电子射野影像装置的位置标定研究

Study of Position Calibration on Electronic Portal Imaging Device
下载PDF
导出
摘要 目的研究电子射野影像装置(EPID)安装时2种不同的高度标定方法和2种不同的中心标定方法,找出方便、快捷且准确的EPID位置标定方法。方法对于1台装有EPID的加速器,分别采用金属球法和3D Frame法进行EPID的高度标定,分别采用金属球法和Collimator QA法进行EPID的中心标定。结果根据不同方法采集的图像,计算对应的高度标定值以及中心标定值。结论 3D Frame法与Collimator QA法更加稳定可靠,适合作为临床上EPID的安装标定与质检方法。 Objective To find a more convenient, more effective and more accurate method to implement the postion calibration of electronic portal imaging device (EPID) by studying two kinds of height calibration method and two kinds of center calibration method. Methods For the same accelerator equipped with EPID, metal ball method and 3D Frame method was used to implement height calibration respectively, and mental ball method and Collimator QA method was used to implement center calibration respectively. Results According to the captured images,the height calibration value and the center calibration value were worked out respectively. Conclusion 3D Frame method and Collimator QA method are more stable and reliable, and more suitable for clinical EPID calibration and quality assurance.
作者 李勇 杜鹏
出处 《肿瘤基础与临床》 2015年第3期232-234,共3页 journal of basic and clinical oncology
关键词 电子射野影像装置 高度标定 中心标定 electronic portal imaging device height calibration center calibration
  • 相关文献

参考文献9

二级参考文献35

  • 1张春光,高兴旺,黄劭敏,祁振宇.利用EPIDs进行直线加速器剂量对称性测量的研究[J].中国医学物理学杂志,2004,21(6):318-320. 被引量:1
  • 2沈捷,张福泉,邱杰,胡克,张明杰.应用EPID对放射治疗摆位误差的研究[J].中华放射医学与防护杂志,2006,26(3):308-308. 被引量:39
  • 3鞠忠建,王运来,马林,徐寿平,戴相昆,王连元.用二维电离室矩阵验证多叶准直器叶片到位精度[J].中华放射肿瘤学杂志,2006,15(4):335-338. 被引量:18
  • 4潘才住,潘建基,陈传本,张秀春,李金莲.电子射野影像系统测量鼻咽癌调强放疗的摆位误差[J].福建医科大学学报,2007,41(1):49-52. 被引量:24
  • 5Woo MK,Nico A.Impact of multileaf collimator leaf positioning accuracy on intensity modulation radiation therapy quality assurance ion chamber measurements[J].Med Phys,2005,32(5):1440-1445.
  • 6Parent L,Seco J,Evans PM,et al.Evaluation of two methods of predicting MLC leaf positions using EPID measurements[J].Med Phys,2006,33(9):3174-3182.
  • 7Baker SJ,Budgell GJ,MacKay RI.Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration[J].Phys Med Biol,2005,50(7):1377-1392.
  • 8Prescribing recording and reporting photon beam therapy(Supplement to ICRU report 50)[R]. IGRU. Report no. 62,1999.
  • 9Van Lin EN, Van derVightL, Huizenga H, et al. Set-up improvement in head and neck radiotherapy using a 3D off-line EPID-based correction protocol and a customized head and neck support [J]. Radiotherapy Oncol, 2003, 68(2):137-148.
  • 10TateokaK OouchiA NakataK等.液体电离室型电子射野影像装置在剂量验证中的应用.放射物理技术杂志,2008,1(2):171-177.

共引文献46

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部