摘要 将电喷雾解吸电离与常压离子迁移率谱进行联用,通过在离子迁移率谱仪漂移管进样口前端设置与进样口同电位的辅助电极提升了检测信号强度及稳定性。利用该系统对玻片表面吸附的美莎芬样品进行直接检测,得到的美莎芬约化迁移率为1.29 cm2/(V·s)。同时研究了电喷雾解吸电离源参数对被测物特征峰强度的影响,发现该系统可在较低的雾化气气压下获得较大的特征峰强度。在优化的参数下对玻片表面不同量的美沙芬样品进行直接检测,根据10倍噪声计算系统对美沙芬的检测限为1.6 ng/cm2,表明电喷雾解吸电离常压离子迁移率谱系统具有对物体表面痕量物质进行直接检测的应用潜力。 Desorption electrospray ionization (DESI) source was coupled to an atmospheric pressure ion mobili- ty spectrometer (IMS), the signal intensity was improved by attaching an auxiliary electrode with fixed voltage in front of the ion inlet of IMS tube. Dextromethorphan (DXM) on glass was detected directly by the DESI-IMS. The reduced mobility of DXM measured by the DESI-IMS is 1.29 cm2/( V ·s). Experimental parameters of DESI were optimized for rapid and sensitive detection of DXM. Strong signal is obtained by the DESI-IMS under low nebulizing gas pressure of DESI. The calculated detection limit of the DESI-IMS is 1.6 ng/cm2. The method has the potential for rapid detection of trace substance on solid surface.
出处 《科学技术与工程》 北大核心 2015年第21期28-31,37,共5页 Science Technology and Engineering
基金 国家自然科学基金(61471341)资助
关键词 电喷雾解吸电离 离子迁移率谱 联用 辅助电极 美沙芬 desorption electrospray ionization ion mobility spectrometry coupling auxiliary eleetrode dextromethorphan
  • 相关文献

参考文献9

  • 1Borsdorf H, Eiceman G A. Ion mobility spectrometry: principles and applications. Applied Spectroscopy Reviews, 2006, 41 ( 4 ) : 323-375.
  • 2刘林,花榕,王姜,石磊,李金英,周书民,陈焕文.离子迁移谱仪的电离技术研究进展[J].质谱学报,2014,35(2):97-107. 被引量:5
  • 3Takats Z, Wiseman Jm Fau-Gologan B, Gologan B Fau-Cooks R G, et al. Mass spectrometry sampling under ambient conditions with de- sorption electrospray ionization. Science,306(5695) :471-473.
  • 4Nefliu M, Smith J, Venter A, et al. Internal energy distributions in desorption electrospray ionization (DESI). J Am Soc Mass Spectrom, 2008 , 19 ( 3 ) : 420-427.
  • 5Roscioli K M, Tufariello J A, Zhang X, et al. Desorption electros- pray ionization (DESI) with atmospheric pressure ion mobility spec- trometry for drug detection. Analyst. 2014, 139(7) : 1740-1750.
  • 6李泳明,高晓光,李建平,贾建,李艳志,何秀丽.用于美沙芬快速检测的电喷雾离化离子迁移率谱技术研究[J].仪表技术与传感器,2014(9):99-101. 被引量:2
  • 7Wu C, Siems W F, Hill H H. Secondary electrospray ionization ion mobility spectrometry/mass spectrometry of illicit drugs. Analytical Chemistry, 2000 , 72 (2) : 396-403.
  • 8Mirmahdieh S, Khayamian T, Saraji M. Analysis of dextromethor- phan and pseudoephedrine in human plasma and urine samples using hollow fiber-based liquid-liquid-liquid microextraction and corona dis- charge ion mobility spectrometry. Microchimica Acta, 2012, 176 (3-4) : 471-478.
  • 9Takats Z, Wiseman J M, Cooks R G. Ambient mass spectrometry using desorption elcctrospray ionization (DESI): instrumentation, mechanisms and applications in forensics, chemistry, and biology. Journal of Mass Spectrometry, 2005 , 40(10) : 1261-1275.

二级参考文献85