期刊文献+

一种离心式微流控生化分析芯片 被引量:4

A Centrifugal Microfl uidic Chip for Biochemical Analysis
下载PDF
导出
摘要 文章介绍了一种低成本、快速、自动化的离心式微流控生化分析芯片。该芯片通过整合样本前处理和多生化指标检测,采用了多级微流道与虹吸微阀相结合的方式把样本的前处理、样本输送、样本与生化试剂反应等过程集成于一体,并通过比色检测获得各个生化检测项目的结果。实验结果显示芯片比色孔光程的精密度的变异系数在0.08%~0.52%,通过血糖结果的精密度评估的芯片对样本和稀释液定量以及两者混合的精密度变异系数为1.4%,同时使用芯片测定15个生化项目的日间和批内精密度均小于3.5%,这表明该离心式微流控生化分析芯片符合临床检测要求。 In this paper, a low-cost, rapid, automated centrifugal microfluidic chip for biochemical analysis was presented. The chip integrates sample pre-processing and multi-biochemical items assay. Through multi-level microfluidic channels and micro-siphon valves, the chip combines the pre-treatment of samples, sample transportation and the reaction processes between sample and biochemical reagent. The results are obtained by colorimetric detection. It is shown that the reproducibility of cuvette pathlength for chip is from 0.08% to 0.52%. The coefficient of variation (CV) of sample, diluents metering and mixing for chip by the CV of blood glucose results is 1.4%. The CV of day-to-day and within-run precision for 15 biochemical items are less than 3.5%. Therefore, it is concluded that the centrifugal microfluidic chip for biochemical analysis can satisfy the clinical testing requirements
出处 《集成技术》 2015年第3期86-92,共7页 Journal of Integration Technology
关键词 微流控芯片 生化分析 床旁诊断 microfluidic chip biochemical analysis point-of-care testing
  • 相关文献

参考文献1

二级参考文献8

  • 1苑广信,何巧红,陈恒武,方群,张燕冰.芯片毛细管电泳激光诱导荧光快速分离检测麻黄碱类兴奋剂[J].分析化学,2006,34(12):1693-1696. 被引量:8
  • 2Fang Zhao-Lun(方肇伦).Microfluidic Chip(微流控分析芯片).Beijing(北京):Science Press(科学出版社),2003:1-12
  • 3Fang Qun(方群),Fu Jing-Lin(富景林),Huang Yan-Zhen(黄艳贞).Chinese Patent(中国发明专利),Appl.No:03141572.5,2003
  • 4Fu J L,Fang Q,Zhang T,Jin X H,Fang Z L.Anal.Chem.,2006,78:3827-3837
  • 5Jia Z J,Fang Q,Fang Z L.Anal.Chem.,2004,76:5597-5602
  • 6Jacobson S C,Koutny L B,Hergenroder R,Moore A W Jr,Ramsey J M.Anal.Chem.,1994,66:3472-3476
  • 7Jacobson S C,Hergenroder R,Koutny L B,Warmack R J,Ramsey J M.Anal.Chem.,1994,66:1107-1113
  • 8Du X G,Fang Z L.Electrophoresis,2005,26:4625-4631

共引文献12

同被引文献50

  • 1Sia SK, Kricka LJ. Microfluidics and point-of care testing [J]. Lab Chip,2008,8(12) .1982-1983.
  • 2NCCLS. Preliminary Evaluation of Quantitative Clinical La- boratory Methods[S]. Approved Guideline EP10-A2. 2002.
  • 3Franke TA, Achim W. Microfluidics for miniaturizedlaboratories on a chip [J]. Chemphyschem, 2008,9(15):2140-2156.
  • 4Yeo LY,Chang HC, Chan PPY, et al. Microfluidicdevices for bioapplications [J]. Small, 2011, 7(1):12-48.
  • 5Studer V,Pepin A, Chen Y, et al. An integratedAC electrokinetic pump in a microfluidic loop forfast and tunable flow control [J]. Analyst, 2004,129(10):944-949.
  • 6Wang G, Ho HP, Chen Q, et al. A lab-in-a-dropletbioassay strategy for centrifugal microfluidics withdensity difference pumping, power to disc andbidirectional flow control [J]. Lab on A Chip, 2013,13(18):3698-3706.
  • 7Hartshome H, Backhouse CJ, Lee WE. Ferrofluid-based microchip pump and valve [J]. Sensors &Actuators B Chemical, 2004,99 (2-3): 592-600.
  • 8Erickson D, Serey X,Chen YF, et al. Nanomani-pulation using near field photonics [J]. Lab on AChip, 2011, 11(6):995-1009.
  • 9Grigorenko AN, Roberts NW,Dickinson MR,et al. Nanometric optical tweezers based onnanostructured substrates [J]. Nature Photonics,2008,2 (6): 365-370.
  • 10Chen J, Kang Z, Wang G,et al. Optofluidic guiding,valving, switching and mixing based on plasmonicheating in random gold nanoisland substrate [J].Lab on A Chip, 2015, 15(11): 2504-2512.