期刊文献+

64层螺旋CT空间分辨力评价方法解析 被引量:4

Spatial resolution evaluation of 64-slice helical computed tomography
下载PDF
导出
摘要 目的:针对目前CT空间分辨力的检验现状,提出更为科学严谨的64层螺旋CT空间分辨力检测要求。方法 :对现行行业标准中规定的空间分辨力测试要求进行分析,用试验数据来说明现行标准中存在的不足,同时提出64层螺旋CT空间分辨力的检测要求,并分析影响空间分辨力的因素。结果:64层螺旋CT空间分辨力测试要求能够很好地评价空间分辨力,避免了原有标准测试要求中存在的不足。结论:64层螺旋空间分辨力检测要求的提出,能够更好、更科学地评价其空间分辨力,保证图像质量更好地应用于临床。 Objective To put forward spatial resolution detection requirements of 64-slice helical CT based on analyzing the present situation of CT spatial resolution detection.Methods The requirements for spatial resolution in the existing industrial standards were analyzed, and some test data were used to prove the deficiency of the standards. Spatial resolu-tion detection requirements were proposed for 64-slice helical CT, and the factors affecting spatial resolution were ex-plored. Results The deficiencies in the original standards were eliminated. Conclusion The proposed requirements may facilitate the spatial resolution detection of 64-slice CT, and thus can enhance image quality for clinical diagnosis and treatment.
出处 《医疗卫生装备》 CAS 2015年第9期93-94,100,共3页 Chinese Medical Equipment Journal
基金 国家科技支撑计划项目(2012BAI22B00)
关键词 64层螺旋CT 空间分辨力 调制传递函数 64层螺旋CT 64-slice helical CT spatial resolution modulation transfer function
  • 相关文献

参考文献9

  • 1张敬理.医疗器械监管技术基础[M].北京:中国医药工业出版社,2009:421-424.
  • 2YY 0310-2005 X射线计算机体层摄影设备通用技术条件[S].
  • 3GB/T 19042.5-2006医用成像部门的评价及例行试验第3-5部分:X射线计算机体层摄影设备成像性能验收试验[S].
  • 4施裕新,陆普选,张志勇.影像医疗器械临床试验手册[M].北京:人民卫生出版社,2013:100-101.
  • 5刘传亚,张建春,吕守臣.多层CT螺距与z轴空间分辨率的关系探讨[J].医学影像学杂志,2010,20(10):1559-1561. 被引量:4
  • 6郭智敏,倪培君.CT系统空间分辨率测试方法研究进展[J].兵器材料科学与工程,2010,33(2):113-117. 被引量:13
  • 7Droge R T,Morin R L.A practical method to measure the MTF of CT scanners[J].Med Phys,1982,5(9):758-760.
  • 8GB9706.18-2006医用电气设备第2部分:X射线计算机体层摄影设备安全专用要求[S].
  • 9IEC 60601-2-44:2012 Medical electrical equipment-part 2-44:particular requirements for the basic safety and essential performance of X-ray equipment for computed tomography[S].

二级参考文献24

  • 1曾祥照.图像分辨率测试卡的结构与标准[J].CT理论与应用研究(中英文),2005,14(3):29-34. 被引量:5
  • 2杨克柽,林意群.CT图像空间分辨力的实用检测方法[J].放射学实践,2006,21(3):303-305. 被引量:7
  • 3刘传亚,秦维昌,王巍,卢传友.64层螺旋CT的层敏感度曲线测试[J].中华放射学杂志,2006,40(11):1210-1212. 被引量:11
  • 4国防科学技术工业委员会.GJB5311-2004工业CT系统性能测试方法[S].北京:国防科学技术工业委员会出版发行部,2004.
  • 5国防科学技术工业委员会.GJB5312-2004.工业射线层析成像(CT)检测[S].北京:国防科工委军标出版发行部.2004.
  • 6OPRAX MEDICAL. Line pair/resolution test phantom [EB/ OL].American http ://www.opraxmedical.com/ Accessories/ Phantoms/Test/Linepair.
  • 7The Phantom laboratory Inc,Catphan manual[EB/OL]. American[ 1997-07]. http//www.phantomlab.com/catphan.html.
  • 8Japanese Industrial Standard. Z 4916-1997. Test charts for resolution characteristics of X-ray equipment[S]. Japan, 1997.
  • 9QRM. 3D spatial resolution phantom [EB/OL]. Germany http: //www.qrm.de/content/pdf/QRM-3 DSR.pdf.
  • 10Hye Jung Park,Seung Eun Jung,et al. The relationship between subjective and objective parameters in CT phantom image evaluation [ J ]. Korean J Radiol, 2009,10 (5) : 490-495.