二维半导体阵列在调强放疗平面剂量验证中的应用 被引量:1

On Application of 2D Semiconductor Array in the Verification of Planar Does in IMRT
下载PDF
导出
摘要 目的:二维半导体阵列在调强放疗中剂量验证中的作用。方法:在二维半导体阵列探头上附加5cm等效水模体,对96例患者的IMRT计划以Map CHECK 2二维半导体阵列进行计量学验证,对比分析。结果:Map CHECK 2二维半导体阵列具有较好的测量重复性与稳定性最大偏差不足0.4%,而在剂量线性验证中表现较好,剂量与距离误差为3%/3mm与2%/2mm条件下,DTA标准的绝对剂量为99.46%与91.11%,相对剂量为98.21%与96.73%,γ标准的绝对剂量为99.94%与92.93%,相对剂量为99.16%与98.27%。结论:Map CHECK 2二维半导体阵列能够准确有效验证强调放疗中的绝对剂量与相对剂量,同时就方便简单的优势,适合推广应用。 Objective:To investigate the application of 2D semiconductor array in the verification of Planar does in IMRT. Methods:Attach 5cm of equivalent water phantom to the probes of 2D semiconductor array. Verify the IMRT plan of 96 patients with the 2D semiconductor array Map CHECK 2 metrologieally and make eontrastive analysis. Results. The repeatability and stability of the measurement of 2D semiconductor array Map CHECK 2 was significant,the maximum deviation was less than 0.4% ,and it also worked well in the linear verification of does. When does and range error was 3 %/3turn and 20%0/2mm ,the absolute does DTA was 99.46% and 91.11% respectively ,and relative does was 98.21% and 96.73% ,the absolute does of γ standard was 99. 94% and 92. 93% respectively, and the relative does was 99. 16% and 98. 27%. Conclusion:2D semiconductor array Map CHECK 2 can verify the absolute does and the relative does in IM- RT definitely and effectively,and it is quite convenient and simple,it is worthy of promotion and application.
出处 《数理医药学杂志》 2015年第7期1077-1078,共2页 Journal of Mathematical Medicine
关键词 二维半导体阵列 MAP CHECK 调强放疗 平面剂量验证 2D semiconductor array Map CHECK intensity modulated radiotherapy(IMRT) verification of planar does
  • 相关文献

参考文献5

二级参考文献40

  • 1张惠玲,李勤,全红,韩军.MatriXX在IMRT剂量验证中的应用[J].中国医学物理学杂志,2011,28(2):2455-2459. 被引量:16
  • 2马金利,蒋国樑.调强适形放射治疗剂量学验证体系的建立[J].中国肿瘤,2004,13(8):485-489. 被引量:8
  • 3徐建一,邹绪春,张友德,戴黎明.46道医用剂量分布仪的研制[J].中国测试技术,2005,31(5):134-137. 被引量:1
  • 4Herzen J, Todorovic M, Cremers F, et al. Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine[J]. Phys. Med. Biol. 2007,52(4):1197-1208.
  • 5Meeks SL, Harmon JJ, Langen KM, et al. Performance characterization of megavoltage computed tomagraphy imaging on a helical tomotherapy unit [J]. Med Phys, 2005,32(8):2673-2681.
  • 6Nelms BE,Simon JA. A survey on planar IMRT QA analysis [J]. J Appl Clin Med Phys, 2007,8:876-890.
  • 7Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method [J]. Med Phys, 2003,30:2455-2464.
  • 8Low DA, Harms WB, Mutic S, et al. A technique for the quantitative evaluation of dose distributions [J]. Med Phys, 1998,25(5):656-661.
  • 9Buonamici FB, Compagnucci A, Marrazzo L, et al. An intercomparison between film dosimetry and diode matrix for IMRT quality assurance [J]. Med. Phys. 2007,34(4):1372-1379.
  • 10Jursinie PA, Nelms BE. A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery [J]. Med. Phys. 2003,30(5):870-879.

共引文献41

同被引文献11

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部