摘要
目的:评定高效液相色谱-质谱联用法(HPLC-MS/MS)测定人血浆中替加环素(TGC)浓度的不确定度。方法:分析HPLC-MS/MS法测定人血浆中TGC浓度的整个过程,分别评定测量重复性、称量、工作液配制、生物样品配制、回收率、仪器允差和标准曲线拟和引起的不确定度大小,最后计算合成不确定度并进行扩展。结果:人血浆中替加环素低、中、高浓度的测定值和不确定度可分别表示为:(48.737±4.587),(396.067±19.096)和(1 489.000±103.563)ng·ml-1,(P=95%,k=2)。结论:HPLC-MS/MS法测定人血浆中TGC浓度的不确定度主要来源在低浓度时曲线拟合和高浓度时回收率。
Objective: To evaluate the uncertainty of tigecycline (TGC) determination in human plasma by HPLCJMS/MS. Methods: The whole process of TGC determination by HPLC-MS/MS in human plasma was analyzed and the uncertainty caused by re- peatability, weighing, standard solution preparation, biological sample preparation, recovery, instrument precision and calibration curve fitting were evaluated respectively and the combined and expanded uncertainty were calculated. Results: The measured value and the uncertainty of TGC determined at low, medium and high levels in human plasma could be expressed as (48. 737 ± 4.587 ), (396.067 ± 19. 096) and ( 1489. 000 ± 103. 563 ) ng · ml - 1, respectively ( P = 95 %, k = 2). Conclusion: The uncertainty of TGC in human plasma determined by the HPLC-MS/MS method is mainly caused by calibration curve fitting and recovery at low and high concentration.
出处
《中国药师》
CAS
2015年第9期1433-1437,共5页
China Pharmacist
基金
国家自然科学基金项目(编号:81301118)
多肽及小分子药物北京市重点实验室开放课题(编号:2014DTYW03)