摘要
目的:探讨低管电压联合迭代重建技术(IR)在眼眶CT检查中的应用价值。方法 :2014年6月—2015年1月于我院行眼眶CT平扫患者90例,随机分为120 k V组(A组)、100 k V组(B组)、80 k V组(C组),每组30例。A组行滤波反投影(FBP)重建,B组和C组行IR重建(idose3)。对各组进行图像质量主观及客观评价,记录各组辐射剂量。对图像质量及辐射剂量进行统计学分析。结果:A、B、C组各组的有效剂量分别为(7.3±0.7)m Sv、(4.5±0.3)m Sv、(2.2±0.2)m Sv,各组间差异有统计学意义。各组图像均满足诊断要求,图像质量评分均高于3分,且两观察者对图像质量评价的一致性好或非常好。低剂量组的图像背景噪声较常规剂量组高,且差异有统计学意义(P<0.05)。结论:低管电压联合IR技术在眼眶CT检查中能明显降低CT辐射剂量同时图像满足诊断要求。
Objective: To investigate the combined application value of low tube voltage and iterative reconstruction(IR) for CT imaging of the orbit. Methods: From June 2014 to January 2015 in our hospital, 90 patients were referred for orbit CT. Patients were randomly assigned into three groups: 120 kV group(group A), 100 kV group(group B) and 80 kV group(group C), with 30 cases in each group. Group A underwent filtered back projection(FBP) reconstruction while group B and C underwent IR reconstruction. Subjective and objective evaluations were performed on image quality and radiation dose was recorded for each group, which were statistically analyzed. Results: The average effective dose of group A, B and C were (7.3±0.7) mSv, (4.5±0.3) mSv and (2.2±0.2) mSv respectively. There were significant difference among groups. The images of the three groups all met the diagnostic requirement. All image quality scores were higher than 3 points. The consistency of the two observers in evaluating image quality was either good or very good. The image background noise of low-dose group was higher than that of standard dose group, and the difference was statistically significant (P〈0.05). Conclusion: Low tube voltage combined with IR algorithm in orbit CT examination can significantly reduce radiation dose, and the image quality can meet the diag- nostic requirement.
出处
《中国临床医学影像杂志》
CAS
北大核心
2015年第9期623-626,630,共5页
Journal of China Clinic Medical Imaging