OQAS测量散射光和客观视觉质量的可重复性研究 被引量:2

Repeatability of measurement by OQAS for assessment straylight and objective optimal quality
原文传递
导出
摘要 目的 评价OQAS光学质量分析系统对健康人眼、屈光手术术后眼及早期年龄相关性白内障测量客观散射指数(OSI)和MTF截止频率(MTF cut off)、斯特列尔比(SR)等客观视觉质量指标的可重复性.方法 临床病例对照研究.选取2013年3~8月就诊于中国医科大学附属第四医院的50只健康人眼、50只屈光手术(LASIK)术后眼、50只早期年龄相关性白内障眼,使用OQAS测量受试眼的客观散射指数(OSI)、MTF截止频率(MTF cut off)、斯特列尔比(SR),每项指标测量3次,使用组内相关系数(ICC)及变异系数(CV)为统计学指标评估其可重复性.结果 健康人眼组散射光、MTF cut off及SR测量的ICC及CV值分别为0.921、1.21;0.90、0.2;0.87、0.57;屈光手术术后组散射光、MTF cut off及SR测量的ICC及CV值分别为0.97、1.27;0.90、0.78;0.89、1.91;早期白内障组散射光、MTF cut off及SR测量的ICC及CV值分别为0.96、1.54;0.89、0.71;0.86、2.28.屈光手术术后眼及早期白内障眼的散射光较健康人眼增加,MTF cut off及SR较健康人眼降低,差异均有统计学意义(P<0.05).结论 OQAS可以反映客观散射光及视觉质量,在健康人眼、屈光手术术后眼和早期年龄相关性白内障眼中使用重复性均较好。 Objective To assess the repeatability of objective scatter index and other objective optimal quality measured by OQAS doubled-pass system.Methods Fifty healthy subjects,50 post-LASIK eyes and 50 early cataract eyes were recruited in the Eye Hospital of China Medical University from March 2013 to August 2013.The objective scatter index (OSI),MTF cut off and Strehl ratio (SR) of the eyes were measured by OQAS.Three repeated consecutive measurements were completed by the same examiner.Results The intraclass correlation coefficient (ICC) and coefficient of variation of OSI,MTF cut off and SR in healthy eyes were 0.92,1.21;0.90,0.2;0.87,0.57;The intraclass correlation coefficient (ICC) and coefficient of variation of OSI,MTF cut off and SR in post-LASlK eyes were 0.97,1.27;0.90,0.78;0.89,1.91;The intraclass correlation coefficient (ICC) and coefficient of variation of OSI,MTF cut off and SR in early cataract eyes were 0.96,1.54;0.89,0.71;0.86,2.28.The OSI both in post-LASIK eyes and early cataract eyes were larger than healthy eyes,MTF cut off and SR in post-LASIK eyes and early cataract eyes were smaller than healthy eyes,and there were statistically significant differences.Conclusions OQAS can reflect the objective scatter index and other objective optimal quality.The measurements of OSI,MTF cut off and SR by OQAS doubled-pass system has a high reproducibility.
出处 《中国实用眼科杂志》 2015年第7期769-773,共5页 Chinese Journal of Practical Ophthalmology
关键词 OQAS 散射光 视觉质量 重复性 OQAS Scatter light Optimal quality Repeatability
  • 相关文献

参考文献14

  • 1Franssen L, Coppens JE, van den Berg T]. Compensation com- parison method for assessment of retinal straylight. [J].Invest Ophthalmo! Vis Sci,2006,47(2) :768-776.
  • 2Van Rijn LJ,Wilbelm H,Emesz M,et 81. Relation between per- ceived driving disability and scores of vision screening tests. [J]. Br J Ophthalmol, 2002,86(11) : 1262-1264.
  • 3Guber I, Bachmann LM, Guber J, et al.Reproducibility of stray- light measurement by C-Quant for assessment of retinal stray- light using the compensation comparison method. [J].Graefes Arch Clin Exp Ophthalmol,2011,249(9) : 1367-1371.
  • 4Coppens JE, Franssen L, van Rijn I_J, et al.Reliability of the compensation comparison stray-light measurement method. [J].J Biomed Opt,2006,11(3) :34027.
  • 5Kruijt B, Franssen L, Prick L J, et al. Ocular Straylight in Albi-nism.[J].Optom and vision sei,2011,88(5):585-592.
  • 6万修华,张烨,乔利亚,蔡啸谷,谭家璿,李蕾,王宁利.双通道系统视觉质量分析仪评价正常人视觉质量的可重复性研究[J].眼科,2013,22(3):175-179. 被引量:20
  • 7Vilaseca M,Peris E,Pujol J, et al. Intra-%and intersession re- peatability of double-pass instrument.[J].Optom Vis Sci,2010,87 (9) :675-681.
  • 8Saad A, Saab M, Gatinel D. Repeatability of measurements with a double-pass system.[J].J Cataract Refract Surg,2010,36(1): 28-33.
  • 9Shrout PE. Measurement reliability and agreement in psychiatry. [J].Statistical Methods in Medical Research, 1998,7(3):301- 317.
  • 10Van Den Berg TJ, van Rijn LJ, Michael R et al.Straylight Ef- fects with Aging and Lens Extraction. [J].Am J Ophthalmal, 2007,144(3).358-363.

二级参考文献20

  • 1李镒冲,李晓松.两种测量方法定量测量结果的一致性评价[J].现代预防医学,2007,34(17):3263-3266. 被引量:155
  • 2Cheng X, Thibos LN, Bradley A. Estimating visual quality from wavefront aberration measurements. J Refract Surg, 2003, 19: 579-584.
  • 3Diaz-Douton F, Benito A, Pujol J, et al. Comparison of the retinal image quality with a Hartmann-Shack wavefront sensor and a dou- ble-pass instrument. Invest Ophthalmol Vis Sci, 2006, 47: 1710-1716.
  • 4Saad A, Saab M, Gatinel D. Repeatability of measurements with a double-pass system. J Cataract Refract Surg, 2010, 36: 28-33.
  • 5Benito A, Perez GM, Mirabet S, et al. Objective optical assessment of tear-film quality dynamics in normal and mildly symptomatic dry eyes. J Cataract Refract Surg, 2011, 37: 1481-1487.
  • 6Artal P, Benito A, P6rez GM, et al. An objective scatter index based on double-pass retinal images of a point source to classify cataracts. PLoS ONE, 20310, 6: e16823.
  • 7Cuell JL, Pujol J, Arjona M, et al. Optical quality analysis system: instrument for objective clinical evaluation of ocular optical qual- ity. J Cataract Refract Surg, 2004, 30: 1598-1599.
  • 8Vilaseca M, Peris E, Pujol J, et al. Intra- and intersession repeata-bility of double-pass instrument. Optom Vis Sci, 2010, 87: 675-681.
  • 9Martinez-Roda JA, Vilaseca M, Ondategui JC, et al. Optical qual- ity and intraocular scattering in a healthy young population. Clin Exp Optom, 2011, 94: 223-229.
  • 10International Organization for Standardization (ISO). Statistics vo- cabulary and symbols. Geneva, Switzerland: ISO, 1977: ISO 3534.

共引文献19

同被引文献11

引证文献2

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部