活髓状态对电子根尖定位仪测量准确性影响的研究 被引量:4

THE STUDY ON VITAL PULP EFFECT ON THE ACCURACY OF ELECTRONIC APEX LOCATORS
下载PDF
导出
摘要 目的:在临床根管治疗过程中,分析活髓状态对DENTAPORT ZX,RAYPEX 5,PROPEXⅡ三种根尖定位仪测量根管工作长度的准确性的影响。方法:收集因急慢性牙髓炎、牙髓坏死、根尖周炎需行根管治疗的恒牙活髓根管50个。用三种根尖定位仪分别测量根管工作长度并记录数据,再测量并计算根管实际工作长度。比较根管实际工作长度与根尖定位仪所测长度的差值,并进行统计学分析。结果:活髓状态下,DENTAPORT ZX与RAYPEX 5、PROPEXⅡ之间的测量差别有统计学意义,而RAYPEX 5与PROPEXⅡ之间的测量差别无统计学意义。结论:活髓牙DENTAPORT ZX的测量更准确。 Objective: In the process of clinical root canal treatment,analyzed the vital pulp of influence the accuracies of DENTAPORT ZX,RAYPEX 5 and PROPEX Ⅱ for measuring the working length of the root canal. Methods: Collect 50 vital teeth and permanent root canals which need root canal treatment because of the acute and chronic pulpitis,pulp necrosis,and periapical periodontitis.Three electronic apex locator in root canal working length were measured and recorded data. And then Measurement and calculation of root canal working length. Compare the difference between the root canal working length and the measured length of root apex locator,and statistical analysis. Results: although there is no significant difference between RAYPEX 5 and PROPEX measured,the significant difference can be found between DENTAPORT ZX,RAYPEX 5 and PROPEX measured. Conclusion:DENTAPORT ZX is more accurate measurement under the vital teeth status.
作者 王子 牛一山
出处 《内蒙古医科大学学报》 2015年第3期238-241,245,共5页 Journal of Inner Mongolia Medical University
基金 内蒙古自治区卫生厅科研项目(2010037)
关键词 DENTAPORT ZX RAYPEX5 PROPEXⅡ 根管测量仪 根管工作长度 准确性 DENTAPORT ZX RAYPEX5 PROPEXⅡ apex locator work length accuracy
  • 相关文献

参考文献14

二级参考文献66

共引文献75

同被引文献89

  • 1常缨,董文亮.口腔内科门诊老年患者的就诊特点及分析[J].宁夏医科大学学报,2013,35(6):707-709. 被引量:9
  • 2沈家平,吴友农.口腔疾病患者就医观念及行为特点浅析[J].口腔医学,2004,24(5):303-305. 被引量:5
  • 3中华人民共和国统计局.2004-2013年新农合筹资标准与医疗费用支出情况[EB/OL].(2014-09-30)[2015-01-06].http://data.stats.gov.cn/workspace/index?m=hgnd.
  • 4樊明文,周学东.牙体牙髓病学[M].4版.北京:人民卫生出版社,2014:3-68.
  • 5Wonq AW, Zhanq C, Chu CH. A systematic review of nonsurgical single-visit versus multiple-vist endodontic treatment[J].Clin Cosmet Investiq Dent, 2014,6 : 45 -- 56.
  • 6Singh SV,Nikhil V, Singh AV,et aLAn in vivo comparative evaluation to determine the accuracy of working length between radiographic and electronic apex loeators[J].Indian J Dent Res,2012,23(3) 359--362.
  • 7Krishnan IS, Sreedharan S. A comparative evaluation of electronic and radiographic determination of root canal length in primary teeth; an in vitro study[J].Contemp Clin Dent, 2012,3(4) ,416--420.
  • 8Kim YJ, Chandler NP. Determination of working length for teeth with wide or immature apices a review[J].Int Endod J,2013,46(6):483--491.
  • 9Chakravarthy Pishipati KV.An in vitro comparison of propex II apex locator to standard radiographic method [J]. Iran Endod J,2013,8(3):114-117.
  • 10Aydin U,Karataslioglu E, Aksoy F, et al. In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters[J].J Conserv Dent, 2015,18( 1 ) : 66 -- 69.

引证文献4

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部