期刊文献+

Sysmex XN-9000全自动血细胞分析仪性能评价 被引量:9

Performance evaluation of Sysmex XN-9000 automated hematology analyzer
下载PDF
导出
摘要 目的对日本Sysmex公司XN-9000全自动血细胞分析仪(简称XN-9000分析仪)进行性能评价。方法参照多项国际、国内标准对分析仪进行性能评价。结果 XN-9000分析仪批内及批间精密度、携带污染率、线性范围和准确性均符合相关质量控制要求。白细胞分类结果与手工分类比较,中性粒细胞、淋巴细胞、单核细胞和嗜酸性粒细胞相关性较好,嗜碱性粒细胞相关性不太理想。结论 XN-9000分析仪是一种较理想的全自动血细胞分析仪,性能良好,可以满足临床检验工作及疾病诊治的需要。 Objective To evaluate the performance of Sysmex XN‐9000 automated hematology analyzer .Methods According to international and domestic standards ,performance of analyzer was evaluated .Results The within‐batch and between‐batch preci‐sion ,carryover pollution rate ,linearity range and the accuracy of Sysmex XN‐9000 analyzer were all conform to related require‐ments .Leukocyte classification results compared with manual classification ,the correlation of neutrophil ,lymphocyte ,monocyte and eosinophil were fine ,but correlation of basophil was not very ideal .Conclusion The performance of Sysmex XN‐9000 analyzer could be satisfying ,could meet the needs of clinical inspection and diagnosis and treatment .
出处 《国际检验医学杂志》 CAS 2015年第10期1373-1375,共3页 International Journal of Laboratory Medicine
关键词 SYSMEX XN-9000全自动血细胞分析仪 性能评价 精密度 携带污染率 线性 准确性 白细胞分类 Sysmex XN-9000 automated hematology analyzer performance evaluation precision carryover pollution rate linearity range accuracy white blood cell classification
  • 相关文献

参考文献12

  • 1International Clinical Hyperthermia Society. Guidelines for the e- valuation of blood cell analysers including those used for differen- tial leucocyte and retieulocyte counting and cell marker applica- tions[J]. Clin Lab Haemat,1994,16(2) : 157-174.
  • 2Clinical and Laboratory Standards Institute. Method comparison and bias estimation using patient samples: Approved guideline- second edition: Document EP9-A2 [S]. Wayne, PA, USA: CLSI, 2002.
  • 3Clinical and Laboratory Standards Institute. User verification of performance for precision and trueness; Approved guideline-sec- ond edition: Document EP5-A2 IS:. Wayne, PA, USA: CLSI, 2005.
  • 4Clinical and Laboratory Standards Institute. Evaluation of lineari- ty oi quantitative measurement procedures:a statistical approach; Approved guideline; Document EP6-A [S]. Wayne, PA, USA: CLSI, 2003.
  • 5Enter for Devices and Radiological Health(CDRH), U. S. Depart- ment of Food and Drug Administration(FDA). Class U special controls guidance document:premarket notification for automated differential cell counters for immature of abnormal blood cells; Fi- nal Guidance for Industry and FDA[S]. MD, USA: FDA, 2001.
  • 6中华人民共和国卫生部.医疗机构临床实验室管理办法[Z].2006.
  • 7中国合格评定国家认可委员会.IS015189医学实验室质量和能力的通用要求[S].北京:中国合格评定国家认可委员会,2006.
  • 8叶应妩,王毓三,申子瑜.国临床检验操作规程[M].3版.南京:东南大学出版社,2006:123.
  • 9李惠玉,祁静,杨琛.366例不合格血标本原因分析及护理对策[J].实用临床医药杂志(护理版),2010,14(3):76-77. 被引量:28
  • 10Pierre RV. Peripheral blood film review. The demise of the eye- count leukocyte differential[J]. Clin Lab Med, 2002,22 ( 1 ) : 279- 297.