床旁防护装置对冠状动脉造影术者辐射防护效果的体模研究 被引量:1

Analysis of radiation dose to operators involved in coronary angiography with radiation protection shields: a phantom study
原文传递
导出
摘要 目的 测量冠状动脉造影8个投照体位在有与无床旁防护装置防护下术者所受辐射剂量,为冠心病介入治疗中减少术者辐射暴露提供参考.方法 在第一及第二术者站位,距地面20至180 cm处,每隔20 cm放置一个实时剂量测量仪.采用冠状动脉造影8个体位投照,测量在有与无床旁防护装置防护下,术者在不同投照体位的不同高度接受辐射剂量情况.结果 在第一术者位,除1.2m高度仍可测到较高剂量(剂量率0.35 ~4.78 mSv/h,屏蔽率27.67% ~ 89.33%),其余各点屏蔽率均在91%以上.左前斜尾位、左前斜位、左前斜头位辐射剂量较高.第二术者位屏蔽率较第一术者位低,剂量峰值可出现在0.8、1.0及1.4m高度(剂量率0.27 ~ 1.86 mSv/h,屏蔽率30.34% ~92.13%).右前斜尾位、左前斜尾位、正头位、左前斜位辐射剂量较高.结论 床旁防护装置防护下,术者在左前斜尾位、左前斜位、左前斜头位、右前斜尾位的辐射暴露较高,应尽量少采用上述投照体位长时间曝光.同时应加强0.8~1.4m高度的辐射防护. Objective To measure the dose to the primary operator and assistant operators by employing eight beam projections commonly used in coronary angiography with and without radiation protection shields in order to supply helpful guidance on radiation protection in cardiac intervention.Methods From 20 to 180 cm above the ground at the primary and assistant operators' locations,a DoseAware personal dose meter was placed in terms of an increment of 20 cm to measure radiation dose.Eight commonly used beam projections were performed,including LAO (left anterior oblique) 45°,RAO (right anterior oblique) 30°,CRAN (cranial)25°,cranial LAO (LAO45°/25°),caudal LAO(LAO45°/ 25°),CAUD (caudal) 25°,cranial RAO (RAO30°/25°),caudal RAO (caudal RAO30°/25°).Under the two different conditions,with or without radiation protection shields,the doses to the operators in the selected beam projections were respectively recorded at nine measuring positions and the shielding factor were calculated.Results The primary operator was effectively protected with radiation protection shields.In the standing area of the primary operator,except for the position at the height of 120 cm (radiation dose rate:0.35-4.78 mSv/h;shielding factor:27.67%-89.33%),the shielding factor for each measuring position was above 91%.Higher radiation doses were found at caudal LAO,LAO,and cranial LAO.The shielding factor for the assisting operator was lower than for the primary operator.In the standing area of the assisting operator (radiation dose rate:0.27-1.86 mSv/h;shielding factor:30.34%-92.13%),the peak levels were found at the height of 80,100,140 cm.And caudal RAO,caudal LAO,CRAN,LAO were found to have received higher radiation doses.Conclusions Emphasis should be attached to the use of radiation shields in coronary angiography.With radiation protection shields,higher dose is still recorded in caudal LAO,LAO,cranial LAO,caudal RAO.Furthermore,it should be paid more attention to radiation protection at 80-140 cm height,and less prolonged exposure should be employed in those beam projections mentioned above.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2015年第8期623-627,共5页 Chinese Journal of Radiological Medicine and Protection
基金 NSFC-广东联合基金重点支持项目(U1401255) 国家“十二五”科技支撑计划项目(2011BAI11B22) 广东省科技支撑计划项目(20098030801257,20138031800006)
关键词 冠状动脉造影 辐射剂量 辐射防护 Coronary angiography Radiation dose Radiation protection
  • 相关文献

参考文献13

  • 1Durn A, Hian SK, Miller DL, et al. A summary of recommendations for occupational radiation protection in interventional cardiology[J]. Catheter Cardiovasc Interv, 2013, 81(3) : 562-567.
  • 2MUller MC, Welle K, Strauss A, et al. Real-time dosimetry reduces radiation exposure of orthopaedic surgeons [ J ]. Orthop Traumatol Surg Res, 2014, 100(8) : 947-951.
  • 3Mesbahi A, Mehnati P, Keshtkar A, et al. Comparison of radiation dose to patient and staff for two interventional cardiology units : a phantom study[ J]. Radiat Prot Dosim, 2008, 131 ( 3 ) : 399 -403.
  • 4Fetterly KA, Magnuson DJ, Tannahill GM, et al. Effective use of radiation shields to minimize operator dose during invasive cardiology procedures [ J ]. JACC Cardiovasc Interv, 2011, 4 (10): 1133-I139.
  • 5Abatzoglou I, Koukourakis M, Konstantinides S. Reduction of the radiation dose received by interventional cardiologists following training in radiation protection[ J]. Radiat Prot Dosim, 2013, 155(1): 119-121.
  • 6宋扬,李正才,陈硕,贺强,陈大伟.DSA冠脉造影致医护人员辐射剂量的研究[J].中国辐射卫生,2014,23(4):329-331. 被引量:13
  • 7Fink GE. Radiation safety in fluoroscopy for neuraxial injections [J]. AANA J, 2009, 77(4): 265-269.
  • 8Miller DL, Vail6 E, Bartal G, et al. Occupational radiation protection in interventional radiology: a joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology [ J]. J Vase Interv Radio/, 2010, 21(5) : 607-615.
  • 9Koukorava C, Carinou E, Simantirakis G, et ah Doses to operators during interventional radiology procedures: focus on eye lens and extremity dosimetry[ J]. Radiat Prot Doslm, 2011, 144 (1-4) : 482-486.
  • 10Haqqani OP, Agarwal PK, Halin NM, et al. Defining the radiation "scatter cloud" in the interventional suite[ J ]. J Vase Surg, 2013, 58(.5): 1339-1345.

二级参考文献24

  • 1郑钧正,贺青华,李述唐,岳保荣,王琪,杨霞,张建峰.我国电离辐射医学应用的基本现状[J].中华放射医学与防护杂志,2001,21(S1):13-20. 被引量:63
  • 2郁鹏,程玉玺,刘澜涛,燕树林,李田昌,尉可道.介入诊疗中重要站立区域辐射剂量的测定与评价[J].中华放射医学与防护杂志,2004,24(6):573-575. 被引量:20
  • 3刘彬,白玫,彭明辰.介入诊疗过程中的确定性效应监控方法[J].医疗设备信息,2006,21(2):25-27. 被引量:9
  • 4[2]Stephen Baiter Douglas W.Fletcher,Hsin M.Kuan.Techniques to estimate radiation dose to skin during fluoroscopically guided procedures.Skin Dose Measurements[M].AAPM July 2002.
  • 5[3]STUK Radiation and Nuclear Safety Authority.Patient Dosimetry Protocols in Digital and Interventional Radiology[J].Radiation Protection Dosimetry Vol.94,No.12,PP.105 -108(2001).
  • 6[5]S van de Putte,MSc,F Verhaegen,PhD,Y Taeymans,mD,PhD and H Thierens,PhD.Correlation of patient skin doses in cardiac interventional radiology with dose-area product[J].BJR,vol 73,PP 504-513(2000).
  • 7[6]International Commission on Radiological Protection.Avoidance of radiation injuries from medical interventional procedures[J].ICRP Pubhcation 85.Ann ICRP 2000;30:7-67.
  • 8[7]A.Servomaa and M.Tapiovaara.Organ dose calculation in medical X ray examinatiom by the program PCXMC[J].Radiation Protection Dosimetry,Vol.80,Nos 1-3,PP.213-219(1998).
  • 9[8]Donald L.Miller,MD,Stephen Balter,PhD,Louis K.Wagner,PhD,John Cardella,MD.Quality Improvement Guidelines for Recording Patient Radiation Dose in the Medical Record[J].Vase Interv Radiol 2004;15:423-429.
  • 10中华人民共和国国家质量监督检验检疫总局,GB18871-2002电离辐射防护与辐射源安全基本标准[S].北京:中国标准出版社,2002

共引文献17

同被引文献7

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部