期刊文献+

16台便携式血糖仪与生化分析仪检测血糖结果对比分析 被引量:3

Comparative Analysis of Blood Glucose Testing Results of 16 Portable Blood Glucose Meter and Biochemical Analyzer
下载PDF
导出
摘要 目的通过与生化仪血糖检测结果对比分析,评估便携式血糖仪的准确性和精确度。方法选取本院16台德国罗氏血糖仪,检测50份涵盖低中高不同浓度的血糖水平,与生化仪对比,计算偏差及变异系数,评估便携式血糖仪的准确性及精确度。结果 16台罗氏血糖仪中3台精密度不合格,4台准确度不合格,根据回归分析结果,证实4台不合格,不合格率为25.0%。结论加强便携式血糖仪与生化仪的对比校正试验,是保证便携式血糖仪检验质量的重要措施。 Objectives To assess the accuracy and precision of the portable blood glucose meter by comparative analysis of levels of blood glucose with biochemical analyzer and portable blood glucose meter.Methods 50 covering with high concentration and low sugar in the blood specimens were detected with a total of 16 sets of Germany Roche glucose meter selected and biochemical analyzer in our hospital.The deviation and coefficient of variation were compared between two methods.The accuracy and precision of the portable blood glucose meter were assessed.Results About the precision 3sets were not qualified in 16 Roche glucose meters,while about the accuracy 4sets were not qualified.According to the results of regression analysis,4sets were confirmed unqualified.The unqualified rate was 25%.Conclusions Enhancing Contrast and correction test of portable blood glucose meter and biochemical analyzer is an important measure to ensure the quality of the portable blood glucose meter test.
出处 《临床输血与检验》 CAS 2015年第4期305-308,共4页 Journal of Clinical Transfusion and Laboratory Medicine
关键词 便携式血糖仪 生化分析仪 对比分析 Portable blood glucose meter Biochemical analyzer Comparative analysis
  • 相关文献

参考文献10

二级参考文献32

  • 1张家庆.血糖及其他体液葡萄糖测定进展[J].中华内分泌代谢杂志,2003,19(4):333-335. 被引量:59
  • 2池胜英,陈筱菲,徐克,陈晓东,袁谦.33台快速血糖仪调查结果分析[J].临床检验杂志,2005,23(2):160-160. 被引量:28
  • 3池胜英,袁谦,陈筱菲,陈晓东,胡朋华.医院内快速血糖检测质量管理的实施[J].中华检验医学杂志,2006,29(6):565-566. 被引量:50
  • 4卫生部办公厅.医疗机构便携式血糖检测仪管理和临床操作规范(试行)[Z].2010-12-30.
  • 5Whifley RJ,Santrach PJ,Phillips DL.Establishing a quality management system for unit use testing based on NCCLS proposed guideline(EP18-P)[J].Clin Chim Acta,2001,307(12):145-149.
  • 6Bezdek J. Pattern Recognition with Fuzzy Objective Function Algorithms. New York: Plenum, 1981
  • 7BG/T19634-2005.体外诊断检验系统自测用血糖监测系统通用技术条件[S].中国国家标准化管理委员会,2005.
  • 8Nationalcommittee for Clinical Laboratory StandardsNCCLS,2002:524-526.
  • 9WS/T226-2002.便携式血糖仪血液葡萄糖测定指南[s].全国临床检验标准委员会,2002.
  • 10Khan AI, Vasquez Y, Gray J, et al. The variability of results between point-of-care testing glucose meters and the central laboratory analyzer[J]. Arch Pathol Lab Med,2006,130(10) : 1527-1532.