摘要 利用一款基于瞬发γ中子活化原理设计的主要针对皮带散状物料元素进行在线检测的工业仪器,研究这款仪器用于测量铝土矿组成分析的可行性,实验主要进行了两组样品中Ca、Si、Al、Fe、Ti五种元素的测试,其中一组样品由高纯氧化物组成,一组样品由铝土矿和高纯氧化物组成,测量结果表明,CaO、SiO2、Al2O3、Fe2O3、TiO2的测量精度分别为±0.23%、±0.18%、±0.35%、±0.14%、±0.08%,能够满足原料自动配比及入厂原料质量控制的要求。 An industrial prompt-gamma neutron activation analysis(PGNAA) system, designed for the real-time elemental analysis of bulk materiel on a conveyor belt,has been studied to examine the feasibility of using such a system for bauxite composition analysis. Experiments were conducted to measure Ca Si Fe A1 in two simple,one simple is comprised of pure oxide, and the other simple is comprised of bauxite and pure oxide. The static accuracy are ± 0. 23% (CaO), ± 0. 18% (SiO2), ± 0. 35%(Al2 O3 ), ± 0. 14% (Fe2 03 ), ± 0.08 % (TiO2). The instrument can meet the requirements of automatic mixing system of row material and quality controlling after entering factory.
出处 《中国矿业》 北大核心 2015年第10期171-174,共4页 China Mining Magazine
基金 国家重大科学仪器设备开发专项资助(编号:2012YQ-240121)
关键词 中子活化 瞬发γ 铝土矿 neutron activation prompt-gamma bauxite
  • 相关文献

参考文献8

  • 1李春潮,吕子剑.我国氧化铝行业理化检测现状及发展趋势[J].轻金属,2006(1):21-26. 被引量:2
  • 2D. L. Chichester, E. Empey. Measurement of nitrogen in the body using a commercial PGNAA system-phantom experiments[J]. Applied Radiation and Isotopes, 2004,60 : 55-61.
  • 3张兰芝,倪邦发,田伟之,黄东辉,张桂英,刘存兄,王平生,刘立坤,李德红.瞬发γ射线中子活化分析的现状与发展[J].原子能科学技术,2005,39(3):282-288. 被引量:33
  • 4Kheli R, Idiri Z, Omari L, et al. Prompt Gamma Neutron Activation Analysis of Bulk Concrete Samples With an Am-Be Neutron Source[J]. Applied Radiation and Isotopes, 1999,51 (1),913.
  • 5Naqvi AA, Nagadi MM, Omar S, et al. Elemental Analysis of Concrete Samples Using an Accelerator-based PGNAA Setup [J]. Nuclear Instruments and Methods in Physics Research B, 2004,2005(3): 331-338.
  • 6Nair A GC,Sudarshan K,Raje N,et al. Analysis of Alloys by Prompt Gamma-ray Neutron Activation [ J ]. Nuclear Instruments and Methods in Physics Research A, 2004,516 (1) :143- 148.
  • 7Carlos Oliveiral, Josea Salgadol, Francisco Leitaa' O. Density and Water Content Corrections in the Gamma Count Rate of a PGNAA System for Cement Raw Material Analysis Using the MCNP Code [ J]. Applied Radiation and Isotopes, 1998, 49 (8) :923-930.
  • 8宋兆龙,瞿国圣,谈剑,等.火力厂入炉煤质在线检测装置的研究开发[C]//中国核学会核技术应用分会成立大会暨学术报告会论文集.北京:中国核学会核技术工业应用分会,2004.

二级参考文献43

  • 1联合法生产氧化铝控制分析[M].冶金工业出版社,1977.
  • 2OJ/SL.01.07.12—88..山东铝厂技术标准..,,....
  • 3Q/ZLHNB-2002.中国铝业河南分公司企业标准.,..
  • 4[美]罗伯特 D·布朗著.最新仪器分析技术全书[M].化学工业出版社,1990.9.
  • 5山东.全国第八届轻金属分析学术会议论文集[Z].中国有色金属分析情报网轻金属分网,2001-5.
  • 6山西.全国第七届轻金属分析学术会议论文集[Z].中国有色金属分析情报网轻金属分网,1998-10.
  • 7Byun SH, Choi HD. Design Features of a Prompt Gamma Neutron Activation Analysis at HANARO[J]. J Radioanal Nucl Chem, 2000,244(2) : 413-416.
  • 8Molnar G, Belgya T, Dabolczi L. The New Prompt Gamma-activation Analysis Facility at Budapest[J]. J Radioanal Nucl Chem, 1997, 215(1): 111-115.
  • 9Paul RL, Lindstrom RM, Heald AE. Cold Neutron Prompt Gamma-ray Activation Analysis at NIST-recent Development[J]. J Radioanal Nucl Chem, 1997, 215(1): 63-68.