维持性血液透析患者自体动静脉内瘘血流量与心输出量和平均动脉压变化的相关性探讨 被引量:8

The correlation between arteriovenous fistula access flow and the changes of cardiac output and mean arterial pressure in maintenance hemodialysis patients
下载PDF
导出
摘要 目的观察使用自体动静脉内瘘(arteriovenous fistula,AVF)进行维持性血液透析(maintenance hemodialysis,MHD)患者的内瘘血流量(Qa)、心输出量(CO)和平均动脉压(MAP)的变化。方法选取使用自体动静脉内瘘进行MHD的患者54例。采用超声稀释法观察血液透析开始后30min、120min和180min3个时间点Qa、CO和MAP的变化,观察Qa的变化趋势及CO和MAP在不同时间点对Qa的影响。结果一次血液透析过程中3个时间点检测的患者动静脉内瘘血流量,以及各时间点心输出量、平均动脉压无明显波动,变化均无显著性差异(P>0.05)。透析过程中患者的动静脉内瘘Qa与即刻CO及MAP均成正相关(P<0.05)。透析30min至透析180min之间Qa变化的百分比(△Qa1-3%)与CO变化的百分比(△CO1-3%)(r=0367,P=0.009),与MAP变化的百分比(△MAP1-3%)(r=0.455,P=0.001)均存在正相关性。超滤量大于干体质量5%的患者Qa的变化及CO的变化显著大于超滤量小于干体质量5%的患者。(P<0.05)。结论MHD患者在血液透析过程各个时间段的动静脉内瘘血流量、心输出量、平均动脉压基本稳定并呈正相关,随着超滤脱水量的增加会影响内瘘血流量及心输出量的明显变化,进而影响血流动力学的稳定性。 Objective To observe the relationship between blood access from arteriovenous fistula(Qa)and the changes of cardiac output(CO) and mean arterial pressure(MAP) in maintenance hemodialysis(MHD) patients.Method A total of 54 MHD patients with arteriovenous fistula(AVF) were enrolled in this study.Variations of Qa by ultrasound dilution technique,CO,and MAP were observed at 30 minutes,2 hours,and 3 hours after starting a hemodialysis session to study Qa affected by the changes of CO and MAP in a hemodialysis session.Result Qa,CO and MAP changed insignificantly at the 3 time points in a hemodialysis session(P〉0.05).Qa was positively correlated with CO and MAP at the same time point(P〈0.05).,The percentage of Qa variation within the 3 time points(△Qa1-3%) was positively correlated with the percentage of CO variation(△CO1-3%)(r=0367,P=0.009),and the percentage of MAP variation(△MAP1-3%)(r=0.455,P=0.001).The variation of Qa and CO were significantly greater in patients having ultrafiltration volume of 〉5% dry body weight than in those having ultrafiltration volume of〈 5% dry body weight(P〈0.05).Conclusion Qa,CO and MAP were relatively stable in a hemodialysis session.Qa was positively correlated with CO and MAP.The increase of ultrafiltration volume in a session may result in the changes of Qa and CO and thus interfere with the stability of hemodynamics.
出处 《中国血液净化》 2015年第5期300-303,共4页 Chinese Journal of Blood Purification
关键词 血液透析 自体动静脉内瘘 内瘘血流量 超声稀释技术 Maintenance hemodialysis Arteriovenous fistula Access flow Ultrasound dilution tech-nology
  • 相关文献

参考文献11

  • 1钱家麒.关注血液透析中若干问题,进一步提高血液透析治疗水平[J].中华肾脏病杂志,2005,21(2):63-64. 被引量:30
  • 2Sehgal A R, Dor A, Tsai A C. Morbidity and cost implica tions of inadeqIlate hemodialysis[J]. Am J Kidney Dis 2001,37:1223- 1231.
  • 3KDOQI:Clinical practice guidelines for vascular access Update2006[J]. Am J Kidney Dis, 2006, Vo148, Nol, Suppll s176- s307.
  • 4华参,叶朝阳,李林,梅长林.HD02型血透监测仪对血液透析患者内瘘的监测[J].中华肾脏病杂志,2006,22(12):730-733. 被引量:24
  • 5Besarab A. Access monitoring is wor thw hile and valu- able. [J]. Blood Purif, 2006, 24(1) : 77- 89.
  • 6Casey E T, Murad M tt, Rizvi A Z, et al. Survoillance of ar- teriovenous hemodialysis access: a systematic review and meta-analysis[J]. J Vasc Surg, 2008,48:48-54.
  • 7Krivistski NM, Ganntela S. Access Flow measurement as a predictor of hemodialysis graft thrombosis:Making clin- ical decision[J].Semin Dial, 2001,14:181-185.
  • 8Christoph S, Christa M, Maryla B, et al. Flow monitoring: performance characteristics of ultrasound dilation ver- sus color Doppler ultrasound compared with fistulogra- phy[J]. Am J Kidney Dis, 2003, Vo142, No3:S31-545.
  • 9Tonelli M, Jindal R, Hirsch D, et al. Screening for subclin- ical stenosis in native vessel arteriovernous fistulae [J].J Am Soc Nephrol, 2001,12:1729-1733.
  • 10Basile C, Lomonte C, Vernaglione L, et al. The relationship between the flow of arteriovenous fistula and cardiac output in haemodialyMs patients[J].Nephrol Dial Trans- plant, 2008,23:282 -287.

二级参考文献13

  • 1张颖秋,梅长林,叶朝阳.维持性血透患者血管通路两种再循环率测定方法的比较[J].第二军医大学学报,2005,26(7):812-814. 被引量:10
  • 2National Kidney Foundation : K-DOQI clinical practice guidelines for vascular access : update 2000. Am J Kidney Dis, 2001,37 Suppl 1: S137-S181.
  • 3Jindal KK, Ethier J, Lindsay RM, et al. Clinical practice guidelines for vascular access. Canadian Society of Nephrology. J Am Soc Nephrol, 1999, 10 Suppl 13: S287-S321.
  • 4Besarab A. Access monitoring is worthwhile and valuable.Blood Purif, 2006,24:77-89.
  • 5Depner TA, Krivitski NM, MacGibbon D. Hemedialysis access recirculation measured by ultrasound dilution. ASAIO J, 1995,41: M749-M753.
  • 6Krivitski NM. Theory and validation of access flow measurement by dilution technique during hemodial-ysis. Kidney Int, 1995,48 : 244-250.
  • 7Krivitski NM. Novel method to measure access flow during hemodialysis by ultrasound velocity dilution technique. ASAIO J, 1995, 41: M741-M745.
  • 8Depner TA, Krivitski NM. Clinical measure mental blood flow in hemodialysis access fistulae and grafts by ultrasound dilution. ASAIOJ, 1995, 41: M745-M749.
  • 9Tonelli M, Jindal K, Hirsch D, et al. Screening for subclinical stenosis in native vessel arteriovenous fistulae. J Am Soc Nephrol, 2001,12:1729-1733.
  • 10Basile C, Ruggiefi G, Vernaglione L, et al. A comparison of methods for the measurement of hemodialysis access recirculation. J Nephrol, 2003, 16: 908-913.

共引文献52

同被引文献86

引证文献8

二级引证文献38

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部