Novel calibration optical path of cryogenic radiometer 被引量:4

Novel calibration optical path of cryogenic radiometer
原文传递
导出
摘要 We introduce a novel calibration optical path of the cryogenic radiometer, which can avoid the repeated dis- mounting measurement and eliminate the negative influence of Brewster window effect on calibration result. The novel calibration optical path is used to calibrate the absolute spectral responsivity of the standard transfer detector at 633 nm, the results of which are compared with the ones of the previous structure. It is shown that comparing the previous results to the structure optimization, the measurement uncertainty of laser power reduces by a factor of 2, the measurement uncertainty on the absolute spectral responsivity of the transfer detector decreases by 15%, and the consistence of the calibration on absolute spectral responsivity is 4.0 × 10^-3. The experiment result proves that the novel calibration optical path of cryogenic radiometer can effectively reduce the calibration uncertainty against standard detector and improve the accuracy of calibration. We introduce a novel calibration optical path of the cryogenic radiometer, which can avoid the repeated dis- mounting measurement and eliminate the negative influence of Brewster window effect on calibration result. The novel calibration optical path is used to calibrate the absolute spectral responsivity of the standard transfer detector at 633 nm, the results of which are compared with the ones of the previous structure. It is shown that comparing the previous results to the structure optimization, the measurement uncertainty of laser power reduces by a factor of 2, the measurement uncertainty on the absolute spectral responsivity of the transfer detector decreases by 15%, and the consistence of the calibration on absolute spectral responsivity is 4.0 × 10^-3. The experiment result proves that the novel calibration optical path of cryogenic radiometer can effectively reduce the calibration uncertainty against standard detector and improve the accuracy of calibration.
出处 《Chinese Optics Letters》 SCIE EI CAS CSCD 2015年第5期35-39,共5页 中国光学快报(英文版)
基金 supported by the National Natural Science Foundation of China under Grant Nos.61275173 and 11204318 the National Defense Technology Foundation of China under Grant Nos.J2920130004
关键词 Cryogenic equipment Cryogenics Radiometers Structural optimization Uncertainty analysis Cryogenic equipment Cryogenics Radiometers Structural optimization Uncertainty analysis
  • 相关文献

参考文献17

  • 1J. E. Martin, N. P. Fox, and P. G. Key, Metrologia 21, 147 (1985).
  • 2P. V. Foukal, H. Kochling, and P. Miller, Appl. Opt. 29,988 (1990).
  • 3T. R. Gentile, J. M. Houston, and J. E. Hard!s, Appl. Opt. 35, 1056 (1996).
  • 4M. Sildoja, F. Manoocheri, M. Merimaa, E. Ikonen, I. Muller, L. Werner, J. Gran, T. Kubarsepp, M. Smid, and M. L. Rastello, Metrologia 50,385 (2013).
  • 5I. Muller, U. Johannsen, U. Linke, L. Socaciu-Siebert, M. Smid, G. Porrovecchio, M. Sildoja, F. Manooeheri, E. Ikonen, J. Gran, T. Kubarsepp, G. Brida, and L. Werner, Metrologia 50, 395 (2013).
  • 6G. P. Eppeldauer, H. W. Yoon, J. Zeng, and T. C. Larason, Metro- logia 49, 112 (2012).
  • 7A. Fehlmannl, G. Kopp, W. Schmutz, and R. Winkler, Metrologia 49, 34 (2012).
  • 8S. Cui, Z. Wang, and S. Yang, Chin. Opt. Lett. 12, 110101 (2014).
  • 9J. Zhang, Y. Lin, J. Shao, and Q. Fan, Aeta MetroIogica Sin. 19, I94 (1998).
  • 10X. Zheng, H. Wu, J. Zhang, Y. Liu, W. Zhou, L. Wang, and Y. Qiao, Chin. Sci. Bull. 45, 1341 (2000).

同被引文献22

引证文献4

二级引证文献11

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部