基于迭代重组的双源CT腰椎低剂量扫描可行性研究 被引量:6

The Application of Low-Dose Scan in Lumbar Vertebrae Based on Iterative Image Reconstruction of Dual-Source CT
原文传递
导出
摘要 目的对比双源CT腰椎低剂量扫描基于原始数据的迭代重组(SAFIRE)和常规剂量扫描滤波反投影重组(FBP)的辐射剂量、图像质量,探讨腰椎CT低剂量扫描的可行性。方法将120例拟行腰椎CT扫描患者随机分为A、B、C三组,A组为常规剂量组,采用FBP重组;B组为低管电流组,C组为低管电压组,B、C组预计降低50%辐射剂量且均采用SAFIRE重组。记录并计算每例患者辐射剂量指标[CT剂量指数(CTDIvol)、剂量长度乘积(DLP)、有效辐射剂量(ED)],对比L4~5水平各组图像各感兴趣区(ROI)客观评价指标:信噪比(SNR)、椎间盘与硬膜囊对比噪声比(CNR)和主观评分,行统计学分析。结果 B组辐射剂量较A组平均降低约62%且各感兴趣区SNR及CNR与A组无显著统计学差异(P〉0.05);C组辐射剂量较A组平均降低约70%,除L5骨松质之外,余ROI客观图像质量均低于A组(P〈0.05);除椎小关节,A、B组图像主观评分无显著性差异;观察者间一致性很好。结论与常规剂量FBP相比,基于SAFIRE的腰椎CT低剂量扫描(管电流降低至常规50%)结合自动管电流调节技术(Care Dose)在明显降低辐射剂量的同时不影响图像质量,推荐将其作为腰椎疾病筛查方法。 Objective To compare radiation dose and image quality of the lumbar vertebrae produced by low-dose dualsource CT scan,Sinogram Affirmed iterative reconstruction(SAFIRE) with conventional dose scan,filter back-projection(FBP) reconstruction,to study the feasibility of low dose scanning of lumbar CT.Methods 120 patients underwent lumbar CT scan were randomly divided into A,B,C groups,group A is the conventional dose group,using FBP reconstruction; group B is the lower tube current group,group C is the lower tube voltage group,B and C groups were treated with SAFIRE reconstruction which intended end radiation dose was 50% less than the group A.Recorded and calculated radiation dose indicator for each case(CTDILOV,DLP,ED),compare the objective evaluation indexes:the signal-to-noise ratio(SNR),the contrast-to-noise ratio between intervertebral disc and the dural sac(D-D CNR) and the subjective scores of different reconstruction image of the every interest(ROI) at L4 /5level,then process statistical analysis.Results The average radiation dose of group B decreased 62% than group A,the SNR of every ROI and DNR had no significant difference compared with group A(P〈0.05).Compared with group A,the average radiation dose of group C reduced about 70%,in addition to L5 vertebrae body,all ROI objective image quality is lower than that of group A(P〈0.05); in addition to the vertebral small joint,A and B groups has no significant difference in subjective image score; All radiologists agreement performed very well.Conclusion Compared with conventional dose with FBP,Lumbar CT low dose scanning based on SAFIRE(the tube current reducing 50% compared with conventional) combined with automatic tube current modulation technique can significantly reduce the radiation dose,but not affect image quality,so it is recommended as a screening program for lumbar vertebral disease.
出处 《临床放射学杂志》 CSCD 北大核心 2015年第7期1166-1170,共5页 Journal of Clinical Radiology
关键词 双源计算机断层扫描 原始数据迭代重建 滤波反投影重建 腰椎 辐射剂量 图像质量 Dual-source CT Sinogram affirmed iterative reconstruction Filter back-projection Lumbar vertebrae Radiation Image quality
  • 相关文献

参考文献14

  • 1Sookpeng S, Martin C J, Gentle DJ. Comparison of different phantom designs for CT scanner automatic tube current modulation system tests[ J ]. J Radiol Prot ,2013,4:735-761.
  • 2Wang Z,Huang Z,Zhang L,et at. Low dose reconstruction algorithm for differentikl phase contrast imaging [ J ]. J Xray Sci Technol, 2011,3:403-415.
  • 3McNitt-Gray MF. AAPM/RSNA Physics Tutorial for Residents: Top- ics in CT. Radiation dose in CT[J]. Radiographics,2002,6: 1541- 1553.
  • 4Yang CH. , Wu TH, Chiou YY, et al. Imaging quality and diagnostic reliability of low-dose computed tomography lumbar spine for evalua- ting patients with spinal disorders[ J ]. The Spine Journal,2014,14: 2682-2690.
  • 5Bongartz G, Golding SJ, Jurik AG, et al. European guidelines on quality criteria for computed Tomography_Eur 16262 EN. Luxem- bourg: European Commission ,2000.
  • 6Zondervan RL, Hahn PF, Sadow CA, et al. Body CT scanning in young adults: examination indications, patient outcomes, and risk of radiation-induced cancer[ J ], Radiology,2013,2:460-469.
  • 7Hu XH, Ding XF,Wu RZ. , et al. Radiation dose of non-enhanced chest CT earl be reduced 40% by using iterative reeonstruetion in image space[J]. Clin Radiol,2011,11:1023-1029.
  • 8王艳,史大鹏,朱绍成,吴青霞,孙明华.比较以自适应性统计迭代重建技术和滤过反投影重建的低剂量腹部CT的图像质量[J].中国医学影像技术,2012,28(10):1902-1905. 被引量:39
  • 9Chen CM, Chu SY, Hsu MY, et al. Low-tube-vohage ( 80 kVp) CT aortography using 320-row volume CT with adaptive iterative recon- struction : lower contrast medium and radiation dose [ J ]. Eur Radi- ol,2014,2:460-468.
  • 10Meyer M, Klein SA, Brix G, et al. Whole-body CT for lymphoma stag- ing: feasibility of halving radiation dose and risk by iterative image reconstruction[ J]. Eur J Radiol,2014,2:315-321.

二级参考文献15

  • 1刘勇,张华.ALARA理论及CT低剂量应用研究进展[J].中国医学影像技术,2008,24(S1):17-20. 被引量:28
  • 2Guillerman RP. Newer CT applications and their alterna- tives.- what is appropriate in children? [J]. Pediatr Radiol, 2011, 41(2) : 534-548.
  • 3Wallace AB, Goergen SK, Schick D, et al. Multidetector CT dose: clinical practice improvement strategies from a success- ful optimization program [J]. J Am Coil Radiol, 2010, 7(8) : 614-624.
  • 4Kalender W (2000) Computed tomography: fundamentals, system technology, image quality, applications, 1st edn. Wi- ley VCH ,Weinheim.
  • 5Wang G, Yu H, De Man B. An outlook on x-ray CT re- search and development [J]. Med Phys, 2008, 35(3) : 1051- 1064.
  • 6Antonio Moscariello, Richard AP, Takx U. Joseph schoepf, Matthias renker, Peter L. Zwerner. Coronary CT angiogra- phy: image quality, diagnostic accuracy,and potential for ra- diation dose reduction using a novel iterative image recon- struction technique-comparison with traditional filtered back projection [J]. European Radiology, 2011, 21(10): 2130- 2138.
  • 7Anna Winklehner, Christoph Karlo, Gilbert Puippe, Bern- hard Schmidt, et al. Raw data based iterative reconstruction in body CTA: evaluation of radiation dose saving potential [J]. European Radiology, 2011, 21(12): 2521-2526.
  • 8Stephan Baumueller, Anna Winklehner, Christoph Karlo, Robert Goetti, Thomas Flohr, et al. Low-dose CT of the lung: potential value of iterative reconstructions [J]. Euro- pean Radiology, 2012, 22(12): 2597-2606.
  • 9Ullrich Ebersberger, Francesco Tricarico U, Joseph Schoep{, Philipp Blanke, et al. CT evaluation of coronary artery stents with iterative image reconstruction: improvements in image quality and potential for radiation dose reduction [J]. Euro- pean Radiology, 2013, 23(1) : 125-132.
  • 10S.Singh,M.K.Kalra,M.A.Moore,R.Shailam,B.Liu,T.L.Toth,秦乃姗.适应患儿大小、临床指征和既往检查次数的小儿CT扫描参数降低剂量及其依从性的研究[J].国际医学放射学杂志,2009,32(5):489-489. 被引量:42

共引文献45

同被引文献37

引证文献6

二级引证文献21

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部