期刊文献+

能谱CT成像多个测量参数分析 被引量:9

Multiple parameters in spectral CT imaging
下载PDF
导出
摘要 目的:探讨能谱CT扫描所测量CT值与碘浓度之间的关系,测量未知成分混合物的衰减系数、混合物密度和有效原子序数。方法:采用GE公司标准定量分析体模,体模圆周上9个孔,共安装3个试管(ROI1,ROI2,ROI3),试管长度100 mm,外径25 mm,内径18 mm,其内装有不同浓度标准碘液,浓度依次为0.6 mg/m L、0.4 mg/m L、0.2 mg/m L。结果:能谱CT成像可以探讨多参数成像,可以获得CT单能量成像,混合能量图像、能谱曲线//有效原子序数、混合物的密度对临床诊断的多参数分析提供了理论依据。结论:通过能谱CT的理论模型和实验数据结合分析,可以测量多个成像参数。 Objective To discuss on the relationship between iodine concentration and CT values measured by CT spectra imaging technology, and to measure the attenuation coefficient of mixture of the unknown composition, the density of mixture and the effective atomic number. Methods The standard quantitative analysis phantom of GE company was applied. There were nine holes on the circumference of a circle phantom. Three tubes (ROll, ROI2 ROI3), with 100 mm in the tube length, 25 mm in the outside diameter, and 18 mm in the inside diameter, were installed, with different concentrations of iodine solution. And the different concentrations were, in order, 0.6 mg/mL, 0.4 mg/mL, 0.2 mg/mL. Results Multi-parameter imaging could be studied by spectral CT imaging, and CT single energy imaging, hybrid energy imaging, spectral curve, effective atomic number and the density of mixture could be obtained, which provided a theoretical basis for the multi-parameter analysis of clinical diagnosis. Conclusion Multiple imaging parameters can be measured by analyzing the theoretical model and experimental data of spectral CT.
出处 《中国医学物理学杂志》 CSCD 2015年第3期317-321,共5页 Chinese Journal of Medical Physics
基金 国家自然科学基金(81470078)
关键词 能谱成像 碘含量 混合物密度 有效原子序数 spectral imaging iodine content attenuation mixture coefficient density effective atomic number
  • 相关文献

参考文献14

  • 1Jung D, Oh Y, Kim M, et al. Usefulness of the virtual monoc- hromatic image in dual-energy spectral ct for decreasing renal cyst pseudo enhancement: a phantom study[J]. AJR, 2012, 199(6): 1316- 1319.
  • 2Matsumoto K, Jinzaki M, Tanami Y, et al. Virtual monochromatic spectral imaging with fast kilo voltage switching: Improved image quality as compared with that obtained with conventional 120-kVp CT[J]. Radiology, 2011,259(1): 257-262.
  • 3Yu L, Leng S, Mccollough C. Dual-energy CT-based monochromatic imaging [J]. AJR, 2012, 199: s9-S15.
  • 4Toepker M, Moritz T, Krauss 13, et al. Virtual non-contrast in second- generation, dual-energy computed tomography: reliability of attenua- tion values[J]. Eur J Radiol, 2012, 81: e398-e405.
  • 5Xin L, Li FY, Andrew N, et al. Quantitative imaging of element composition and mass fraction using dual-energy CT: Threemat- erial decomposition[J]. Med Phys, 2009, 36(5): 1602-1609.
  • 6Primak A J, Ramirez G. Improved dual-energy material discrimina-tion for dual-source CT by means of additional spectral filtration[J]. Med Phys, 2009, 36: 1359-1369.
  • 7王乐,刘斌,余永强,王万勤,汪洁,周勇,吴兴旺,李小虎,朱晓红,张帅.能谱成像测量单能量下CT值与碘含量的相关性分析[J].实用放射学杂志,2012,28(5):784-787. 被引量:23
  • 8刘爱连,沈云.能谱CT临床应用图谱[M].北京:人民军医出版社.2014:2-26.
  • 9吉强,洪洋.医学影像物理学[M].北京:人民卫生出版社,2009:5-25.
  • 10Landry G, Joao S, Mathieu G, et al. Deriving effective atomic numbers from DECT basedon a parameterization of the ratio of high and lowlinear attenuation coefficients[J]. Phys Med Biol, 2013, 58: 6851-6866.

二级参考文献26

  • 1张建军,程广,刘铁.非增强螺旋CT与B超在泌尿系结石诊断中的价值比较[J].放射学实践,2007,22(6):594-596. 被引量:13
  • 2许骅 李剑颖.有效原子序数的原理与临床应用初探.JCT杂志,2011,:60-63.
  • 3Wang SC, Hsu YS, Chen KK, et al. Correlectlon between urianry tract pure stone composition and stone morphology on plainabdom inal film[J]. J Chin Med Assoc ,2004,67 ( 5 ) :235 - 238.
  • 4Deveci S, Coskun M, Tekin MI, et al. Spiral corn putedtom ography role in determination of chemical composition of pure and mixed urinary stones-an in vitro study[J].Urology, 2004,64 (2) :237 -240.
  • 5Motley G,Dalrymple N,Keesling C,et al. Housfidd united detemination of urinary stone composition[J]. Urology, 2001,58(20) : 170- 173.
  • 6Takahashi N, Hamnan RP, Vrtiska TJ, et al. Dual znergy CT iodine-subtraction virtual unenhancod technique to detect urinary stones in an iodine-filled collecting system:a phantom study [ J ]. A JR Am J Roentgenol, 2008,190 ( 5 ) : 1169 - 1173.
  • 7Vrtiska TJ, Hartman RP, Kofler JM, et al. Spatial resolution and radiation dose of a 64-MDCT scanner compaxvd with published CT urography protocols [ J ]. AJR Am J Roentgenol, 2009, 192 (4) :941 -948.
  • 8Yu L,Primak AN ,Liu X,et al. Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT[J]. Med Phys,2009,36(3) :1019 - 1024.
  • 9Primak AN, Ramirez Giraldo JC, Liu X, et al. Improved dual-energy material discrimination for dual-source CT by means of additional spectral filtration [ J ]. Med Phys, 2009,36 ( 4 ) : 1359 - 1369.
  • 10Thomas C,Krauss B,Ketelsen D,et al.Differentiation of urinarycalculi with dual energy CT:effect of spectral shaping by high en-ergy tin filtration[J].Invest Radiol,2010,45(7):393-398.

共引文献101

同被引文献76

引证文献9

二级引证文献55

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部