期刊文献+

保留导丝技术对起搏导线的影响

Effects of retained guidewire on pacing lead
原文传递
导出
摘要 目的 总结因保留导丝至起搏导线断裂的病例,观察保留导丝对起搏导线的影响 方法 观察了自2010年1月至2013年12月因导线断裂在南京鼓楼医院行导线拔除的患者6例,其中5例采用了保留导丝技术.结果 5例患者均成功拔除了导线.2根右心室主动固定导线的保留钢丝及导线内部线圈均断裂,断裂处在锁骨下静脉.2例左心室导线的保留导丝及导线线圈均断裂,断裂部位在三尖瓣环,1例左心室导线仅保留导丝断裂,但内部线圈仍完整,测试导线阻抗未见明显升高.结论 保留导丝可损伤导线绝缘层及内部线圈,建议慎用保留导丝技术. Objective Observe the influence of retained guidewire on the pacing lead.Methods From Jan.2010 to Dec.2013,Nanjing Drum Tower Hospital,six patients received leads extraction because of lead fracture were enrolled in our study.Results Five patients had retained guidewire.All leads were extracted using mechanical tools,including 2 right ventricular fixed leads and 3 left ventricular pacing leads.Fracture of the coil electrodes as well as deterioration of the insulation coating was found in 4 patients.1 retained guide in left ventricular pacing fracture,but the pacing parameters,including impedance was not elevated.Conclusion The retained guidewire technique should not be used because delayed electrode damage can occur.
出处 《中华心律失常学杂志》 2015年第2期113-114,共2页 Chinese Journal of Cardiac Arrhythmias
关键词 保留导丝 起搏器 导线断裂 Retention wire Pacemaker Lead fracture
  • 相关文献

参考文献8

  • 1McAlister FA, Ezekowitz JA, Wiebe N, et al. Systematic review: cardiac resynchronization in patients with symptomatic heart failure [ J ]. Ann Intern M ed ,2004,141:381-390.
  • 2Koos R,Sinha AM, Markus K,et al. Comparison of left ventricular lead placement via the coronary venous approach versus lateral thoracotomy in patients receiving cardiac resynchronization therapy [J]. Am J Cardiol,2004,94: 59-63.
  • 3Cock CC,Jessurun ER, Allaart CA, et al. Repetitive intraoperative dislocation during transvenous left ventricutar Lead implantation: usefulness of the retained guidewire technique[J]. Pacing Clin Electrophysiol, 2004,27 : 1589 - 1593.
  • 4Sharifkazemi MB,Aslani A. Stabilization of the coronary sinus lead position with permanent styler to prevent and treat dislocation [ J]. Europace, 2007,9 : 875- 877.
  • 5陈少伯,姜铁民,梁国庆,赵季红,赵鹏,李玉明.心室同步化治疗保留左室电极导引钢丝技术的初步探讨[J].介入放射学杂志,2013,22(2):99-101. 被引量:3
  • 6Nagele H, Hashagen S, Ergin M, et al. Corona~ sinus lead frag- mentation 2 years after implantation with a retained guidewire[ J]. Pacing Clin Eleetrophysiol, 2007,30:438- 439.
  • 7Arbelo E, Gareia-Quintana A, Caballero E, et al. Late failure of left ventricular leads stabilized using the retained guidewire technique in patients undergoing cardiac resynchronization therapy [ J ]. Rev Esp Cardio1,2008,61 : 91-94.
  • 8Charles K,Christian B, Roger W,et al. A single-centre experience of over one thousand lead extractions [ J ]. Europaee, 2009, 11 : 612-617.

二级参考文献11

  • 1Schl(o)sser M,Stellbrink C. Indication for CRT[J].Herzschrittmacherther Elektrophysiol,2009.103-108.
  • 2Al-Khadra AS. Use of a modified introducer sheath with a sidehole to improve access to left ventricular veins with proximal origin[J].Europace:European Pacing,Arrhythmias and Cardiac Electrophysiology,2006.56-59.
  • 3Papiashvili G,Ann6 W,Duytschaever M. Just another case of lead dislocation[J].Journal of Cardiovascular Electrophysiology,2010.216-218.
  • 4De Cock CC,Jessurun ER,Allaart CA. Repetitive intraoperative dislocation during transvenous left ventricular Lead implantation:usefulness of the retained guidewire technique[J].Pacing and Clinical Electrophysiology,2004.1589-1593.
  • 5Yi F,Shen M,Wu F. Improving left ventricular pacing threshold using retained guidewire technique:a case report[J].Europace:European Pacing,Arrhythmias and Cardiac Electrophysiology,2010.1792-1793.
  • 6Sankaranarayanan R,James MA. Use of a retained guidewire to improve the left ventricular pacing threshold[J].Pacing and Clinical Electrophysiology,2007.1285-1287.
  • 7Sherzer AI,Feigenblum DY,Pina JW. Use of the retained guidewire technique facilitates left ventricular epicardial capture during biventricular defibrillator implantation[J].Pacing and Clinical Electrophysiology,2007.436-437.
  • 8de Cock CC,Res JC,Hendriks ML. Usefulness of a pacing guidewire to facilitate left ventricular lead implantation in cardiac resynchronization therapy[J].Pacing and Clinical Electrophysiology,2009.446-449.
  • 9N(a)gele H,Hashagen S,Ergin M. Coronary sinus lead fragmentation 2 years after implantation with a retained guidewire[J].Pacing and Clinical Electrophysiology,2007.438-439.
  • 10Arbelo E,García-Quintana A,Caballero E. Late failure of left ventricular leads stabilized using the retained guidewire technique in patients undergoing cardiac resynchronization therapy[J].Revista Espanola De Cardiologia,2008.91-94.