瘤学杂志,2007,16(4):307-312. 被引量:12
  • 6A gazaryan N, Solber g TD, DeMarco JJ. Patient specific quality assurance for the delivery of intensity modulated radiotherapy[J]. J Appl Clin Med Phys, 2003, 4(1): 40-50.
  • 7Dong L, Antolak J, Salehpour M, et al. Patlent-specific point dose measurement for IMRT monitor unit verification [ J ]. Int J Radiat Oncol Biol Phys, 2003, 56 (3) : 867-877.
  • 8Dogan N, Leybovich LB, Sethi A. Comparative evaluation of Kodak EDR2 and XV2 films for verification of intensity modulated radiation therapy [J]. Phys Med Biol, 2002, 47(22): 4121-4130.
  • 9Stock M, kroupa B, Georg D. Interpretation and evaluation of the gamma index and gamma Index angle for the verification of IMRT hybrid plans[J]. Phys Med Biol, 2005, 50(3): 399-411.
  • 10Zhu XR, Jursinie PA, Grimm DF, et al. Evaluation of Kodak EDR2 film for dose verification of intensity modulated radiation therapy delivered by a statie multileaf collimator[J]. Med Phys, 2002, 29 (8) : 1687-1692.