摘要
目的采用高效液相色谱法测定盐酸氨溴索口服溶液中杂质B[反式-4-(6,8-二溴-四氢喹唑啉环己醇)盐酸盐]与杂质E(2-氨基-3,5-二溴苯甲醛)的含量。方法色谱柱为Ultimate XB-C18(250 mm×4.6mm,5μm),以0.01 mol·L^-1磷酸氢二铵(用磷酸调节pH至7.0)-乙腈(50:50)为流动相,检测波长为250 nm(检测杂质B)与238 nm(检测杂质E),柱温为40℃,流速为1 mL·min^-1,进样量为20μL。结果杂质B进样浓度在1.215-19.44μg·mL^-1与峰面积呈良好的线性关系(r=0.999 9),检测限为2.43 ng,平均回收率均在100.0%-100.6%(n=3),RSD均〈1.6%;杂质E进样浓度在1.032-16.512μg·mL^-1与峰面积呈良好的线性关系(r=0.999 4),检测限为0.825 ng;平均回收率均在95.9%-100.1%(n=3),RSD均〈0.65%。结论该方法简单可行、灵敏度高,适用于盐酸氨溴索口服液中有关物质的质量控制。
Objective To establish an HPLC method to determine the content of impurity B [trans-4-(6, 8-dibromo-1, 4-dihydroquinazolin-3(2H)-yl) cyclohexanol hydrochloride] and impurity E(2-amino-3, 5-dibromobenzaldehyde) in Ambroxol hydrochlorid oral solution. Methods Ultimate XB-C18 column was used with the mobile phase composed of acetonitrile-0.01 mol·L^- 1 ammonium hydrogen phosphate(50 :50). The detection wavelength was 250 nm for impurity B and 238 nm for impurity E. The column temperature was 40 ℃, and the flow rate was 1.0 mL·min^- 1, the sample size was 20 μL. Results The calibration curve of impurity B and impurity E had good linearity at 1.215-19.44 μg·mL^- 1(r = 0.999 9) and 1.032- 16.512 μg·mL^- 1(r = 0.999 4), the minimum detection limit of impurity B and impurity E were 2.43 ng and 0.825 ng. The average recoveries of impurity B and impurity E was at 100.0%- 100.6%(RSD 〈1.6%) and 95.9%- 100.1%(RSD〈 0.65%). Conclusion The method is simple and highly sensitive and suitable for the quality control of related substances in Ambroxol hydrochloride oral solution.
出处
《中南药学》
CAS
2015年第6期638-641,共4页
Central South Pharmacy