期刊文献+

气相色谱-质谱法同时测定土壤中30种农药残留 被引量:4

Simultaneous Determination of 30Pesticide Residues in Soil by GC-MS
下载PDF
导出
摘要 采用气相色谱-质谱法同时测定土壤中30种农药残留。样品经过丙酮-正己烷(1+1)混合溶液均质提取,再用正己烷进行萃取,弗罗里硅土柱净化后,采用HP-5MS色谱柱分离,选用电子轰击离子源和选择离子监测模式进行质谱分析,外标法定量。30种农药的线性范围为0.1-1.0mg·L^-1,检出限(3S/N)为3.8-9.6μg·kg^-1。加标回收率在75.8%-103%之间,测定值的相对标准偏差(n=6)在5.6%-9.7%之间。 A GC-MS method was developed for the determination of 30 pesticide residues in soil.The sample was extracted with acetonen-hexane(1+1)and hexane in succession.Then the extract was cleaned up on a florisil cartridge and separated on a HP-5MS column.EI-SIM was adopted in MS and external standard was used for quantification.The linearity range of 30 pesticide residues was 0.1-1.0mg·L^-1 with detection limits(3S/N)in the range of 3.8-9.6μg·kg^-1.The recovery rates measured by standard addition method were in the range of75.8%-103% and RSD(n=6)were in the range of 5.6%-9.7%.
出处 《理化检验(化学分册)》 CAS CSCD 北大核心 2015年第7期925-929,共5页 Physical Testing and Chemical Analysis(Part B:Chemical Analysis)
基金 国家质检总局科研项目(2012IK176) 山东检验检疫局科研项目(SK201353)
关键词 气相色谱-质谱法 农药残留 土壤 GC-MS Pesticide residues Soil
  • 相关文献

参考文献15

二级参考文献154

共引文献270

同被引文献47

引证文献4

二级引证文献32

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部