期刊文献+

便携式铵氮荧光检测仪 被引量:1

Design of Portable Fluorescence Analyzer of Ammonium
下载PDF
导出
摘要 基于铵氮的荧光分析方法研制了一种便携式铵氮荧光检测仪,由样品检测模块、信号放大与压频转换模块、测频模块、数据显示与存储模块和电源模块组成。仪器的基线和恒定信号响应可稳定2 h;铵氮浓度在0-4 000 nmol/L范围内时,其测定结果与通用仪器测定结果有线性相关性;仪器可测定实际水样中的铵氮浓度,与通用仪器的测定结果无显著差异。 A portable fluorescence analyzer of ammonium was developed based on fluorescence analysis method. The analyzer was made of sample detection module,signal amplifying and voltage-frequency converting module,frequency measurement module,data displaying and memory module and power module. Both the baseline and response to a constant signal of the analyzer can maintain stable for 2 hours. The signals of ammonium solutions whose concentration were in the range of 0 - 4 000 nmol / L were separately detected by both the analyzer and a universal apparatus,and the results of the analyzer had a linear correlation with those of the universal apparatus. The analyzer is used to determine ammonium concentration in actual water samples,and the results have no significant differences with those of the universal apparatus.
出处 《仪表技术与传感器》 CSCD 北大核心 2015年第6期31-33,52,共4页 Instrument Technique and Sensor
基金 国家自然科学青年基金项目(41206077)
关键词 便携式 铵氮 荧光 LED 光电二极管 portable ammonium fluorescence LED photodiode
  • 相关文献

参考文献6

二级参考文献54

  • 1乔长安,陈运涛,张富平,李彬.等精度频率测量方法[J].火力与指挥控制,2003,28(z1):61-62. 被引量:8
  • 2Aoki T,Uemura S,Munemori M,1983.Continuous-flowfluorometric-determination of ammonia in water.AnalyticalChemistry,55(9): 1620–1622.
  • 3Camargo J A,Alonso A′,2006.Ecological and toxicological ef-fects of inorganic nitrogen pollution in aquatic ecosystems:A global assessment.Environment International,32(6):831–849.
  • 4de Jong E P,Lucy C A,2006.Low-picomolar limits of detectionusing high-power light-emitting diodes for fluorescence.Analyst,131(5): 664–669.
  • 5Egawa T,Durand J L,Hayden E Y,Rousseau D L,Yeh SR,2009.Design andevaluation of a passive alcove-basedmicrofluidic mixer.Analytical Chemistry,81(4): 1622–1627.
  • 6Fukushi K,Ito H,Kimura K,Yokota K,Saito K,Chayama Ket al.,2006.Determination of ammonium in river waterand sewage samples by capillary zone electrophoresis withdirect UV detection.Journal of Chromatography A,1106(1-2): 61–66.
  • 7Gao D,Wei H B,Guo G S,Lin J M,2010.Microfluidiccell culture and metabolism detection with electrosprayionization quadrupole time-of-flight mass spectrometer.An-alytical Chemistry,82(13): 5679–5685.
  • 8Gray S M,Ellis P S,Grace M R,McKelvie I D,2006.Spectrophotometric determination of ammonia in EstuarineWaters by hybrid reagent-injection gas-diffusion flow anal-ysis.Spectroscopy Letters,39(6): 737–753.
  • 9Hellman A N,Rau K R,Yoon H H,Bae S,Palmer J F,Phillips K Set al.,2007.Laser-induced mixing in microfluidic channels.Analytical Chemistry,79(12): 4484–4492.
  • 10Hisamoto H,Horiuchi T,Tokeshi M,Hibara A,Kitamori T,2001.On-chip integration of neutral ionophore-based ionpair extraction reaction.Analytical Chemistry,73(6): 1382–1386.