锥形束CT引导下前列腺癌调强放疗图像配准的研究 被引量:2

Image alignment assessed with k V cone beam CT for image guided prostate radiotherapy
下载PDF
导出
摘要 目的:通过千伏级锥形束CT(KV-CBCT)引导的前列腺癌调强放疗的数据,探讨前列腺癌图像配准的最佳模式。方法:入选16例接受根治性调强放疗的前列腺癌患者,共行KV-CBCT在线校正治疗体位214次。采用常规皮肤标记激光对位后采集图像,离线将所获得CBCT图像与计划CT图像进行自动配准、骨性配准、软组织配准及手动配准。比较四种配准方式之间的差异。结果:16例患者四种配准方式在X(左右)、Y(升降)、Z(头脚)三个方向的摆位误差分别为-0.060±0.28、-0.062±0.45、-0.065±0.38;-0.071±0.27、-0.098±0.45、-0.085±0.41;-0.077±0.26、-0.027±0.44、-0.111±0.40;-0.059±0.29、-0.075±0.51、-0.098±0.39。经分析四种配准方式之间没有统计学差异。结论:综合各种因素考虑,应用在线自动配准KV—CBCT引导放疗技术治疗前列腺癌患者更为合适。 Objective: To investigate the optimal alignment methods with kilovohage cone-beam CT ( KV-CBCT) for the treat-ment of prostate cancer.Metheds: 16 patients with prostate cancer undergoing IMRT were enrolled onto the study and 214 CBCT im-agings were obtained from this study.After initial set-up using the skin marks, a CBCT scan was acquired and registered with the planning CT.The difference in auto alignment, bone alignment, soft tissue alignment and manual alignment was compared.Results:There was no statistical difference between the four alignment methods.The comparisons of setup errors in the vertical direction、lateral direction and longitudinal direction were -0.060 ±0.28、 -0.062 ±0.45、 -0.065 ±0.38; -0.071 ±0.27、 -0.098 ±0.45、-0.085 ±0.41; -0.077 ±0.26、 -0.027 ±0.44、 -0.111 ±0.40; -0.059 ±0.29、 -0.075 ±0.51、 -0.098 ±0.39 re-spectively.Conclusions Due to the comprehensive factors, the auto alignment should be used in the treatment.
机构地区 北京医院放疗科
出处 《医疗装备》 2015年第6期27-30,共4页 Medical Equipment
关键词 前列腺肿瘤/调强放射疗法 在线图像引导 锥形束CT 图像配准 Prostate neoplasms / intensity modulated radiotherapy Online image guided CBCT Alignment
  • 相关文献

参考文献5

  • 1Remeijer P, Geerlof E, Ploeger L, et al. 3 -D portal im- age analysis in clinical practice: an evaluation of 2 - D and 3 - D analysis techniques as applied to 30 prostate cancer patients, Int J Oncol Biol Phys, 2000, 46:1281 - 1290.
  • 2van Herk M. Errors and margins in radiotherapy. Semin Ra- diat Oncol Biol, 2004, 14 : 52 - 64.
  • 3Remeijer P, Geerlof E, Ploeger L, et al. 3 - D portal im- age analysis in clinical practice: an evaluation of 2 -D and 3 -D analysiS techniques as applied to 30 prostate cancer patients. Int J Oncol Biol Phys, 2000, 46:1281 -1290.
  • 4Kupel Jan PA, Langen KM, Zerdan OA, et al. Daily var- iations in delivered doses in patients' treated with radiother- apy for localized prostate cancer. Int J Radiat Oncol Bi ol Phys, 2006, 66:876-882.
  • 5Schallenkamp JM, Herman MG, Kruse JJ, et al. Prostate position relative to pelvic bony anatomy based on intrapros- tatic gold markers and electronic portal imaging. Int J Radi- at Oncol Biol Phys, 2005, 63:800-811.

同被引文献28

  • 1戴建荣,胡逸民.图像引导放疗的实现方式[J].中华放射肿瘤学杂志,2006,15(2):132-135. 被引量:184
  • 2Piotrowski T, Kaczmarek K, Bajon T, et al. Evaluation of image-guidance strategies for prostate cancer[J]. Technol Cancer Res Treat, 2014, 13(6):583-591. DOI:10.7785/tcrtexpress.2013.600258.
  • 3International Federation of Gynecology and Obstetrics Committee on Gynecologic Oncology. Figo staging for carcinoma of the vulva, cervix, and corpus uteri[J]. Int J Gynaecol Obstet, 2014, 125(2):97-98. DOI:10.1016/j.ijgo.2014.02.003.
  • 4Lim K, Small W, Portelance L, et al. Consensus guidelines for delineation of clinical target volume for intensity-modulated pelvic radiotherapy for the definitive treatment of cervix cancer[J]. Int J Radiat Oncol Biol Phys, 2011, 79(2):348-355. DOI:10.1016/j.ijrobp.2009.10.075.
  • 5Toita T, Ohno T, Kaneyasu Y, et al. A consensus-based guideline defining the clinical target volume for pelvic lymph nodes in external beam radiotherapy for uterine cervical cancer[J]. Jpn J Clin Oncol, 2010, 40(5):456-463. DOI:10.1093/jjco/hyp191.
  • 6Stroom JC, Heijmen BJ. Geometrical uncertainties, radiotherapy planning margins, and the ICRU-62 report[J]. Radiother Oncol, 2002, 64(1):75-83.
  • 7Van Herk M. Errors and margins in radiotherapy[J]. Semin Radiat Oncol, 2004, 14(1):52-64. DOI:10.1053/j.semradonc.2003.10.003.
  • 8Hui B, Zhang Y, Shi F, et al. Association between bone marrow dosimetric parameters and acute hematologic toxicity in cervical cancer patients undergoing concurrent chemoradiotherapy:comparison of three-dimensional conformal radiotherapy and intensity-modulated radiation therapy[J]. Int J Gynecol Cancer, 2014, 24(9):1648-1652. DOI:10.1097/IGC.0000000000000292.
  • 9Boswell S, Tome W, Jeraj R, et al. Automatic registration of megavoltage to kilovoltage CT images in helical tomotherapy:an evaluation of the setup verification process for the special case of a rigid head phantom[J]. Med Phys, 2006, 33(11):4395-4404.
  • 10Laursen LV, Elstr?m UV, Vestergaard A, et al. Residual rotational set-up errors after daily cone-beam CT image guided radiotherapy of locally advanced cervical cancer[J]. Radiother Oncol, 2012,105(2):220-225. DOI:10.1016/j.radonc.2012.08.012.

引证文献2

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部