期刊文献+

医用电气设备听觉报警信号特征检测方法研究 被引量:1

Research on the Detection Methodology for Audible Alarming Signals of Medical Electrical Equipment
下载PDF
导出
摘要 目的探讨医用电气设备听觉报警信号特征的检测方法。方法采用噪声频谱分析仪配合高带宽泰克数字示波器与B&K耳声分析系统配合声音分析仪两种测试平台,对多参数监护仪进行听觉报警信号检测。结果在相同环境条件下,两种测试平台所得结果差别不大;相较于普通实验室,两种测试平台在消音室均得到更准确的测试结果。结论对于听觉报警信号特征的检测在消声室中干扰较少,检测结果更为准确。 Objective To explore the detection methodology for audible alarming signals of medical electrical equipment. Methods Two testing platforms were adopted to detect the audible alarming signals of the multi-parameter monitor, including the combination of the noise spectrum analyzer and high bandwidth Tektronix digital oscilloscope as well as the combination of the ear acoustic analysis system and the voice signal analyzer. Results Noobvious differences were seen between the testing results on two platforms under the same environmental conditions. In contrast with the common laboratory, application of the two platforms inthe anechoic chamber could achieve more accurate testing results. Conclusion Detection of the features of audible alarming signals in the anechoic chamber had less interference and could obtain more accurate testing results.
出处 《中国医疗设备》 2015年第7期22-24,12,共4页 China Medical Devices
关键词 医用电气设备 听觉报警信号特征 消声室 medical electrical equipment features of audible alarming signals anechoic chambers
  • 相关文献

参考文献8

  • 1lEC61}(iltl 1 :O03,Mcdical electrical equipmcn{ Part 1 8:General requirements (or basic safety and csscntiR perf'ormance Collateral standard:GCl:cral requirements,tests H]d guidance fbr alarm systems m medical electrical equipment and medical electrical systems[S].
  • 2YY0709-2009,医用电气设备第1-8部分:安全通用要求并列标准通用要求,医用电气设备和医用电气系统中报警系统的测试和指南[S].
  • 3齐丽晶,张海明.解读YY0709-2009医用报警系统的测试[J].中国医疗器械杂志,2011,35(4):291-293. 被引量:7
  • 4齐丽晶,张海明,张赟.YY0709—2009医用报警系统检测方法研究[J].医疗卫生装备,2014,35(2):125-126. 被引量:2
  • 5GB/T378522/1EC61672-2:2013医用保警声学声级计第1部分:型式评价试验[S].
  • 6GB50800-2012,中华人民共和国国家标准:消声室和半消声室技术规范[S].
  • 7纳特.西尔弗信号与噪声[M].北京:中信出版社2013.
  • 8刘翔,葛斌,刘锦初.各类医用电气设备标准间电气安全指标的差异[J].中国医疗设备,2013,28(12):120-121. 被引量:3

二级参考文献22

  • 1李永安.光电池光特性的电子测量[J].物理实验,2005,25(6):9-11. 被引量:9
  • 2YY0709.2009医用电气设备第1-8部分:安全通用要求并列标准:通用要求医用电气设备和医用电气系统中报警系统的测试和指南.
  • 3Dan O' Brien From Mallory Sonalert Products, Inc. Using audible alarms in medical equipment (IEC 60601-1-8) http://www.mallory- sonalert.com/Articles/TechArticles/Audible%20Alarms%20in%20 Medical%20Equipment.pdf.
  • 4Lau, N. A signal processing approach to auditory alarm design and evaluation: An analysis of melodic medical equipment a/arm[A]. Systems, Man and Cybernetics, 2007. ISIC. IEEE International Conference[C], 7-10 Oct. 2007.
  • 5P. M. Sanderson, A. Wee and P. Lacherez, Learnability and discriminability of melodic equipment alarms[J], Anaesthesia, 2006. 61: 142-147.
  • 6P. M. Sanderson. The multimodal world of medical monitoring displays[J]. Applied Ergonomics, 2006, 37: 501-512.
  • 7国家食品药品监督管理局.YY/T0841-2O11医用电气设备周期性测试和修理后测试[M],1版.北京:中国标准出版社,2011.
  • 8YY0709-2009医用电气设备第1-8部分:安全通用要求并列标准:通用要求,医用电气设备和医用电气系统中报警系统的测试和指南[S].
  • 9IEC 60601-1-8:2006 Medical electrical equipment Part 1-8: General requirements for basic safety and essential performance Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems[S].
  • 10王娟,周万珍,李媛,等.C#.net与Matlab混合编程的快速实现[J].科技咨询,2006(33):60-61.