期刊文献+

碲化镉量子点荧光猝灭法测定镍 被引量:3

Determination of nickel by cadmium telluride quantum dots fluorescence quenching method
下载PDF
导出
摘要 基于镍离子对碲化镉(CdTe)量子点荧光具有猝灭作用,建立了对镍离子的定量检测方法。实验表明,在5mL比色管中依次加入0.5mL 1.5×10^-4 mol/L CdTe量子点溶液(以Cd^2+计)、50μL不同镍离子浓度的标准溶液,用pH 10.0硼砂缓冲溶液定容,室温下反应10min,以400nm为激发波长,在荧光发射波长为608nm处测定其相对荧光强度,CdTe量子点荧光衰减程度与镍离子浓度在2.0×10^-7-7.8×10^-5 mol/L范围内呈线性关系,其线性回归方程为F0/F=1.008 9+0.030 8ρ(μmol/L),相关系数r=0.998 7,检出限为1.5×10^-7 mol/L。方法用于水样中镍离子的测定,测得结果与原子吸收光谱法(AAS)一致,相对标准偏差(RSD,n=5)为5.6%。 A quantitative method for determination of nickel was established based on the fluorescence quenching effect of nickel ion on cadmium telluride(CdTe)quantum dots.The experimental results indicated that the fluorescence attenuation degree of CdTe quantum dots was linear to the concentration of nickel ions in range of 2.0×10^-7-7.8×10^-5 mol/L under the following conditions:0.5mL of 1.5×10^-4 mol/L CdTe quantum dots solution(in Cd^2+)and 50μL nickel standard solution with different concentrations were added into 5mL colorimetric tube successively;the mixed solution was diluted to the mark with borax buffer solution at pH 10.0;after reaction at room temperature for 10 min,the relative fluorescence intensity was determined at fluorescence emission wavelength of 608 nm with 400 nm as excitation wavelength.The equation of linear regression was F0/F=1.008 9+0.030 8ρ(μmol/L)with correlation coefficient of r=0.998 7.The detection limit was 1.5×10^-7 mol/L.The proposed method was applied to the determination of nickel ions in water sample.The results were consistent with those obtained by atomic absorption spectrometry(AAS).The relative standard deviation(RSD,n=5)was 5.6%.
出处 《冶金分析》 CAS CSCD 北大核心 2015年第6期49-52,共4页 Metallurgical Analysis
基金 国家级大学生创新创业训练项目(20141012499) 山西省大学生创新创业训练项目(2014382)
关键词 碲化镉(CdTe)量子点 镍(Ⅱ) 荧光猝灭法 cadmium telluride(CdTe)quantum dot nickel(Ⅱ) fluorescence quenching method
  • 相关文献

参考文献12

二级参考文献148

共引文献77

同被引文献36

引证文献3

二级引证文献4

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部