冠状动脉造影中汉族人群体重指数与剂量面积乘积和透视时间相关性研究 被引量:3

Study on the relationship between body mass index and dose area product as well as fluoroscopy time in Han nationality during coronary angiography
下载PDF
导出
摘要 目的研究汉族受检人群体重指数(BMI)与冠状动脉造影(CAG)检查辐射剂量面积乘积(DAP)值、透视时间相关性,早期判断和降低高辐射风险。方法回顾性分析451例汉族人群CAG患者资料,男276例,年龄32-87岁;女175例,年龄42—84岁;BMI值17.30~35.42kg/m^2。根据BMI值分为4组,A组25例(BMI〈20kg/m^2),B组204例(BMI20~24.9kg/m^2),C组192例(BMI25~29.9kg/m^2),D组30例(BMI〉30kg/m^2)。计算各组DAP和透视时间均值,采用非参数Kruskal-Wallis检验多重比较各组BMI患者DAP值,单因素方差分析或LSD-t检验比较各组BMI患者透视时间。结果DAP均值在A组为(1070.07±541.33)μGy/m^2,B组为(1326.82±606.91)μGy/m^2,C组为(1937.99±1030.31)μGy/m^2.D组为(2654.53±1296.69)μGy/m^2;透视时间均值在A组为(3.53±2.08)min,B组为(2.70±1.80)min,C组为(2.75±1.88)min,D组为(2.71±1.69)min。组间比较,除A组与B组间DAP值差异无统计学意义(P=0.232),余各组间差异均有统计学意义(P〈0.05);A组与其它各组间透视时间差异均有统计学意义(P〈0.05),余各组间差异均无统计学意义(P〉0.05)。结论成年汉族受检人群CAG手术中DAP值随BMI增高而增加,透视时间在BMI〈20kg/m^2时明显延长。介入手术中不仅要了解BMI增高造成DAP值增加.更需特别关注BMI〈20kg/m^2时手术难度提高所致透视时间延长造成的DAP值增加。 Objective To investigate the relationship between body mass index (BMI) and dose area product (DAP) as well as fluoroscopy time during coronary angiography (CAG) in Han nationality subjects in order to improve early judging and reducing high radiation risk. Methods CAG materials of 451 Han nationality subjects were retrospectively analyzed. The patients included 276 males (age 32-87 years) and 175 females (age 42-84 years), and the BMI values ranged from 17.30 to 35.42 kg/m^2. According to BMI values, the patients were divided into group A(BMI〈20 kg/m^2,n=25) ; group B (BMI 20-24.9 kg/m^2,n=204), group C (BMI 25-29.9 kg/m^2,n=192) and group D (BMI〉30 kg/m^2,n=30). The mean DAP and fluoroscopy time of each group were calculated; the DAP values were compared between each other among the four groups by using nonparametric Kruskal-Wallis test, while the fluoroscopy time was compared between each other among the four groups by using single factor analysis of variance or LSD-t test. Results The mean DAP of group A, B, C and D was(1 070.07±541.33) μGym^2, (1 326.82±606.91) μGym^2, (1 937.99±1 030.31) μGym^2 and (2 654.53±1 296.69) μGym^2 respectively. The mean fluoroscopy time of group A, B, C and D was (3.53± 2.08) min, (2.70±1.80) min, (2.75±1.88) min and (2.71±1.69) min respectively. Statistically significant difference in DAP values existed between each other among the four groups (P〈0.05) except between group A and group B (P=0.232). Statistically significant difference in fluoroscopy time existed between group A and other three groups (P〈0.05), while the difference in fluoroscopy time between each other among B, C and D groups was not significant (P〉0.05). Conclusion The DAP values of adult Han subjects receiving CAG are increased with the increase of BMI values; the fluoroscopy time is significantly prolonged in patients with BMI 〈20 kg/m^2. In performing interventional procedure, the operator should not only understand that the increase of BMI can cause the increase of DAP, but also need to pay special attention to the increase of DAP value caused by operation difficulty when the patient's BMI is 〈20 kg/m^2. (J Intervent Radiol, 2015, 24: 763-766)
出处 《介入放射学杂志》 CSCD 北大核心 2015年第9期763-766,共4页 Journal of Interventional Radiology
关键词 剂量面积乘积 体重指数 透视时间 dose area product body mass index fluoroscopy time
  • 相关文献

参考文献15

  • 1Lloyd- Jones D, Adams R, Carnethon M, et al. Heart disease and stroke statistics-2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee[Z].Circulation, 2009, 119: 480- 486.
  • 2Miller JM, Rochitte CE, Dewey M, et al. Diagnostic performance of coronary angiography by 64- row CT[J].N Engl J Med, 2008, 359: 2324- 2336.
  • 3Chida K, Saito H, Otani H, et al. Relationship between fluoroscopic time, dose- area product, body weight, and maximum radiation skin dose in cardiac interventional procedures[J].AJR Am J Roentgenol, 2006, 186: 774- 778.
  • 4Shah A, Das P, Subkovas E, et al. Radiation dose during coronary angiogram: relation to body mass index[J].Heart Lung Circ, 2015, 24: 21- 25.
  • 5Bahreyni Toossi MT, Zare H, Bayani S, et al. Organ and effective doses of patients arising from coronary angiography and percutaneous transluminal coronary angioplasty at two hospitals in Mashhad- Iran[J].Radiat Prot Dosimetry, 2008, 128: 363- 366.
  • 6李江.医用诊断X射线辐射源空气比释动能率的测量不确定度评定[J].计量与测试技术,2011,38(5):62-63. 被引量:5
  • 7王智廷,曹国全,闻彩云,黄周青.冠状动脉介入治疗中操作者所受剂量的综合研究[J].介入放射学杂志,2013,22(10):834-837. 被引量:5
  • 8Pi- Sunyer FX. Obesity: criteria and classification[J].Proc Nutr Soc, 2000, 59: 505- 509.
  • 9陆再英,钟南山.内科学[M].北京:人民卫生出版社,2008:121.
  • 10冯俊,王爱玲,程景林,郭杰.不同类型心血管介入手术辐射剂量分析[J].中华放射医学与防护杂志,2012,32(4):416-419. 被引量:17

二级参考文献28

  • 1郁鹏,程玉玺,刘澜涛,燕树林,李田昌,尉可道.介入诊疗中重要站立区域辐射剂量的测定与评价[J].中华放射医学与防护杂志,2004,24(6):573-575. 被引量:20
  • 2Chida K, Saito H, Otani H, et al. Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. AJR Am J Roentgenol, 2006,186(3):774-778.
  • 3Chida K, Kato M, Kagaya Y, et al. Radiation dose and radiation protection for patients and physicians during interventional procedure. J Radiat Res, 2010,51(2): 97-105.
  • 4Valentin J. Avoidance of radiation injuries from medical interventional procedures. Ann ICRP,2000,30(2):7-67.
  • 5Miller DL, Balter S, Cole PE, et al. Radiation doses in interventional radiology procedures: the RAD-IR study. PartⅠ: overall measures of dose. J Vasc Interv Radiol, 2003,14(6):711-727.
  • 6Miller DL, Balter S, Cole PE, et al. Radiation doses in interventional radiology procedures: the RAD-IR study. PartⅡ: skin dose. J Vasc Interv Radiol, 2003, 14(8): 977-990.
  • 7Mercuri M, Xie CC, Levy M, et al. Predictors of increased radiation dose during percutaneous coronary intervention. Am J Cardiol, 2009, 104(9): 1241-1244.
  • 8Kuon E, Empen K, Robinson DM, et al. Efficiency of a minicourse in radiation reducing techniques: a pilot initiative to encourage less irradiating cardiological interventional techniques. Heart, 2005, 91(9): 1221-1222.
  • 9Rivolta A, Emanuelli S, Tessarin C, et al. Method of patient dose evaluation in the angiographic and interventional radiology procedures. Radiol Med, 2005, 110(5-6): 689-698.
  • 10Suzuki S, Furui S, Yamakawa T, et al. Radiation exposure to patients’ skin during cardiac resynchronization therapy. Europace, 2009, 11(12): 1683-1688.

共引文献1724

同被引文献17

引证文献3

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部