5种常用口腔材料的磁共振伪影比较 被引量:8

Comparison of magnetic resonance imaging artifacts of five common dental materials
下载PDF
导出
摘要 目的比较口腔科常用的3种金属和2种陶瓷材料在0.35、1.5及3 T 3种场强磁共振成像(MRI)不同序列中形成伪影的差异。方法将3种金属和2种陶瓷材料制作成同样大小及厚度的中切牙内冠,将其置入盛满生理盐水的塑料罐体中,分别在0.35、1.5和3.0 T的磁共振(MR)机中扫描获得T1WI、T2WI影像。比较各种材料在不同场强的MR扫描中伪影的差异。结果氧化锆在3种场强的MR机中扫描无明显伪影;铸造陶瓷伪影极轻微;贵金属合金、烤瓷镍铬合金与烤瓷钴铬合金在3种场强MR扫描中均出现不同程度的伪影。结论氧化锆、铸造陶瓷这两种常用口腔陶瓷材料基本无伪影或伪影极轻微。贵金属合金、烤瓷镍铬合金与烤瓷钴铬合金3种常用口腔金属材料易引起MRI伪影,且随着主磁体场强增加,伪影面积增大。 Objective To compare five materials commonly used in dentistry, including three types of metals and two types of ceramics, by using different sequences of three magnetic resonance imaging (MRI) field strengths (0.35, 1.5, and 3.0 T). Methods Three types of metals and two types of ceramics that were fabricated into the same size and thickness as an incisor crown were placed in a plastic tank filled with saline. The crowns were scanned using an magnetic resonance (MR) machine at 0.35, 1.5, and 3.0 T field strengths. The T1WI and T2WI images were obtained. The differences of various materials in different artifacts of field MR scans were determined. Results The zirconia crown presented no significant artifacts when scanned under the three types of MRI field strengths. The artifacts of casting ceramic were minimal. All dental precious metal alloys, nickel-chromium alloy dental porcelain, and cobalt-chromium ceramic alloy showed varying degrees of artifacts under the three MRI field strengths. Conclusion Zirconia and casting ceramics present almost no or faint artifacts. By contrast, precious metal alloys, nickel-chromium alloy dental porcelain and cobalt-chromium ceramic alloy display MRI artifacts. The artifact area increase with increasing magnetic field.
出处 《华西口腔医学杂志》 CAS CSCD 北大核心 2015年第3期230-233,共4页 West China Journal of Stomatology
关键词 口腔材料 金属 陶瓷 磁共振成像 伪影 dental material metal ceramics magnetic resonance imaging artifact
  • 相关文献

参考文献10

  • 1高元贵,蔡幼拴,蔡祖龙.磁共振成像诊断学[M].3版.北京:人民军医出版社.1999:122-128.
  • 2霍沃德里.颅脑MR和CT诊断学[M].西安:世界图书出版西安公司,2001:75-99.
  • 3Shellock FG, Shellock VJ. Metallic stents: evaluation of MR imaging safety[J]. AJR Am J Roentgenol, 1999, 173 (3):543-547.
  • 4Viano AM, Gronemeyer SA, Haliloglu M, et al. Improved MR imaging for patients with metallic implants[J]. Magn Reson Imaging, 2000, 18(3):287-295.
  • 5Shafiei F, Honda E, Takahashi H, et al. Artifacts from den- tal casting alloys in magnetic resonance imaging[J]. J Dent Res, 2003, 82(8):602-606.
  • 6Lissac M, Metrop D, Brugirard J, et al. Dental materials and magnetic resonance imaging[J]. Invest Radiol, 1991, 26(1): 40-45.
  • 7Tartaglino LM, Flanders AE, Vinitski S, et al. Metallic ar- tifacts on MR images of the postoperative spine: reduction with fast spin-echo techniques[J]. Radiology, 1994, 190(2): 565-569.
  • 8戈明媚,王秋良,刘志钦,刘晓林,廖玉珍,孙玮,孙楠,张英魁.MR螺旋桨扫描技术在消除伪影方面的临床应用[J].中华放射学杂志,2006,40(2):208-212. 被引量:37
  • 9陈财忠,李韧晨,张澍杰,王忠,李静,曾蒙苏.去金属伪影序列在脊柱金属植入物患者MR成像中的应用[J].中华放射学杂志,2014,48(4):320-323. 被引量:27
  • 10刘广顺,任庆云,孟令强,雷立存,王瑶.口腔金属材料对磁共振成像的影响[J].华西口腔医学杂志,2010,28(5):505-508. 被引量:14

二级参考文献30

  • 1赖寿伟,杨华岳.金属固定假牙对头部MR图像的影响研究[J].实用医技杂志,2005,12(02A):350-352. 被引量:11
  • 2邵晨婴,俞立英.口腔内金属修复物与磁共振伪影的关系[J].国际口腔医学杂志,2007,34(2):143-145. 被引量:12
  • 3van Everdingen KJ,van der Grond J,Kappelle IJ,et al.Diffusion-weighted magnetic resonance imaging in acute stroke[J].Stroke.1998,29(9):1783-1790.
  • 4Nishimura DG,Irarrazabal P,Meyer CH.A velocity k-space analysis of flow effects in echo-planar and spiral imaging.Magn Reson Med,1995,33:549-556.
  • 5Glover GH,Pauly JM.Projection reconstruction techniques for reduction of motion effects in MRI.Magn Reson Med,1992,28:275-289.
  • 6Meyer CH,Hu BS,Nishimura DG,et al.Fast spiral coronary artery imaging.Magn Reson Med,1992,28:202-213.
  • 7Ehman R,Felmlee J.Adaptive technique for high-definition MR imaging of moving structures.Radiology,1989,173:255-263.
  • 8Ordidge RJ,Helpern JA,Qing ZX,et al.Correction of motional artifacts in diffusion-weighted MR images using navigator echoes.Magn Reson Imaging,1994,12:455-460.
  • 9Sachs T,Meyer CH,Irarrazabal P,et al.The diminishing variance algorithm for real-time reduction of motion artifacts in MRI.Magn Reson Med,1995,34:412-422.
  • 10Pipe JG.Motion correction with PROPELLER MRI:application to head motion and free-breathing cardiac imaging.Magn Reson Med,1999,42:963-969.

共引文献78

同被引文献55

  • 1石旭旭,朱晓敏,毛艳,张铁刚.两种非贵金属烤瓷冠对牙龈影响[J].黑龙江医药科学,2009,32(5):39-40. 被引量:2
  • 2刘玉华,孙樱琳.固定义齿修复材料对MRI图像的影响[J].现代口腔医学杂志,2005,19(6):588-589. 被引量:19
  • 3Laakman RW, Kaufman B, Han JS, et al. Mr imaging in patients with metallic implants [ J ]. Radiology, 1985,157 ( 3 ) :711 - 714.
  • 4Vikhoff B, Ribbelin SK, Hler B, et al. Artefacts caused by dental filling materials in Mr imaging[ J]. Acta Radiologica, 1995,36 ( 3 ) : 323 - 325.
  • 5ChristopherR, Camacho. Nonsusceptibility Artifacts Due toMetallic Objects in MR Imaging [J]. Ori Res,1995,8(5)∶75-88.
  • 6JohnF S. The role of magnetic susceptibility in magnetic resonance imaging:magnetic compatibility of the first and second kinds [J]. MedPhys,1996,21(23)∶815-850.
  • 7KressB, BuhlY, AndersL, et al. Quantitative analysis of MRI signals intensity as a tool for evaluating tooth pulp vitality [J]. DentomaxillofacRad,2004,33(1)∶241-244.
  • 8Tymofiyeva OS, Rottner K, Boldt J, et al. Influence of dental materials on dental MRI [J]. DentomaxillofacRad,201342(1)∶1-9.
  • 9Abu BE, AbeerE, SamahS, et al. Effect of Non-ionizing Radio Frequency Signals of Magnetic Resonance Imaging On Physical Propertoes of Dental Alloys and Metal-Ceramic Adhesion [J]. J Adhes Dent,2014(16)∶407-413.
  • 10昌仁民,周新韩,文戈,陈斌,胡罢生,曲华丽,邱士军,张雪林.3.0T磁共振常见伪影的产生机理、表现及解决措施[J].中国医疗设备,2008,23(6):95-99. 被引量:7

引证文献8

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部