TEG5000血栓弹力图仪性能评价 被引量:7

Evaluation of the performance of thrombelastography TEG5000
原文传递
导出
摘要 目的评估分析TEG5000血栓弹力图仪(TEG仪)主要参数性能,以验证其是否符合临床实验室的检测要求。方法按标准操作程序(SOP)对TEG仪的主要参数,如从血样开始检测至描记图幅度达2 mm所需的时间(R时间,R)、从R时间终点至描记幅度达20 mm所需时间(K时间,K)、从血凝块形成点至描记图最大曲线弧度作切线与水平线的夹角(α角度,α)、描记图上的最大振幅即最大切应力系数(MA)、凝血综合指数(CI)、任一时刻描记图曲线两点间的距离(A)和血凝块强度(G)等进行重复性、线性、可比性试验,同时观测其对血小板(PLT)和纤维蛋白原(Fbg)的影响。结果重复测定全血标本R、K、MA、CI等主要参数的CV范围为1.8%~11.1%;线性试验R等的相关系数(r)绝对值均〉0.95,表明仪器的实测值和理论值具有良好的相关性;R和MA与PT、APTT、TT和Fbg呈正相关(P〈0.01)。PLT和Fbg的量对血栓弹力图各参数的影响不同,根据参数的变化可以看出标本中PLT和Fbg的变化。结论 TEG仪检测主要参数性能符合要求,可推广应用于临床。但应注意PLT和Fbg对其测定结果的影响。 Objective To evaluate the main performance parameters of thromboelastography TEG5000,and to test if it meets the examination requirements of clinical laboratory test. Methods According to standard operational procedure( SOP),the main parameters of the TEG were detected from the blood samples to the time reaching to the 2 mm( R) required time( R time,R),the time from the end of R to the graph amplitude reaching 20 mm( K time,K),from the blood clot formation point to the maximal curve radian as the angle of tangent and horizontal line( α angle,α),the maximal amplitude,namely the maximal coefficient of the shear stress( MA),coagulation index( CI),the distance between two point( A) at any time in the graph,and blood clot strength( G) were tested to show the repeatability,liner and comparability. The influence of platelet( PLT) and fibrinogen( Fbg) were also observed at the same time. Results The main parameters of K,MA,CI and R of CV were in the range of1. 8%- 11. 1%. The absolute value of the correlation coefficient( r) of linear test R was more than 0. 95,showing that the instrument has a good correlation with the measured value and theoretical value. The correlation between R,MA and PT,APTT,TT and Fbg was positive(P〈0. 01). The influences of platelets and fibrinogen on thromboelastography parameters were different,and the changes of PLT and Fbg in samples could be identified according to the parameters changing. Conclusion The main performance parameters of TEG meets the requirement of clinical work,which can be promoted. But the influence effect by PLT and Fbg should also be concerned.
出处 《中国卫生检验杂志》 CAS 2015年第15期2573-2575,共3页 Chinese Journal of Health Laboratory Technology
基金 贵州省科学技术基金项目(黔科合SY20103129)
关键词 TEG5000血栓弹力图仪 性能评价 血液凝固 Thromboelastography TEG5000 Performance evaluation Blood coagulation
  • 相关文献

参考文献10

二级参考文献32

  • 1曹忠平,陶国才,刘怀琼.15例肝移植患者术中凝血功能改变及常规凝血监测和血栓弹性图监测指标相关性研究[J].临床麻醉学杂志,2003,19(12):713-714. 被引量:7
  • 2陈秉学,中华麻醉学杂志,1998年,18卷,209页
  • 3Mc Nicol P L,Anaesth Intens Care,1994年,22卷,659页
  • 4Kang Y,Anesth Analg,1985年,64卷,888页
  • 5Malett SV,Cox DJ.Thromboelastograph.Br J Anaesth,1992,69:307-313.
  • 6Nascimento B, A1 Mahoos M, Callum J, et al. Vitamin K-depend- ent coagulation factor deficiency in trauma : A comparative analysis between international normalized ratio and thromboelastography [J]. Transfusion, 2012,52(1) : 7-13.
  • 7Agren A,Wikman AT. Holmstrom M, et al. Thromboelastogra- phy(TEG) compared to conventional coagulation tests in surgical patients: A laboratory evaluation[J]. Seand J Clin Lab Invest, 2013,73(3) : 214-220.
  • 8De Pietri L,Montalti R,Begliomini B,et al. Thromboelastographic changes in liver and pancreatic cancer surgery: hypercoagulability, hypocoagulability or normocoagulability? [J]. EJA, 2010,27 (7) : 608-616.
  • 9Sankarankutty A, Nascimento B, da Luz L T, et al. TEG and ROTEM in trauma: Similar test but different results? [J]. World J Emerg Surg,2012,7(Suppl 1) :S3.
  • 10Hagemo JS, Naess PA, Johansson P, et al. Evaluation of TEG and RoTEM inter-changeability in trauma patients[J]. Injury, 2013,44(5) :600-605.

共引文献82

同被引文献48

  • 1王梅,庄逢源,田天.血栓弹力图的数学模型[J].中国生物医学工程学报,1989,8(3):131-136. 被引量:2
  • 2王红,张淑文,任爱民,张丽霞,王超,黄樱,苏艳丽,王宝恩.重度脓毒症凝血功能紊乱与病情严重度及预后的关系[J].中华急诊医学杂志,2005,14(10):804-806. 被引量:54
  • 3毕波,吕元.定量检测方法学性能验证的系统设计[J].中华检验医学杂志,2007,30(2):143-145. 被引量:153
  • 4林孟相,魏宏建,付春来,蔡和伦.脓毒症患者血小板膜糖蛋白CD62p和CD63的表达及其临床意义[J].中国呼吸与危重监护杂志,2007,6(3):205-207. 被引量:9
  • 5KANG Y G, MARTIN D J, MARQUEZ J, et al. Intraoper-ative changes in blood coagulation and thrombelasto- graphic monitoring in liver transplantation [ J ]. Anesthesia and analgesia, 1985,64 (9) : 888 - 896.
  • 6BRAZZEL C. Thromboelastography-guided transfusion Therapy in the trauma patient[ J]. AANA journal,2013, 81(2) :127-132.
  • 7TAPIA N M,CHANG A,NORMAN M,et al. TEG-guided resuscitation is superior to standardized MTP resuscitation in massively transfused penetrating trauma patients [ J ]. Journal of Trauma and Acute Care Surgery,2013,74(2) : 378-386.
  • 8KAUFMANN C R,DWYER K M, CREWS J D, et al. Usefulness of thrombelastography in assessment of trauma patient coagulation [ J ]. Journal of Trauma and Acute Care Surgery, 1997,42 (4) :716-722.
  • 9LUDDINGTON R J. Thrombelastography/thromboelastome- try [ J ]. Clinical &Iaboratory Haematology,21305,27 (2) :81 --90.
  • 10高兴波.智能前照灯系统角度传感器数字电路设计[D].天津:天津大学,2012.

引证文献7

二级引证文献22

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部