期刊文献+

不同厚度无托槽隐形矫治器矫治应变力学效果研究 被引量:4

Treating mechanical strain effect of different thickness without brackets invisible appliance
下载PDF
导出
摘要 目的探讨不同厚度无托槽隐形矫治器产生的矫治应变力学效果研究。方法制作厚度为0.75mm和1.00mm无托槽隐形矫治器各10副,将矫治器戴入治疗前上颌光固化树脂模型,位移设计量均为0.2mm。比较两组切牙牙冠唇面和舌侧共11个区的应变情况。结果厚度为0.75mm的无托槽隐形矫治器在切1/3近中、中1/3中、中1/3近中、颈1/3近中、舌侧中央窝、舌侧隆突位置矫治应变大于1.00mm厚度的无托槽隐形矫治器(P<0.01)。结论厚度为0.75mm的无托槽隐形矫治器较1.00mm的无托槽隐形矫治器产生的应力更大。 Objective To discuss treating mechanical strain effect of different thickness without brackets invisible appliance. Methods Each no invisible appliance of 0.75mm and 1.00mm was produced for 10, which were worn on maxillary light-cured resin model, and the amount of displacement design were 0.2mm. Treating mechanical strain effects of incisor crown lingual lip surface and a total of 11 zones of two groups were compared. Results Treating mechanical strain effects of zones of incisor crossing 1/3 mesial, mesial crossing 1/3 mesial, central crossing 1/3 mesial, neck crossing 1/3 mesial, lingual fovea, and lingual protuberance of 0.75mm group were bigger than those of 1.00mm group(P 〈 0.01). Conclusion Treating mechanical strain effect of 0.75mm thickness without brackets invisible appliance is greater than 1.00mm.
作者 邢锐
出处 《中国医药科学》 2014年第23期167-169,共3页 China Medicine And Pharmacy
关键词 无托槽隐形矫治器 矫治应变力学 Without brackets invisible appliance Treating mechanical strain
  • 相关文献

参考文献11

二级参考文献89

  • 1白玉兴,任超超.无托槽隐形矫治技术临床应用中的相关问题[J].中国实用口腔科杂志,2009,2(1):13-16. 被引量:25
  • 2王邦康.口腔正畸矫治方法的新进展——无托槽隐形矫治器的研究与展望[J].北京口腔医学,2005,13(1):2-5. 被引量:27
  • 3王凡,白玉兴,祁鹏,周洁珉,厉松,岐红恩.无托槽隐形矫治尖牙整体移动的生物力学研究[J].医用生物力学,2007,22(2):133-136. 被引量:18
  • 4Suhling JC, Jaeger RC. Silicon piezoresistive stress sensors and their application in electronic packaging. IEEE Sensors J, 2001, 1(1) :14-30.
  • 5Kusy RP. Influence of force systems on archwire-bracket combinations. Am J Orthod Dentofacial Orthop, 2005, 127 (3) : 333-342.
  • 6Tuncay, Orhan C. The Invisalign System. New Malden, UK: Quintessence Publishing Co, Ltd, 2006 : 187-194.
  • 7Lapatki BG, Paul O. Smart brackets for 3D-force-moment measurements in orthodontic research and therapy - developmental status and prospects. J Orofac Orthop,2007,68 (5) :377-396.
  • 8Lapatki BG, Bartholomeyczik J, Ruther P, et al. Smart bracket for multi-dimensional force and moment measurement. J Dent Res, 2007, 86(1) :73-78.
  • 9Sweet JN. Integrated test chips improve IC assembly. IEEE Circuits Dev Mag, 1990,6(5) :39-45.
  • 10Wu JD, Ho SH, Zheng PJ, et al. An experimental study of failure and fatigue life of a stacked CSP subjected to cyclic bending- Electronic Components and Technology Conference, 2001. Proceedings, 51 st. 2002:1081-1086.